SPACELABS MEDICAL 90482 BISPECTRAL INDEX ANALYSIS (BIS) MODULE
K973596 · Spacelabs Medical, Inc. · OLW · Feb 3, 1998 · Neurology
Device Facts
Record ID
K973596
Device Name
SPACELABS MEDICAL 90482 BISPECTRAL INDEX ANALYSIS (BIS) MODULE
Applicant
Spacelabs Medical, Inc.
Product Code
OLW · Neurology
Decision Date
Feb 3, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Indications for Use
The Spacelabs Medical 90482 BIS Module is intended to monitor the state of the brain by data acquisition of EEG signals in the intensive care unit, operating room and for clinical research. The Bispectral Index (BIS), a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents.
Device Story
Module acquires/processes 1-2 channels of EEG signals; derives Bispectral Index (BIS) to correlate EEG activity to brain state. Calculates spectral edge frequency, electromyogram activity, median power frequency, signal quality index, and EEG suppression. Uses licensed Aspect Medical Systems bispectral analysis algorithm and proprietary sensor. Integrates with Spacelabs Medical Patient Care Information System (PCIS) via SDLC communications for waveform/digital data display, review, and editing. Used in ICU, OR, and research settings by clinicians. Provides real-time monitoring of anesthetic effects; aids clinical decision-making regarding anesthesia depth.
Clinical Evidence
Bench testing only. Performance testing conducted to ensure device meets functional specifications and safety standards, including IEC 601-1 and IEC 601-2-26 for electroencephalographs.
Technological Characteristics
Modular unit; interfaces with Spacelabs Medical PCIS. Uses proprietary Aspect Medical sensor for EEG acquisition. Bispectral analysis algorithm licensed from Aspect Medical. Connectivity via SDLC communications. Complies with IEC 601-1 and IEC 601-2-26 safety standards.
Indications for Use
Indicated for monitoring brain state via EEG signal acquisition in ICU, OR, and clinical research settings. Used as an aid in monitoring anesthetic agent effects. Prescription use only. Applied externally to the scalp.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
EEG Monitor, Models A-1000/A-1050 with the Processed Monitor (K963644)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Spacelabs Medical. The logo consists of the word "Spacelabs" in a bold, sans-serif font, with the word "medical" in a smaller font below it. Above the word "Spacelabs" is a stylized graphic that resembles a curved line or arc. The logo is black and white.
FEB - 3 1998 ×973596
15220 N.E. 40th Street
P.O. Box 97013
Redmond, Washington 98073-9713
206-882-3700
## 510(k) SAFETY AND EFFECTIVENESS SUMMARY
## Spacelabs Medical 90482 Bispectral Index Analysis (BIS) Module
- Russ Garrison Submitter's Name: 1. Director of Regulatory Affairs
- Spacelabs Medical Inc. Company: 15220 N.E. 40th Street Redmond, WA 98073
- (425) 882-3913 Telephone:
- (425) 867-3550 Facsimile:
- Spacelabs Medical 90482 BIS Module Name of Device: 2.
Flectroencephalograph (EEG) Monitor Classification: OLW, ORF, OLT, ONC
- We consider the Spacelabs Medical 90482 BIS Predicate Device: 3. Module to be substantially equivalent to the EEG Monitor, Models A-1000/A-1050 with the Processed Monitor, Models, currently marketed by Aspect Medical Systems, Inc. (510[k] reference K963644). Both monitoring systems provide the means to monitor the state of the brain by data acquisition and processing of EEG signals in a clinical hospital environment or research setting. Both also use the identical bispectral analysis algorithm, a processed EEG variable displayed as a numerical value, as an aid in monitoring the effects of certain anesthetic agents. The bispectral analysis algorithm used in the Spacelabs Medical 90482 BIS Module has been licensed from Aspect Medical Systems and the 90482 BIS Module uses a Digital Signal Converter (DSC-2) and a proprietary sensor developed by Aspect Medical to operate in the BIS mode.
{1}------------------------------------------------
- The Spacelabs Medical 90482 BIS Module is a slim, Device Description: 4. lightweight singular modular unit that, when used in conjunction with a Spacelabs Medical Patient Care Information System (PCIS), provides the capability to acquire, process and display one (1) or two (2) channels of brain electrical activity (EEG). The EEG data is processed to derive a Bispectral Index (BIS) to correlate the EEG activity to the state of the brain. Other trended parameters may also be processed, including spectral edge frequency, electromyogram activity, median power frequency, a signal quality index, and suppression of the EEG signal.
The Module is the primary interface to the patient The Module is capable of being monitored. acquiring and processing EEG signals for a single patient. The Module accumulates the patient physiological data of interest and provides both waveform and digital data to a Spacelabs Medical Patient Care Information System (PCIS) monitor via SDLC communications. The PCIS system provides the display, review and editing capabilities for the care provider.
- The Spacelabs Medical 90482 BIS Module is Intended Use: 5. intended to monitor the state of the brain by data acquisition of EEG signals in the intensive care unit, operating room and for clinical research. The Bispectral Index (BIS), a processed EEG variable, may be used as an aid in monitoring the effects of certain anesthetic agents.
- The design, components, storage technology and Comparison of 6. energy source of the 90482 BIS Module are similar Technological to its predicate device. Both the 90482 BIS Module Characteristics: and the Aspect EEG/BIS Monitor provide the means for interfacing with a patient, collecting EEG data, and processing the data for alarm generation and display of EEG waveforms and numeric BIS values on a bedside or central monitoring system. The bispectral analysis algorithm has been licensed from Aspect Medical and, to operate in the BIS mode, a specific proprietary sensor provided by Aspect Medical Systems must be used.
{2}------------------------------------------------
The only significant difference between the The only significant BIS Module and the comparable Spacelabs Medical by Aspect Medical are in the system Tharketed By Rop 90482 BIS Module, which hardware packaging of the spatibility with the Patient is designed for modular companisms offered by Spacelabs Medical.
- The Spacelabs Medical 90482 BIS Module will be Testing: The Spacellabs Medical be loss performance testing 7. prior to release. Final testing for the system includes various performance tests designed to includes various peno manos all of its functional ensure that the device morning as specifications. Safety testing will be performed by third party Safety lesting will be pononewice complies to agencies - to - enours safety standards for medical applicable industry and only and on 1, C222.2 No. 601-1, and s, including OELOS for for the safety of IEC IEC 601-1.
electroencephalographs specifically, IEC 601-2-26.
In conclusion, the Spacelabs Medical 90482 BIS Module is as safe and offective In conclusion, the Spacelabs Medical School by Aspect Medical Systems and as the prodicate state or effectiveness issues.
.ભૂર-
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Russ Garrison Director of Regulatory Affairs Spacelabs Medical, Incorporated 15220 North East 40th Street Redmond, Washington 98073
APR - 9 2012
Re: K973596
Trade/Device Name: Spacelabs Medical 90482 Bispectral Index Analysis (BIS) Module Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLW, ORT, OLT, OLT, OMC Dated (Date on orig SE ltr): December 18, 1997 Received (Date on orig SE Itr): December 22, 1997
Dear Mr. Garrison:
This letter corrects our substantially equivalent letter of February 3, 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportalroblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE
Page 1 of 1
510(k) Number (if known): Not Known (New Submission)
Device Name: Spacelabs Medical 90482 Bispectral Index Analysis (BIS) Module
Indications for Use:
1
Condition to be screened, monitored, treated or diagnosed. Coridition to be sereoned, monito, abnormalities in electroencephalograph (EEG) signals in the brain.
Prescription use only.
Prescription asc offity.
Yes. Caution statement is provided in the introductory page of the Patient Tes. Gaution System which includes the operating instructions for this Module.
Parts of body applied to . . ... ... ... .. .. . Faris or body upplies to the Module are applied externally to the scalp.
Frequency of use. Frequency as directed by physician.
Physiological purpose.
- To monitor the state of the brain by data acquisition of EEG signals ● in the intensive care unit, operating room and for clinical research, and
- To display a Bispectral Index (BIS), a processed EEG variable, ● which may be used as an aid in monitoring the effects of certain anesthetic agents.
(Please Do Not WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter Use (Optional Format 1-2-96) |
|---------------------------------------|----|-----------------------------------------------|
|---------------------------------------|----|-----------------------------------------------|
(Division Sign-Off)
Division of General Restorative Devices
| 510(k) Number | KA73596 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.