K973516 · Terumo Medical Corp. · KDI · Jan 30, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K973516
Device Name
CAPIOX HEMOCONCENTRATOR
Applicant
Terumo Medical Corp.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Jan 30, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CAPIOX Hemoconcentrator is designed to remove excess fluid from the blood in order to maintain proper hematocrit and protein concentration during cardiopulmonary bypass and to enable reinfusion of blood remaining in the circuit after bypass. It is intended to be used during and after surgical procedures requiring cardiopulmonary bypass (up to 6 hours) when the removal of excess fluid from blood is required. It should not be used as a dialyzer, hemofilter or other device.
Device Story
CAPIOX Hemoconcentrator is a hollow-fiber device used during cardiopulmonary bypass to remove excess fluid from blood while retaining essential plasma proteins. Device consists of glycerin-free polysulfone fibers within a tubular casing. Blood flows inside fibers; hydrostatic pressure differential across membrane drives ultrafiltration of excess fluid. Available in two models (HC05, HC11) with varying surface areas. Operated by perfusionists or clinicians in surgical settings. Transparent ports allow air bubble observation during priming. Device requires no pre-rinse. Benefits include maintenance of hematocrit and protein levels and enabling reinfusion of residual circuit blood post-bypass.
Clinical Evidence
Bench testing only. Device performance validated through comparison of specifications and materials with predicate. Biocompatibility testing conducted per ISO-10993. Sterilization validated to SAL 10^-6. Ethylene oxide residuals meet regulatory limits.
Indicated for patients undergoing surgical procedures requiring cardiopulmonary bypass where removal of excess fluid is necessary to maintain hematocrit and protein concentration; intended for use during and up to 6 hours after bypass.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
SUMMARY OF SEPTEMBER SUPERSANTIAL FOUNVALENCE PERTAINING TO SUBSTANTIAL EQUIVALENCE
K973516
Proprietary Device Name: CAPIOX® Hemoconcentrator
Classification Name: High permeability dialyzer
Reason for Submission:
New device.
### Intended Use:
The CAPIOX Hemoconcentrator is designed to remove excess fluid from the blood in The CAPIOX Hemocorcentrator is accept and protein concentration during cardiopulmonary
order to maintain proper hematocrit and recorpsion in the citization of the bypass. order to maintain proper nematocrit and proceen could after bypass.
It is intended to be used during and after surgical procedures required. It should It is intended to be used during and arter survey of excess fluid from blood is required. It should
bypass (up to 6 hours) when the removal of excess fluide not be used as a dialyzer, hemofilter or other device.
### Description
CAPIOX® Hemoconcentrator consists of Glycerin-free polysulfone fibers, casing, blood port, O-ring and adhesives.
The hemoconcentrator is available in two model has a total membrane area I he nemoconcentrator is a valuable in a 1.08 m² surface area.
Three types of blood ports are available in each model: ¼" slip, 3/16" slip, and 3/16" Three types of blood ports are available in oden model.
The ports are transparent allowing easy observation of air bubbles passing through the ports when priming.
Dimensions of filtrate ports meet the requirements specified in ISO 8637 which permits connection of fast couplings.
The CAPIOX Hemoconcentuator provides high ultra filtration rates which permits the The CAPIOX Hemoval of excess fluid by a slight hydrostatic pressure differential across the membrane without loss of essential plasma proteins
{1}------------------------------------------------
II - Summary and Certification Summary and Cectiveness
The CAPIOX Hemoconcentrator provides high ultration rates which permits the The CAPIOX Hemoconcentrator provides mgr differentic pressure differential across the sufficient removal of excess that plasma proteins.
..
Pre-rinse is not needed. Priming with blood is acceptable.
## Substantial Equivalence
. . .
The CAPIOX® Hemoconcentrator is substantially equivalent to the Minntech Hemocor HPH
anting of the concerners and as follows The CAI 101&ator (510(k) K923139).as follows:
### Intended use: same
## Design and Materials
Both devices consist of a tubular casing which contains glycerine-free polysulfone bollow
wall and the contraction for for first while execuse fluid passes through the membr Both devices consist of a fibers while excess fibid passes through the membrane walls
fibers. Blood flows inside the fibers while excess through the membrane walls
to the sam fibers. Blood flows inside the fibers withe excess nun-passes and exist of the casing.
# Technology and Principles of Operation
Both devices use hollow fiber membrane technology. Some form of punping mechanism is Both devices use hollow fiber memoratie technology. Some to may be and thus through the hemoconcentrator.
{2}------------------------------------------------
II - Summary and Certification Summary of Safety and Effectiveness
## Specifications
1
The specifications for both devices are substantially equivalent. The table below summarizes the The specifications for bour devices are centrator with the Minntech Hemocor Hemoconcentrator.
| Table | - |
|-------|---|
|-------|---|
| | CAPIOX Hemoconcenterator | Minnetch Hemocor |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Used during and after surgical<br>procedures requiring<br>cardiopulmonary bypass when<br>the removal of excess fluid is<br>required. (Should not be used<br>as a dialyzer, hemofilter or for<br>any other function. | Used during and after surgical<br>procedures requiring<br>cardiopulmonary bypass when<br>the removal of excess fluid is<br>required. |
| Membrane technology | Hollow Fiber | Hollow Fiber |
| Membrane material | Glycerin-free polysulfone | Glycerin-free polysulfone |
| Blood flow relative to fiber | Inside | Inside |
| Effective surface area | HC11: 1.08m2<br>HC05: 0.5 m2 | HPH1000: 1.1m2<br>HPH400 : 0.3m2 |
| Blood Ports | 1/4" slip<br>3/16" slip<br>3/16"luer lock | 1/4" slip |
| Filtrate ports (*see note) | ½" Hansen quick connect | ¼" slip |
| Maximum Blood Flow | 500 mL/min | 500 mL/min |
| Maximum transmembrane<br>pressure | 500 mmHg | 500 mmHg |
| Priming volume | HC11:70 mL<br>HC05: 35 mL | HPH1000: 88 mL<br>HPH400 : 34 mL |
| Priming | No rinse required | No rinse required |
| Sterilization method | Ethylene oxide | Ethylene oxide |
*1/2" Hansen quick connect is the most commonly used filtrate port which is also used in the Baxter HQ-7000, Cobe Hemofilter, and Bard H-4207.
p. 3/4
{3}------------------------------------------------
Summary and Certification II -Summary of Safety and Effectiveness
In summary, the CAPIOX® and the Minntech Hemocor HPH Hemoconcentrator are substantially equivalent in intended use, design and materials, technology/principles of operation, specifications and performance. Differences as described above do not raise new issues of safety or effectiveness.
Terumo's statement that this device is substantially equivalent to any other device is Foreme of stocomply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.
### Additional Safety Information
- ·Pyrogen Testing
- · Sterilization conditions have been validated to provide a Sterility Assurance Level (SAL) of 10-6.
- · Ethylene oxide residuals will not exceed the maximum residue limits proposed for Part 821 of Title 21 in the Federal Register of June 23, 1978 (or as finalized or amended).
- · Manufacturing control testing
- · Blood contacting materials were tested in accordance with the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993, " Biological Evaluation of Medical Devices Part 1: Evaluation and Testing (External communicating devices/Circulating Blood/Limited contact duration).
Date Prepared September 3, 1997
- Prepared by: Sandi Hartka, Manager Regulatory Affairs
- for: Terumo Medical Corporation 2100 Cottontail Lane Somerset, NJ 08873
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the edge. Inside the circle is a stylized graphic of an eagle with three stripes representing its wings. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Sandi Hartka Manager Regulatory Affairs TERUMO.Medical Corporation 125 Blue Ball Road Elkton, MD 21921
Re. K973516
CAPIOX® Hemoconcentrator Models CX*HC05, CX*HC11 Dated: December 17, 1997 Received: December 19, 1997 Regulatory Class: III 21 CFR 876.5860/Procode: 78 KDI JAN 3 0 1998
Dear Ms. Hartka:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have are memor We have reviewed your Security from and of on use stated in the enclosure) to devices marketed in internet commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been recitasified in pror to may 26, 1970, the enabilied and on the medical of (Act). You may, therefore, market the ceries, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements of a column to the general controls provisions of the same of the sense and prohibitions against misbranding and additions of
If your device is classified (see above) into either class III (Prematet Approval) in may be subject to If your device is classified (see above) and the could in the Cound in the Coast of Essent Regulations. The 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Crook Post COST Postice 21, Falls 800 to 693. A substition (QS) for Medical Devices: General regulation (21 CFR Part 20) and requirement, as set to ut in the Quality System (V) its ) vill verify such assumptions. Filing on one oncerners assessming volu with the GMP regulation may results are action. In addition, FDA may publish further announcements concerning Viour device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections \$31 through \$42 of the Act for devices under the Electronic Provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitte if you done opening and the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807 97) Other general information on your resons (200) the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
h.7liau
Lillian Yiu, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
.. . ..
Enclosure
{5}------------------------------------------------
K973516 510(k) Number (if known):
CAPIOX® Hemoconcentrator Device Name:
## Indications For Use:
The CAPIOX Hemoconcentrator is designed to remove excess fluid from the I the CAYION Tiemoconoomiator is estages
blood in order to maintain proper hematocrit and protein concentration during cardiopulmonary bypass and to enable reinfusion of blood remaining in the circuit after bypass.
It is intended to be used during and after surgical procedures requiring it is michede to or or or 6 hours) when the removal of excess fluid from blood is required. It should not be used as a dialyzer, hemofilter or other device.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Over-The-Counter Use_ Prescription Use OR (Per 21 CFR 801.109) (Division Sign-Off) Division of Reproductive, Abduminal, ENT, (Optional Format 1-2-96) and Radiological Devices 510(k) Number
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.