The TDx PRM is intended to be used as an accessory to emergency resuscitators to monitor vontilatory functions including peak inspiratory pressure, respiratory rate, positive end expiratory pressure, and inspiratory/expiratory time ratio.
Device Story
TDx PRM is a portable, battery-operated airway pressure monitor; attaches to emergency resuscitators (bag mask, demand valves, transport ventilators) lacking pressure gauges. Device uses electronic pressure transducer to monitor peak inspiratory pressure, breath frequency, I/E time, and PEEP; includes CPR metronome for pacing resuscitation. No direct patient contact; uses standard bacterial filters to prevent pathogen transmission. Used in emergency settings by clinicians; provides real-time monitoring of ventilatory parameters to assist in resuscitation management. Device performs no diagnostic or therapeutic functions; provides no alarms.
Clinical Evidence
Bench testing only. Testing included functional performance, environmental stress (impact/drop, storage temperature), electromagnetic interference, electrostatic discharge, spill resistance, surface temperature, and power supply/battery life testing. No clinical data provided.
Technological Characteristics
Microprocessor-controlled monitor; electronic pressure transducer sensing; battery-powered; portable form factor. Attaches to airway circuits via standard bacterial filters. No direct patient contact. Software-based parameter calculation.
Indications for Use
Indicated for use as an accessory to emergency resuscitators (bag mask, oxygen-powered demand valves, or automatic transport ventilators) to monitor ventilatory parameters including peak inspiratory pressure, respiratory rate, PEEP, and I/E ratio in patients requiring emergency ventilation.
Regulatory Classification
Identification
An airway pressure monitor is a device used to measure the pressure in a patient's upper airway. The device may include a pressure gauge and an alarm.
Sechrist Industries, Inc., Airway Pressure Monitor Model 400 (K814608)
Submission Summary (Full Text)
{0}------------------------------------------------
JUL 23 1998
K973347
### Spiracle Technology TDx PRM Pulmonary Resuscitation Monitor 510(k) SUMMARY
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
| Submitter | Contact Person |
|------------------------------------|---------------------------------|
| Michael Sarné | C. Stephen Lawrence |
| Spiracle Technology | Hogan & Hartson, L.L.P. |
| 16520 Harbor Boulevard, "D" | 4675 MacArthur Court, Suite 670 |
| Fountain Valley, California 92708 | Newport Beach, California 92660 |
| Telephone: (714) 418-1091 | Telephone: (714) 440-0400 |
| Facsimile: (714) 418-1095 | Facsimile: (714) 833-0976 |
#### Name of Device
| Trade Name: | TDx PRM Pulmonary Resuscitation Monitor |
|----------------------|-----------------------------------------|
| Common Names: | Respiratory Monitor |
| Classification Name: | Airway Pressure Monitor |
| | 21 C.F.R. 868.2600 |
#### Predicate Devices
- (1) Bird Mean Airway Pressure Monitor (K812140)
- (2) Sechrist Industries, Inc., Airway Pressure Monitor Model 400 (K814608)
#### Intended Use
The TDx PRM is intended to be used as an accessory to emergency resuscitators to monitor vontilatory functions including peak inspiratory pressure, respiratory rate, positive end expiratory pressure, and inspiratory/expiratory time ratio.
## Technological Characteristics and Substantial Equivalence
The TDx PRM is intended to be used as an accessory to emergency resuscitators to monitor ventilatory functions. It is a small, portable battery operated device that attaches to bag mask resuscitators, oxygen powered demand valve resuscitators or automatic transport ventilators, which may not be equipped with pressure gauges for the purpose of patient airway monitoring. The TDx PRM monitors the following patient parameters: peak inspiratory pressure, breath frequency, inspiratory and expiratory times, positive end expiratory pressure ("PEEP"). The TDx PRM also offers a cardiopulmonary resuscitation metronome to
{1}------------------------------------------------
assist in the pacing of cardiopulmonary resuscitation efforts. The TDx PRM performs no diagnostic or therapeutic functions; it functions solely as a patient monitor and does not provide any patient alarms. Moreover, the TDx PRM has no direct patient contact; transmission of airborne pathogens is prevented by means of standard bacterial filters such as are routinely used with emergency resuscitation equipment.
The TDx PRM is substantially equivalent to other airway pressure monitors including the Bird Mean Airway Pressure Monitor (K812140) and Sechrist Industries, Inc., Airway Pressure Monitor Model 400 (K814608). These predicate devices, like the TDx PRM are microprocessor-controlled devices that utilize a electronic pressure transducer. Additionally, the predicate devices, like the TDx PRM. calculate and display breath rate, peak inspiratory pressure, inspiratory/expiratory time ratio, and positive end expiratory pressure.
#### Performance Data
Extensive functional testing of the TDx PRM has been performed. In addition, testing of the device has been performed under various environmental conditions, including impact/drop testing, storage temperature testing, electromagnetic interference testing, electrostatic discharge testing, spill resistance testing and surface temperature testing. Power supply testing also was performed: these tests included battery life testing and low power icon testing. The functional, environmental and power supply testing performed on the device demonstrated that it meets its performance objectives and complies with applicable FDA guidelines.
{2}------------------------------------------------
Image /page/2/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 23 1998
Mr. Guy Gansel Spiracle Technology 16520 Harbor Boulevard, Suite "D" Fountain Valley, CA 92708-1360
Re: K973347 TD2 PRM Regulatory Class: II (two) Product Code: 73 CAP May 8, 1998 Dated: Received: May 11, 1998
Dear Mr. Gansel:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the [kit/tray] have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with
{3}------------------------------------------------
Page 2 - Mr. Guy Gansel
the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callista
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Spiracle Technology TDx PRM Pulmonary Resuscitation Monitor Indications for Use Statement
# 510(k) Reference Number:
This is an initial submission; no number has yet been assigned.
# Statement of Indications for Use:
The TDx PRM is intended to be used as an accessory to emergency Fire TDX I Har is and ventilators to monitor ventilatory functions including peak resulchators and voltators or rate, positive end expiratory pressure, and inspiratory/expiratory time ratio.
Marke Kramer
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
. Prescription use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.