NORLAND BODY COMPOSITION OPTION FOR NORLAND DEXA BONE DENSITOMETERS
Applicant
Norland Corp.
Product Code
KGI · Radiology
Decision Date
Feb 13, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1170
Device Class
Class 2
Indications for Use
Body composition assesses the non-bone tissue determined during the bone density scans performed by Norland DEXA bone densitometers, and estimates soft tissue mass, fat mass, lean mass, total soft tissue mass, % fat, TBMC/LBM (Total Bone Mineral Content / Lean Body Mass), and Siri and Brozek equation equivalent values. These body composition estimates are useful to health care professionals in their management of diseases/conditions where the disease/condition itself, or its treatment, can affect the relative amounts of patient fat and lean tissue. The Norland Body Composition Option does not diagnose disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments. Some of the diseases/conditions for which lean and fat tissue estimates are useful are chronic renal failure, Anorexia Nervosa, excessive obesity, AIDS/HIV, and Cystic Fibrosis. Body composition is also a convenient alternative to hydrostatic weighing and skin fold measurements.
Device Story
Software option for Norland DEXA bone densitometers; utilizes existing bone density scan data to estimate body composition; no additional radiation or scanning required. Inputs: DEXA scan data. Processing: analysis of non-bone tissue to calculate soft tissue mass, fat mass, lean mass, % fat, and TBMC/LBM; applies Siri and Brozek equations. Outputs: quantitative body composition values. Used in clinical settings by healthcare professionals to monitor conditions like renal failure, obesity, or HIV. Does not diagnose or recommend treatment; provides data for clinical management. Benefits: non-invasive alternative to hydrostatic weighing or skin fold measurements.
Clinical Evidence
Clinical study involving hundreds of subjects compared Norland body composition values to underwater weighing (UWW) using Siri and Brozek equations. Results demonstrated that the device provides Siri and Brozek equivalent values for whole body scans.
Technological Characteristics
Software-based analysis module for DEXA bone densitometers. Operates on existing scan data; no hardware changes. Uses mathematical modeling (Siri and Brozek equations) to derive body composition from non-bone tissue data. Standalone software option for existing Norland DEXA systems.
Indications for Use
Indicated for patients requiring assessment of non-bone tissue (soft tissue mass, fat mass, lean mass, % fat) as part of management for conditions affecting body composition, including chronic renal failure, Anorexia Nervosa, obesity, AIDS/HIV, and Cystic Fibrosis. Used by healthcare professionals; no specific age or gender restrictions.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Lunar Total Body Software Option for the DPX bone densitometer (K935454)
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 1 3 1998
## K973459
## 510k Summary
## as required by 807.92(c) for Norland Body Composition Option
Prepared January 1998
Norland Corporation Submitted by: W6340 Hackbarth Road Fort Atkinson. WI 53538 Reg. # 2124648 920-563-8456 Mr. Terry Schwalenberg Contact Person: Director Regulatory Affairs Norland Body Composition Option Device Trade Name: Body Composition Option for DEXA bone densitometers Common Name: Classification: Bone densitometer, (21 CFR 892.1170), product code 90KGI; Class II Predicate Devices: Lunar Total Body Software Option for the DPX bone densitometer (K935454) The Norland Body Composition Option (body composition) is used with Description of Norland DEXA bone densitometers to estimate the relative amounts of lean Device: and fat tissue in the scan area. Body composition does not require any changes to the bone densitometer nor does it require additional scanning or radiation exposure beyond the bone density scans. In most cases, existing scans can be re-analyzed to provide the body composition values. Although body composition is most useful for whole body scans, it is compatible with all scan modes for Norland DEXA bone densitometers. Body composition values for whole body scans were compared to underwater weighing (UWW) values in clinical studies involving hundreds of subjects. The UWW values were determined using both the Siri and Brozek equations. Based on this study, the Norland body composition provides the Siri and Brozek equivalent values for the Whole Body scans. Intended use Body composition assesses the non-bone tissue determined during the bone density scans performed by Norland DEXA bone densitometers, and estimates soft tissue mass, fat mass, lean mass, total soft tissue mass, % fat, TBMC/LBM (Total Bone Mineral Content / Lean Body Mass), and Siri and Brozek equation equivalent values.
{1}------------------------------------------------
These body composition estimates are useful to health care professionals in their management of diseases/conditions where the disease/condition itself, or its treatment, can affect the relative amounts of patient fat and lean tissue. The Norland Body Composition Option does not diagnose disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments. Some of the diseases/conditions for which lean and fat tissue estimates are useful are chronic renal failure, Anorexia Nervosa, excessive obesity, AIDS/HIV, and Cystic Fibrosis. Body composition is also a convenient alternative to hydrostatic weighing and skin fold measurements.
Safety
Body composition doesn't require additional radiation exposure over what is required for the bone density scan alone. The data which is normally analyzed for the bone values can also be analyzed for the lean and fat values. During the scan, the subject lies quietly on a cushioned table (or sits in a chair); they are not required to be in contact with any electrical part or to ingest any chemical/drug.
The delivered dose for Norland bone densitometers range from 0.07 mrem for a whole body scan, to 3 mrem for a pDEXA forearm scan, to 10 mrem for a hip scan. Stray radiation is very small. At a distance of three feet from the x-ray beam, the radiation field is less than 0.1 mRem per hour. This field is present only during the few minutes of the actual patient scan; and is primarily scatter from the patient. Shielded rooms are not required, and the operators are not required to be out of the room during the scan. These dose levels compare favorably to the background radiation level of 100 mrem per year; to a chest x-ray of 35 mrem; and to the increased dose received during a cross country (round trip) airline flight of 5 mrem.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 3 1998
Terry Schwalenberg Director, Regulatory Affairs Norland Corporation W6340 Hackbarth Rd. Fort Atkinson, WI 53538-8999 Re: K973459
Norland Body Composition Option for Norland Dexa Bone Densitometers Dated: January 5, 1998 Received: January 6, 1998 Regulatory class: II 21 CFR 892.1170/Procode: 90 KGI
Dear Mr. Schwalenberg:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitte diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrbdsmamain.html": ---------------------------------------------------------------------------------------------
Sincerely yours,
h. Tiau Yu.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) number (if known):
Device Name: Body Composition Option for Norland DEXA Bone Densitometers
Indications for use:
The Norland Body Composition Option (body composition) assesses the non-bone tissue determined during the bone density scans performed by Norland DEXA bone densitometers, and estimates soft tissue mass, fat mass, total soft tissue mass, % fat, TBMC/LBM (Total Bone Mineral Content / Lean Body Mass), and Siri and Brozek equation equivalent values.
These body composition estimates are useful to health care professionals in their management of diseases/conditions where the disease/condition itself, or its treatment, can affect the relative amounts of patient fat and lean tissue. The Norland Body Composition Option does not diagnose disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments. Some of the diseases/conditions for which lean and fat tissue estimates are useful are chronic renal failure, Anorexia Nervosa, excessive obesity, AIDS/HIV, and Cystic Fibrosis. Body composition is also a convenient alternative to hydrostatic weighing and skin fold measurements.
- (Please do not write below this line - continue on another page if needed) -----------Concurrence of CDRH; Office of Bevice Evaluation (ODE)--------------
Prescription Use (Per 21 CFR 801.109)
(Optional Format 1-2-96)
OR
Over-the-counter Use
David le. Sejimm
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT and Radiological Devices
510(k) Number K973459
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.