← Product Code [MDB](/productcode/MDB) · K973325

# BACT/ALERT FAN CULTURE BOTTLES (K973325)

_Organon Teknika Corp. · MDB · Dec 2, 1997 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K973325

## Device Facts

- **Applicant:** Organon Teknika Corp.
- **Product Code:** [MDB](/productcode/MDB.md)
- **Decision Date:** Dec 2, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.2560
- **Device Class:** Class 1
- **Review Panel:** Microbiology

## Indications for Use

BacT/Alert® FAN™ Culture Bottles are used with the BacT/Alert® Microbial Detection System in qualitative procedures for enhanced recovery and detection of aerobic and anaerobic microorganisms from blood and other body fluids.

## Device Story

BacT/Alert FAN Culture Bottles contain growth media, an adsorbent, and an internal CO2 sensor. Used in conjunction with the BacT/Alert Microbial Detection System, the bottles facilitate the recovery and detection of aerobic and anaerobic microorganisms from blood and other body fluids. The system monitors microbial growth by detecting CO2 levels within the bottle. The device is intended for use in clinical laboratory settings by trained personnel. The output provides qualitative detection of microbial presence, assisting healthcare providers in diagnosing infections and guiding clinical decision-making regarding antimicrobial therapy.

## Clinical Evidence

Performance characteristics were established through comparison to traditional culture methodology. Testing confirmed the device's ability to recover and detect aerobic and anaerobic microorganisms from blood and other body fluids, demonstrating equivalence to conventional culture techniques.

## Technological Characteristics

Contains growth media, an adsorbent, and an internal CO2 sensor. Operates as an in-vitro diagnostic system for microbial growth monitoring. Designed for use with the BacT/Alert Microbial Detection System.

## Regulatory Identification

A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.

## Predicate Devices

- Traditional Culture Methodology

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K973325

510(k) Premarket Notification Organon Teknika Corporation BacT/Alert™ FAN Culture Bottles

# 510(k) Summary BacT/Alert® FAN™ Culture Bottles

DEC - 2 1997

## The submitter's name, address, telephone number, a contact person, and the date the (a)(1) summary was prepared;

Submitter's Name: Organon Teknika Corporation

Submitter's Address: 100 Akzo Avenue, Durham, North Carolina, 27712 USA

Submitter's Telephone: (919) 620-2288

Submitter's Contact: Rebecca A.Rivas

Date 510(k) Summary Prepared:

## The name of the device, including the trade or proprietary name if applicable, the common or (a)(2) usual name, and the classification name, if known;

Trade or Proprietary Name: BacT/Alert® FAN™ Culture Bottle

Common or Usual Name: Culture Bottles

Classification Name: Microbial Growth Monitor

### An identification of the legally marketed device to which the submitter claims substantial (a)(3) equivalence;

Device Equivalent to: Traditional Culture Methodology

#### A description of the device. (a)(4)

1

Device Description: The BacT/Alert® FAN Culture Bottles contain media, an adsorbent and an internal sensor that detects CO2 as an indicator of microbial growth when used in connection with the BacT/Alert® Microbial Detection System.

#### A statement of the intended use of the device. (a)(2)

BacT/Alert® FAN™ Culture Bottles are used with the BacT/Alert® Device Intended Use: Microbial Detection System in qualitative procedures for enhanced recovery and detection of aerobic and anaerobic microorganisms from blood and other body fluids.

{1}------------------------------------------------

## 510(k) Premarket Notification Organon Teknika Corporation BacT/Alert™ FAN Culture Bottles

#### A summary of the technological characteristics of the new device in comparison to those of (a)(6) the predicate device.

The BacT/Alert® FAN Culture Bottles contain media, an adsorbent and an internal sensor that detects CO2 as an indicator of microbial growth when used in connection with the BacT/Alert® Microbial Detection System.

#### A brief discussion of the clinical tests submitted, referenced, or relied on in the premarket (b)(2) notification submission for a determination of substantial equivalency.

Testing was performed to establish the performance characteristics of the new device including comparison to traditional culture.

## The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device (b)3) is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

Organon Teknika's BacT/Alert® FAN Culture Bottles are used with the BacT/Alert® Microbial Detection System in qualitative procedures for enhanced recovery and detection of aerobic and anaerobic microorganisms from blood and other body fluids. It is substantially equivalent to conventional culture methodologies.

The methods are equivalent in the following respects:

દ

1. They have the same intended use: recovery and detection of aerobic and anaerobic microorganisms from body fluids.

- 2. They are both in-vitro diagnostic test systems which are based on microbial growth in media.
- 3. Organon Teknika's BacT/Alert® FAN Culture Bottles vield test results comparable to that seen with conventional culture methodology.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, along with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the eagle in a circular fashion. The logo is black and white.

DEC - 2 1997

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Rebacca A. Rivas Regulatory Specialist Organon Teknika Corp. 100 Akzo Ave. Durham, NC 27712

Re: K973325

> Trade Name: Bact/Alert Fan Culture Bottles Regulatory Class: I Product Code: MDB Dated: November 7, 1997 Received: November 10, 1997

Dear Ms. Rivas:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

## Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a Iegally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely vours.

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 510(k) Premarket Notification Organon Teknika Corporation BacT/Alert™ FAN Culture Bottles

ਤੋਂ ਤੋਂ k97 5 10(k) Number (il known):_

、Device Name:_ BACT/ALERI FAN Quiture Bottles (expended indications)

ations For Use:

d with the BacT/Alert Hicrobial Dete P/Alert FAN Culture Bottles etion of res for from blood and other body Cluide

ાજ્યનુભ

1°° EASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NE_JEDI

ર્ભ

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription use*_*
(Per 21 CFR 801.109)

Over The-Counts: Use ...

(Optional Formal 1-2-96)

. ... ... ....................................................................................................................................................................

John Ticehurst, Med Offices & Interim chief
(Division Sign-Off)
MICB

Division of Clinical Laboratory Devices 510(k) Number_JC 9

ﺘﺔ ﻟﻠﺘﻌﻠﻴﻘﺎﺕ

---

**Source:** [https://fda-staging.innolitics.com/device/K973325](https://fda-staging.innolitics.com/device/K973325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).