K973509 · Kendall Healthcare Products Co. Div.Of Tyco Health · KMF · Dec 4, 1997 · General Hospital
Device Facts
Record ID
K973509
Device Name
KENDALL CURASORB ZN + MG ALGINATE WOUND DRESSING
Applicant
Kendall Healthcare Products Co. Div.Of Tyco Health
Product Code
KMF · General Hospital
Decision Date
Dec 4, 1997
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 880.5090
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Like predicate devices, the Kendall CURASORB® Zn+Mg Alginate Wound Dressings are external wound dressings designed to help protect the wound, provide a moist environment for wound healing, and absorb wound exudate. CURASORB® Zn+Mg Dressings are indicated for use on wounds which present moderate to heavy wound exudate such as dermal ulcers, traumatic wounds, donor sites and seconddegree burns.
Device Story
CURASORB® Zn+Mg Alginate Wound Dressing is a sterile, single-use, nonwoven, absorptive dressing composed of calcium, sodium, zinc, and magnesium alginate fibers. Used in clinical or home settings for wound management. Upon contact with wound exudate, fibers transform into a firm gel/fiber mat, providing a moist healing environment and cushioning the wound. Available in various pad sizes and packing ropes. Healthcare providers apply the dressing to the wound surface to manage exudate; the device does not accelerate healing or act as synthetic skin.
Clinical Evidence
No clinical data. Biocompatibility testing performed per ISO 10993 Part I and FDA memorandum G95-1.
Technological Characteristics
Nonwoven, absorptive dressing composed of calcium, sodium, zinc, and magnesium alginate fibers. Sterile, single-use. Available in various pad dimensions and packing rope lengths. Biocompatibility tested per ISO 10993 Part I.
Indications for Use
Indicated for management of moderate to heavy exudating wounds, including dermal ulcers (venous stasis, pressure, arterial, diabetic), traumatic wounds (lacerations, punctures, abrasions, incisions), donor sites, and second-degree burns. Contraindicated for third-degree burns.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
Predicate Devices
Kendall CURASORB® Calcium Alginate Dressing
Kendall CURASORB® Zn Alginate Wound Dressing
Submission Summary (Full Text)
{0}------------------------------------------------
K973509
### Exhibit 9
DEC - 4 1997
## 510(k) Summary
In accordance with section 513(i)(3) of the SMDA and as described in 21 CFR Part 807.3, final rule dated December 14, 1994, this summary is submitted by:
Kendall Healthcare Products Company 15 Hampshire Street Mansfield, MA 02048 Date: September 4, 1997
#### Contact Person 1.
Jan L. Efflandt, Regulatory Affairs Telephone: (508) 261-8037
- 2. Name of Device
| Classification Name: | Unclassified |
|-----------------------|-------------------------------------------------|
| Common or Usual Name: | Calcium Zinc and Magnesium Alginate<br>Dressing |
| Proprietary Name: | CURASORB® Zn+Mg Alginate Wound<br>Dressing |
#### 3. Statement of Substantial Equivalence
The Kendall CURASORB® Zn+Mg Alginate Wound Dressing is substantially equivalent in composition, form, function and intended use to the commercially available Kendall CURASORB® Calcium Alginate Dressing and Kendall CURASORB® Zn Alginate Wound Dressing.
#### Device Description 4.
The Kendall CURASORB® Zn+Mg Alginate Wound Dressings are sterile, single-use, nonwoven, absorptive dressings. The off-white alginate fibers are composed of calcium, sodium, zinc and magnesium. As the dressing absorbs wound exudate it converts to a firm gel/fiber mat. This gel cushions the wound and forms a moist environment at the wound interface. The product is available in 2" x 2", 4" x 4", and 4" x 8" , 6" x 10", 12" x 24" pads, as well as 12", 24" and 36" long wound packing rope.
#### ട്. Device Intended Use
Like predicate devices, the Kendall CURASORB® Zn+Mg Alginate Wound Dressings are external wound dressings designed to help protect the wound, provide a moist environment for wound healing, and absorb wound exudate. CURASORB® Zn+Mg Dressings are indicated for use on wounds which present moderate to heavy wound exudate
{1}------------------------------------------------
such as dermal ulcers, traumatic wounds, donor sites and seconddegree burns.
#### Product Comparison .
The Kendall CURASORB® Zn+Mg Alginate Wound Dressings are equivalent to the referenced predicate devices in that they are fabricated from similar materials, have a similar function and equivalent indications for use.
#### 7. Nonclinical Testing
Biocompatibility testing of the Kendall CURASORB® Zn+Mg Alginate Wound Dressing has demonstrated that it contains no bioactive components. Testing performed on the product was based upon guidelines presented in the 10993 ISO Standard, Part I, with the FDA modified matrix presented in memorandum G95-1.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & IAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with its wings spread, depicted in a minimalist, flowing design. The seal appears to be a formal emblem, possibly representing a government agency or organization related to health and services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 4 1997
Ms. Jan L. Efflandt 'Regulatory Affairs Administrator Kendall Healthcare Products Company 15 Hampshire Street Mansfield, Massachusetts 02048
Re: K973509 Kendall CURASORB® Zn+Mg Alginate Wound Dressing Regulatory Class: Unclassified Product Code: KMF Dated: September 15, 1997 Received: September 16, 1997
Dear Ms. Efflandt:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market your device subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations:
- 1. This device may not be labeled for use on third degree burns.
- 2. This device may not be labeled as having any accelerating effect on the rate of wound healing or epithelization.
- 3. This device may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin.
- 4. This device may not be labeled as a treatment or a cure for any type of wound.
The labeling claims listed above would be considered a major modification in the intended use of the device and would require a premarket notification submission (21 CFR 807.81). The general controls provisions of the Act include requirements for annual
{3}------------------------------------------------
Page 2 - Ms. Jan L. Efflandt
registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address http://www.fda.gov/cdrh/dsmamain.html.
Sincerely yours,
[signature]
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
973509
Device Name: CURASORB Zn+Mg Alginate Wound Dressing
Indications For Use:
The proposed device is for use in the management of wounds which present moderate to heavy wound exudate such as dermal ulcers (i.e., venous stasis, pressure, arterial, diabetic); traumatic wounds (i.e., lacerations, punctures, abrasions or incisions); donor sites; and second-degree burns.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---------------------------------|----------|
| Division of General Restorative | |
| 510(k) Number | IC973509 |
| Prescription Use<br>(Per 21 CFR 801.109) | <div style="text-align: center;">X</div> | OR | Over-The-Counter Use |
|------------------------------------------|------------------------------------------|----|----------------------|
|------------------------------------------|------------------------------------------|----|----------------------|
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.