MEDTRONIC INSTENT ESOPHACOIL ESOPHAGEAL STENT SYSTEM AND ESOPHACOIL EXTENSION SYSTEM

K973584 · Medtech Instruments · ESW · Jan 9, 1998 · General, Plastic Surgery

Device Facts

Record IDK973584
Device NameMEDTRONIC INSTENT ESOPHACOIL ESOPHAGEAL STENT SYSTEM AND ESOPHACOIL EXTENSION SYSTEM
ApplicantMedtech Instruments
Product CodeESW · General, Plastic Surgery
Decision DateJan 9, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3610
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Medtronic InStent EsophaCoil Esophageal Stent System is intended for use in the treatment of esophageal obstructions produced by malignant neoplasms. The Medtronic InStent EsophaCoil Proximal and Distal Extensions are intended for extending previously placed esophageal stents exhibiting tumor overgrowth or tumor ingrowth on the upper or lower portion of the existing stent respectively.

Device Story

Self-expanding, tightly wound spiral Nitinol wire coil stent; flared ends for fixation. Extension systems feature single flared end for attachment to existing stents. Delivery catheter includes variable release handle with three independently sliding levers; allows sequential release (distal, center, or proximal). Used in clinical setting for esophageal obstruction treatment. Physician operates delivery system to deploy stent; provides mechanical patency to esophagus. Benefits patient by relieving obstruction caused by malignant neoplasms.

Clinical Evidence

Bench testing only. Corrosion testing performed to evaluate safety. No clinical data provided.

Technological Characteristics

Self-expanding Nitinol wire coil; flared ends. Delivery system with variable release handle. Materials and sterilization methods identical to predicate. No software or electronic components.

Indications for Use

Indicated for patients with esophageal obstructions caused by malignant neoplasms. Extension systems indicated for patients with previously implanted esophageal stents experiencing tumor overgrowth or ingrowth at the proximal or distal ends.

Regulatory Classification

Identification

An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.

Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN - 8 1988 K973584 # 510(k) NOTIFICATION SUMMARY Medtronic InStent EsophaCoil Esophageal Stent System and EsophaCoil Proximal and Distal Extension Systems # General Provisions A. | Submitter's Name | Medtronic InStent | |-----------------------|----------------------------------------------------------------------------------------| | Submitter's Address | 6271 Bury Drive<br>Eden Prairie, Minnesota 55346 | | Contact Person | John R. Dalpee<br>Senior Quality Assurance/Regulatory Affairs<br>Manager | | Classification Name: | Esophageal Endoprosthesis 21 CFR Part 878.3610 | | Common or Usual Name: | Esophageal Endoprosthesis | | Proprietary Name: | EsophaCoil Esophageal Stent System<br>EsophaCoil Proximal and Distal Extension Systems | ## Name of Predicate Device B. These devices are considered substantially equivalent to the Medtronic InStent EsophaCoil Esophageal Stent System. (K944187 and K955041) ## C. Device Description The Medtronic InStent EsophaCoil Esophageal Stent is a self-expanding, tightly wound spiral Nitinol wire coil which is flared at both ends to enhance fixation. The EsophaCoil Extension is identical to the EsophaCoil Esophageal Stent except that only one end is flared. The non-flared end of the EsophaCoil Extension is inserted into an implanted esophageal stent to extend the length of the stent. The delivery catheter is identical to the currently marketed device. The variable release handle utilizes three independently sliding levers located on the catheter 's proximal end that can release the stent in any sequence order; distal release, center release, or proximal release depending upon which of the sliding levers is pulled first, second, or third. ## D. Intended Use The Medtronic InStent EsophaCoil Esophageal Stent System is intended for use in the treatment of esophageal obstructions produced by malignant neoplasms. The Medtronic InStent EsophaCoil Proximal and Distal Extensions are intended for extending previously placed esophageal stents exhibiting tumor overgrowth or tumor ingrowth on the upper or lower portion of the existing stent respectively. {1}------------------------------------------------ # Summary of Technological Characteristics E. The modified EsophaCoil Esophageal Stent is identical to the currently marketed device with the exception of the surface preparation. The EsophaCoil Extension Systems are identical to the currently marketed device with the exception that the EsophaCoil Extension has only one flared end, rather than two flared ends as in the EsophaCoil Esophageal Stent. #### F. Non-Clinical and Clinical Test Summary Corrosion testing was performed to evaluate the safety of the device. No other testing was deemed necessary for the new or modified devices. #### G. Conclusions Based on the testing performed and the nature of the changes, it was concluded that the modified devices are substantially equivalent to the currently marketed device. The construction, materials packaging and sterilization methods #### H. Other Information No other information was deemed necessary for the determination of substantial equivalence of the EsophaCoil Esophageal Stent modifications or the new models, the EsophaCoil Distal and Proximal Extension Systems. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing its wings, positioned to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN - 9 1998 Mr. John R. Dalpee Senior Quality Assurance/Regulatory Affairs Manager Medtronic InStent 6271 Bury Drive Eden Prairie, Minnesota 55346 Re: K973584 > Trade Name: Medtronic InStent EsophaCoil Esophageal Stent System and EsophaCoil Extension System Regulatory Class: III Product Code: EWZ Dated: September 19, 1997 Received: September 22, 1997 Dear Mr. Dalpee: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations - Title 21, -----Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for {3}------------------------------------------------ Page 2 - Mr. John R. Dalpee devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page of of ) 510(k) Number (if known): Device Name: Medtronic InStent EsophaCoil Esophageal Stent System and EsophaCoil Extension System Indications For Use: The Medtronic InStent EsophaCoil Esophageal Stent System is intended for use in the treatment of esophageal obstructions produced by malignant neoplasms. The Medtronic InStent EsophaCoil Distal Extension System is intended for use in extending previously implanted esophageal stents exhibiting tumor overgrowth on the distal side of the stent. 1 The Medtronic InStent EsophaCoil Proximal Extension System is intended for use in extending previously implanted esophageal stents exhibiting turnor overgrowth on the _ proximal side of the stent. (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sign-Off) | | |-----------------------------------------|--| | Division of General Restorative Devices | | | 510(k) Number | | | Prescription Use | | |----------------------|--| | (Per 21 CFR 801.109) | | OR | Over-The-Counter Use | | |--------------------------|--| | (Optional Format 1-2-96) | |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...