VARIABLE PARALLEL PIN
Device Facts
| Record ID | K973521 |
|---|---|
| Device Name | VARIABLE PARALLEL PIN |
| Applicant | Sulzer Calcitek, Inc. |
| Product Code | DZE · Dental |
| Decision Date | Nov 20, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
Indications for Use
The Variable Parallel Pin is intended to provide a visual reference for evaluating parallelism, prosthetic fit and implant spacing between implants and natural teeth during implant placement surgery.
Device Story
Variable Parallel Pin; manual dental surgical instrument. Consists of two pivoting pegs; adjustable orientation from 0 to 25 degrees. Cuff width geometry matches prosthetic abutment flare diameters. Used by dentists during implant placement surgery to provide visual reference for parallelism, prosthetic fit, and implant spacing. Device provides physical alignment guidance; aids clinician in assessing spatial relationships between implants and natural teeth. Benefits patient by ensuring accurate implant placement and improved prosthetic fit.
Clinical Evidence
Bench testing only.
Technological Characteristics
Manual dental instrument; pivoting peg design; adjustable angle (0-25 degrees); cuff geometry matches prosthetic abutment diameters. No software or energy source.
Indications for Use
Indicated for evaluating parallelism and prosthetic fit for dental implants and adjacent natural teeth, accommodating angles up to 25 degrees.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- Sulzer Calcitek's Parallel Pins
- Nobelpharma BRANEMARK SYSTEM® Abutment Selection Kit