STERI-OSS 3.25 MM REPLACE(TITANIUM PLASMA SPRAYED) DENTAL IMPLANT

K973433 · Steri-Oss, Inc. · DZE · Dec 5, 1997 · Dental

Device Facts

Record IDK973433
Device NameSTERI-OSS 3.25 MM REPLACE(TITANIUM PLASMA SPRAYED) DENTAL IMPLANT
ApplicantSteri-Oss, Inc.
Product CodeDZE · Dental
Decision DateDec 5, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use for this device is to serve as support for prosthetic devices to restore patient chewing function.

Device Story

Endosseous dental implant designed to serve as artificial tooth root; supports abutment/prosthesis acting as crown. Implant dimensions: 3.25 mm diameter, 10-16 mm length. Fabricated from titanium alloy; features threaded body with 50-micron plasma-sprayed titanium powder coating and anodized magenta external hex. Used by dental clinicians to restore masticatory function. Device mimics natural dentition structure; healing screws and abutments used as intermediaries during restoration process.

Clinical Evidence

Bench testing only.

Technological Characteristics

Titanium alloy construction; 3.25 mm diameter; 10-16 mm length. Threaded body with 50-micron plasma-sprayed titanium powder coating. External hex interface, anodized magenta. Sterile.

Indications for Use

Indicated for patients requiring prosthetic support to restore chewing function via endosseous dental implants.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 73433 #### Original 510(k) Ongmal 016(K) Steri-Oss 3.25 mm Replace (Titanium Plasma Sprayed) Dental Implant #### Section 6 #### 510(k) Summary #### Manufacturer Information: | Submitter's Name: | Steri-Oss Inc. | | |----------------------|------------------------------------------------------|--| | Submitter's Address: | 22895 Eastpark Drive<br>Yorba Linda, CA 92887<br>USA | | | Contact's Name: | Jeff Hausheer, Ph.D. | | | Contact's Telephone: | 714-282-4800, extension 3815 | | | Date Prepared: | September 1997 | | #### Device Name: | Common Name: | Prosthetic Dental Implant | |----------------------|-----------------------------------------------------------------------| | Trade Name: | Steri-Oss 3.25 mm Replace (Titanium Plasma Sprayed)<br>Dental Implant | | Classification Name: | Endosseous dental implant | ### Predicate Device: Substantial equivalence is claimed to Steri-Oss' Replace Implants and to Steri-Oss' (3.25 mm diameter) Titanium Plasma Sprayed Cylindrical Implant. Page 1 of 3 {1}------------------------------------------------ Original 510(k) Steri-Oss 3.25 mm Replace (Titanium Plasma Sprayed) Dental Implant #### Section & ## 510(k) Summary ### Device Description: #### How The Device Functions: Steri-Oss 3.25 mm Replace Implant: This device is designed to serve as support for prosthetic devices to restore patient chewing function. ### Scientific Concepts: Natural dentition is composed of a root (subgingival) and a crown (supragingival). Endosseous dental implant designs in existence are intended to mimic this structure to aid the patient in restoring natural masticatory function. The subject implant is designed to serve as the root of the artificial tooth and the abutment/prosthesis is designed to serve as the crown. The healing screw and healing abutment are intermediary devices which serve to permit satisfactory completion of the implant healing/restoration process. #### Device Characteristics: This implant is 3.25 mm in diameter, 10 to 16 mm in length, and is fabricated from titanium allov. They are threaded and have a superior external hex. The threaded portion of the implant body has a 50 micron thick plasma sprayed titanium powder coating, and the surface of the external hex is anodized magenta in color. Page 2 of 3 {2}------------------------------------------------ Original 510(k) Steri-Oss 3.25 mm Replace (Titanium Plasma Sprayed) Dental Implant Section 6 ### 510(k) Summary #### Intended Use: The intended use for this device is to serve as support for prosthetic devices to restore patient chewing functions. #### Comparison to Predicate: The following table provides a comparison of the principle technological characteristics of the 3.25mm Replace (Titanium Plasma Sprayed) Implant and the predicate. | -----------<br><br>(<br>------<br><br>-<br>1<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>----<br>-------<br><br>140 Children<br>1 | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|-------------| | ------<br> | -<br>-<br>.<br>------<br><br>1 | 1400 1<br>1 | | Specification/<br>Characteristic | Predicate:<br>Steri-Oss Replace<br>Implant | New Product:<br>Steri-Oss 3.25 mm<br>Replace (Titanium<br>Plasma Sprayed)<br>Dental Implant | |----------------------------------|----------------------------------------------------------------|---------------------------------------------------------------------------------------------| | Material | Titanium | Same | | Surface<br>Characteristics | Color-coded hex; TPS<br>coating on threaded<br>portion of body | Same | | Sterility | Sterile | Same | A second predicate, Steri-Oss' 3.25 mm titanium plasma sprayed cylindrical implant, is presented for the 3.25 mm diameter. # Performance Data: Not applicable. Page 3 of 3 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 5 1997 Jeff Hausheer, Ph.D. ·Regulatory Affairs Specialist Steri-Oss, Incorporated 22895 East Park Drive Yorba Linda, California 92687 K973433 Re: Steri-Oss 3.25mm Replace (Titanium Plasma Trade Name: Spray Dental Implant) Requlatory Class: ı il Product Code: DZE September 8, 1997 Dated: September 10, 1997 Received: Dear Dr. Hausheer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 {4}------------------------------------------------ Page 2 - Dr. Hausheer through 542 of the Act for devices under the Electronic through 542 of the Act 101 devices ander and in the Federal laws or requlations. This letter will allow you to begin marketing your device as fills feecer will areas) (k) premarket notification. The FDA described in your stority promote of your device to a legally finding of substancial equivalence of policition for your marketed predicate device rebarts in a camber. device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in regulacion (21 cm rares), please contact the Office of ville diagnoscic devices) / proams of the questions on Compilance at (3017 334 4010. India device, please contact che promocion and adversibling of (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to the regulation entitled, "Misbranding by reference to premation on your responsibilities under the Act may be Information on your responsible of Small Manufacturers Assistance obtained Trom the Division of 638-2041 or (301) 443-6597 or at at its coll-free namber (05) -300f-2007 -050f-110smamain.html". Sincerely yours, Timothy A. Ulatowski Timot A Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Original 510(k) Original 510(K) Steri-Oss 3.25 mm Replace (Titanium Plasma Sprayed) Dental Implant Section 9 Indications for Use Page 1 of 1 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ 3.25 mm Replace (Titanium Plasma Sprayed) Dental Implant Device Name: Indications For Use: The intended use for this device is to serve as support for prosthetic devices to restore patient chewing function. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | | Concurrence of CDRH, Office of Device Evaluation (ODE) | |---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------| | (Division Sign-Off) | | | Division of Dental, Infection Control, and General Hospital Devices | | | 510(k) Number | K233433 | | Prescription Use | <div style="display:inline-block;">✓</div> OR Over-The-Counter Use _______ (Per 21 CFR 801.109) | (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...