K973450 · Inventive Products, Inc. · IYO · Nov 14, 1997 · Radiology
Device Facts
Record ID
K973450
Device Name
SENSOR PAD
Applicant
Inventive Products, Inc.
Product Code
IYO · Radiology
Decision Date
Nov 14, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The Sensor Pad® is indicated as an aid for performing breast self-examination.
Device Story
Sensor Pad is a thin, 9.5-inch diameter polyether polyurethane bladder containing lubricating fluid. Designed for use during breast self-examination (BSE), the device is placed between the user's fingers and breast tissue. The low-friction silicone fluid and flexible bladder material reduce friction during palpation, facilitating tactile examination. The device is intended for OTC use by individuals performing self-exams. It functions as a mechanical aid to improve the sensitivity of manual palpation by allowing smoother movement over tissue, potentially assisting in the detection of breast lumps.
Clinical Evidence
Six clinical studies were conducted investigating tactile detection of lumps using the Sensor Pad compared to lubricant-only or unaided detection. Studies evaluated clinical performance during breast self-examination. Physical bench testing measured the coefficient of friction to validate the device's friction-reducing properties. Biocompatibility data for the polyether polyurethane and silicone fluid were provided.
Technological Characteristics
Materials: polyether polyurethane bladder, silicone lubricating fluid. Form factor: 9.5-inch diameter thin flexible bladder. Principle: mechanical friction reduction via fluid-filled interface. No energy source, software, or connectivity.
Indications for Use
Indicated for individuals performing breast self-examination as an aid to tactile detection of lumps.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Premarket Notification Inventive Products Inc. - Sensor Pad® September 9, 1997
973450 8102
NOV 1 4 1997
# 510(k) SUMMARY Sensor Pad® September 9, 1997
This summary is provided in accordance with the Safe Medical Devices Act of 1990 (SMDA). The information provided in the 510(k) premarket notification was in accordance with 21CFR 807.87 and the SMDA.
## 1. Submitter of 510(k)
| Company Name: | Inventive Products, Inc. |
|-----------------|---------------------------------------------------|
| Address: | 1450 East North Street<br>Decatur, Illinois 62521 |
| Contact Person: | Grant Wright, President |
| Telephone: | (217) 423-6911 |
| Fax: | (217)423-7282 |
- 2. Name of Device:
Trade/Proprietary Name:
Sensor Pad®
Common/Usual Name:
Aid for breast self-examination
## Classification Name
21 CFR 892.1710 "Mammographic x-ray system (accessory)" Class I
## ತೆ. Legally Market Predicate Devices
The Sensor Pad® cleared under 510(k) K955249
{1}------------------------------------------------
11973450 ド 201 レ
Premarket Notification Inventive Products, Inc. - Sensor Pad® September 9, 1997
### 4. Device Description
The Sensor Pad® is a thin polyether polyurethane bladder, 9.5 inches in diameter containing a small quantity of a lubricating fluid. The very thin nature of the polyether polyurethane allows the pad to conform to the shape of the tissues underlying it. The very low coefficient of friction property of the silicone fluid in combination with the very thin and flexible bladder, results in an easy sliding relationship between the upper and lower surfaces of the bladder. Therefore when properly used between the finger and the soft tissues of the patient, a reduction of friction is observed.
Performance testing of the Sensor Pad's® use on tactile examination was extensively investigated. These studies were provided and described in detail. The reported data included descriptions of the protocols, findings, and conclusions. Physical tests demonstrated the Sensor Pad's® ability to reduce friction, as measured as the coefficient of friction. Biocompatibility data pertinent to the materials used in the manufacture of the Sensor Pad® were also provided. Six clinical studies were reported. These studies included investigations of the tactile detection of lumps using the Sensor Pad® as compared to using a lubricant and/or unaided detection, as well as the clinical performance of the Sensor Pad® during breast self-examination. Additional testing was also conducted to answer additional questions relating to OTC clearance. Results were provided and demonstrate that a change from "Prescription Use Only" to "OTC Use" does not in any way alter the intended use or the safety or effectiveness of this device.
#### 5. Intended Use
The Sensor Pad® is indicated as an aid for performing breast selfexamination.
#### Substantial Equivalence હ.
The Sensor Pad® is physically identical and has the same indication for use as that previously cleared under K955249. This application is to change from "Prescription Use Only" to OTC. The Sensor Pad® previously cleared thus serves as the legally marketed predicate device. Valid scientific evidence, as discussed above, was provided to demonstrate the substantial equivalence determination.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & H. MAN...S. USA" is arranged in a circular pattern around the bird symbol. The text is in all capital letters and is slightly blurred.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 4 1997
Mr. Grant A. Wright President Inventive Products, Inc. 1450 East North Street Decatur, Illinois 62521 Re: K973450 Sensor Pad® - OTC Dated: September 9, 1997 Received: September 11, 1997 Regulatory class: II 21 CFR §892.1710/Product code: 90 IXH
Dear Mr. Wright:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device are thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitto diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note regulation entitled, the the state of the color of the more to the more of the reneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-fromuner (800) 638-204) or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W.Liau Yu
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, . and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page 1 of 1
510(k) Number (if known):_K973450
Device Name: Sensor Pad
Indications For Use:
"The Sensor Pad is indicated as an aid for performing breast self-examination".
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive Abdominal
Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K973450
Prescription Use (Per 21 CFR 801 109)
.
ੇ ਦ
Over-The-Counter Use_レ
.Counter Use *✓*
(Optional Format 1-2-96)
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 3 0 1998
Mr. Grant Wright President Inventive Products, Inc. 1450 East North Street Decatur, Illinois, 62521
Re: K973450
> Sensor Pad® - OTC Dated: September 9, 1997 Received: September 11, 1997 Regulatory Class: II 21 CFR 892.1560 Product Code: 90 IYO
Dear Mr. Wright:
This letter corrects our substantially equivalent letter of November 11, 1997, regarding the CFR classification number from 892.1710 to 892.1560 and product code for 90 IXH to 90 IYO. We apologize for the error and any inconvenience it may have caused.
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General (QS) regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (I'DA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to continue marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{5}------------------------------------------------
Page -2 - Mr. Grant A. Wright
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) and add Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4699. Other general information on your responsibilities under the Office of obtained from the Division of Small Manufacturers Assistance at their toll frec number (800) 688-204) or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmanain.html".
Sincerely yours, 、『 ﺎﺭ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟ
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page_1_of_1_ ::::「
510(k) Number (if known):_K973450
Device Name:___Sensor Pad
Indications For Use:
"The Sensor Pad is indicated as an aid for performing breast
self-examination" self-examination".
",
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF
EEDED) NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| (Division Sign-Off) |
|--------------------------------------------------------------------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices |
| 510(k) Number | K973450 |
|---------------|---------|
|---------------|---------|
| scription Use |
|---------------|
|---------------|
Pres (Per 21 CFR 801. 109)
.. F
િત્ત
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
্র
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.