K973527 · Data Critical Corp. · MSX · Mar 31, 1998 · Cardiovascular
Device Facts
Record ID
K973527
Device Name
CARDIO-PAGER SYSTEM
Applicant
Data Critical Corp.
Product Code
MSX · Cardiovascular
Decision Date
Mar 31, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Cardio-Pager™ System described in this submission is a paging system that interfaces with Primary Patient Monitoring Systems in order to provide a secondary means of annunciating and displaying patient alarm information to mobile health care providers. This pager is intended to serves as a parallel, redundant mechanism to inform the clinical staff of patient events. This device is indicated for use in real-time monitoring of routine patient status and alarm events. The Cardio-Pager™ is limited to use by qualified medical professionals who have been trained on the use of the device. It is intended for use only in a hospital environment.
Device Story
Cardio-Pager™ (StatView) is a secondary alarm notification system for hospital environments; transmits text and waveform data from primary patient monitoring networks to mobile graphical pagers. System components: Critical-Server (IBM-compatible PC) connected to patient-monitoring network; Transmission Subsystem (paging transmitter, serial-to-Ethernet converter, receiver units). Server acquires patient data, manages paging initiation, formats data, and sends to transmitter via LAN. Transmitter broadcasts via standard POCSAG protocol. Clinicians receive alerts and periodic heart rate/waveform updates on pagers, allowing awareness of patient status away from bedside/central monitors. Serves as parallel, redundant mechanism to primary alarm enunciators; does not interrupt primary monitoring system operation. Benefits include improved clinician mobility and timely notification of critical patient events.
Clinical Evidence
Bench testing only. Software validation and verification performed using Data Acquisition (DA) module simulator. Systems-level testing conducted for Marquette Unity Network and Hewlett-Packard CareNet configurations. Tests challenged functional requirements: receipt/transmission of alarms, scheduled updates, and text messages. Communication requirements between Cardio-Pager and various Marquette/HP patient monitoring devices were verified.
Technological Characteristics
System consists of IBM-compatible PC server, serial-to-Ethernet converter, paging transmitter, and graphical pager units. Uses standard POCSAG paging protocol for wireless communication. Interfaces with Marquette Unity Network and Hewlett-Packard CareNet. Software-based data acquisition and transmission modules. No specific materials or sterilization requirements noted.
Indications for Use
Indicated for real-time monitoring of routine patient status and alarm events in hospital environments. Limited to use by trained, qualified medical professionals.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
K973527
Image /page/0/Picture/1 description: The image shows the logo for Data Critical Corp. The logo features a stylized graphic above the company name. The graphic consists of several horizontal lines with a circle in the middle, creating a three-dimensional effect. The text "Data Critical Corp." is written in a serif font.
System 510(k) -- K973527 lemental Information tial- Data Critical Corp. Redmond, WA
MAR 3 1 1998
# Summary of Safety and Effectiveness
This summary of safety and effectiveness is submitted for public release in accordance with the requirements of SMDA 1990 and 21 CFR 820.92.
Submitter's Name:
Data Critical Corp. 2733 152nd Ave. NE Redmond, WA 98052 (425) 885-3500
Contact Person Drew D. Weaver, Director, Regulatory Affairs Data Critical Corp. (206) 885-3500
Date of Summary:
Cardio-Pager™ System Device Name:
(74DRG) 21 CFR 870.2910 Classification Name: Radiofrequency physiological signal transmitter and receiver
Legally Marketed Devices to Which Substantial Equivalence is Claimed:
February 3, 1998
- MMS Pager-LAN Paging System 1) Marquette Medical Systems Milwaukee, WI K962827
- 2) Palmvue Hewlett Packard K945277
Marquette Impact.wf 3) Marquette Medical Systems Milwaukee, WI K971868
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Data Critical Corp. The logo consists of a black abstract shape above the company name. The company name is written in a serif font. The abstract shape appears to be a stylized representation of a sun or a burst of energy.
#### Device Description:
The Cardio-Pager™ System (also referred to as StatView) is a secondary alarm notification system that transmits text and waveform data from patient monitoring devices to a graphical pager worn by a trained professional in a hospital environment. The system monitors the patient monitoring devices for alarm conditions, and responds by transmitting a page that includes text and waveform information from the patient. This system is not intended to be a primary alarm enunciator. The Cardio-Pager™ System alerts and informs the clinician about critical patient alarms without requiring them to be at or near the bedside or central monitors. The system can also provide periodic updates of heart rate and waveform data at intervals established by the caregivers.
The primary components in the Cardio-Pager™ System are the Critical-Server (also referred to as the WT Server) and the Transmission Subsystem. The server is a standard IBM-compatible PC that connects directly to the patient-monitoring network. The Transmission Subsystem consists of the paging transmitter, a standard serial-to-Ethernet converter, and the Cardio-Pager™ receiver units.
Software modules in the Critical-Server acquire patient information from the patient monitoring network, manage the initiation of paging messages, convert the data to proper format for transmission to the paging transmitter, and send the message to the paging transmitter via a LAN. The paging transmitter then sends the page, using standard POCSAG paging protocol. One or more of the Cardio-Pager™ receiver units may receive the page.
Addition of the Cardio-Pager™ System to the patient monitoring computer network does not in any way affect, modify, or interrupt the normal operation of the hospital patient monitoring system.
#### Intended Use:
The Cardio-Pager™ System described in this submission is a paging system that interfaces with Primary Patient Monitoring Systems in order to provide a secondary means of annunciating and displaying patient alarm information to mobile health care providers. This pager is intended to serves as a parallel, redundant mechanism to inform the clinical staff of patient events.
This device is indicated for use in real-time monitoring of routine patient status and alarm events. The Cardio-Pager™ is limited to use by qualified medical professionals who have been trained on the use of the device. It is intended for use only in a hospital
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white drawing of a dark object with a light circle in the middle. The dark object is rectangular in shape and has a rough texture. The light circle is located in the center of the dark object and is surrounded by a dark halo. The image is simple and abstract.
Data Critical Corp.
environment. The device has been validated for use with the Marquette Unity Network and the Hewlett-Packard Systems CareNet System.
### Descriptive Summary of Technological Characteristics and Those of Predicate
The Cardio-Pager™ System has the same basic intended use as the predicate devices, namely to extend access to information from patient monitoring systems to mobile caregivers. Cardio-Pager™ is the same device as the Marquette Impact.wf. The Cardio-Pager™ System has identifical technological characteristics and is intended for use in the same application as the Marquette Impact.wf device. Both the Marquette Impact.wf and the Cardio-Pager serve as a secondary means of annunciating patient events and relaying information from the primary monitoring station through the use of a graphical interface paging system. The Cardio-Pager™ combines the alarm annunciation features of Marquette's Pager-LAN with the waveform paging capabilities of the Hewlett-Packard Palm Vue. The underlying wireless communication technology, standard alpha-numeric paging based on the POCSAG protocol. is identical in all three devices. The information provided to the caregiver in Cardio-Pager™ is a combination of the textual alarm information provided by Pager-LAN and the waveform information provided by Palm Vue. Essentially, the Cardio-Pager™ System enhances the functionality of the Pager-LAN by adding the important patient waveform to the information set available to the mobile caregiver.
#### Performance Data
Following software validation and verification activities, two types of performance testing were conducted for the Cardio-Pager™ System. First, the software was tested using a Data Acquisition (DA) module simulator to test the software in simulated conditions using no specific manufacturer's patient monitoring equipment (PME). The second type of testing consisted of finished device or "systems level" testing. Systems level testing was performed for two specific Cardio-Pager™ configurations: One with the DA module customized to facilitate interface with the Marquette Unity Network and the other with the DA module customized to facilitate interface with the Hewlett-Packard CareNet System. Both performance tests challenged all functional requirements for the Cardio-Pager™ System including: the receipt and transmission of patient alarms, the request, receipt and transmission of scheduled updates, the transmission of text messages, and the Cardio-Pager™ receipt of transmitted alarms, scheduled updates, and text messages. In addition, all communication requirements between the Cardio-Pager System™ and Marquette or Hewlett-Packard patient monitoring equipment (PME) were challenged by testing with a variety of Marquette and Hewlett-Packard patient monitoring devices.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 3 1 1998
Mr. Drew D. Weaver Data Critical Corporation 2733 152nd Avenue, NE Redmond, WA 98052
Re: K973527 Cardio-Pager™ System Regulatory Class: III (three) Product Code: 74 MSX Dated: February 18, 1998 February 19, 1998 Received:
Dear Mr. Weaver:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices : ____ General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Drew D. Weaver
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Data Critical Corp. The logo features a black graphic element above the company name. The graphic element appears to be a stylized representation of a data storage device or a critical component. The text "Data Critical Corp." is written in a serif font.
Cardio-Pager™ System 510(k)
Supplemental Information
Confidential-Data Critical Corp.
Redmond, WA
510(k) Number (if known): K973527
Device Name:
ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ
Cardio-Pager™ System
## Indications for Use:
The Cardio-Pager™ System described in this submission is a paging system that interfaces with primary patient monitoring systems in order to provide a secondary means of annunciating and displaying patient alarm information to mobile health care providers. The Cardio-Pager™ System has been validated for use with the Marquette Unity Network and the Hewlett-Packard System CareNet. This pager is intended to serve as a parallel, redundant mechanism to inform the clinical staff of patient events.
This device is indicated for use in real-time monitoring of routine patient status and alarm events. The Cardio-Pager™ is limited to use by qualified medical professionals who have been trained on the use of the device. It is intended for use only in a hospital environment.
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR Over-The-Counter Use Per 21 CFR 801.109 Jur Division Sign-Off)
Division of Cardiovascular, Respiratory,
Division of Cardiovascular, Devices 1157 Division of Cardion.
Division of Cardional Devices and Neurological Devices - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.