K973323 · Innovasive Devices, Inc. · MBI · Dec 1, 1997 · Orthopedic
Device Facts
Record ID
K973323
Device Name
ACL CLIP-IN
Applicant
Innovasive Devices, Inc.
Product Code
MBI · Orthopedic
Decision Date
Dec 1, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Innovasive Devices, Inc., ACL Clip-In is intended to be a fixation device in the repair of tendons and ligaments in Cruciate Reconstructive surgery.
Device Story
ACL Clip-In is a two-component endosteal fixation device for cruciate ligament reconstruction; consists of an externally threaded Base and a linkage Eyelet. The Base is endoscopically delivered through a tibio-femoral tunnel and inserted into the antero-lateral femur using a hex driver. The Eyelet, featuring a transverse opening for graft attachment, locks into the Base via a flexible wire mechanism. The design allows for axial locking while permitting rotational adjustment of the graft position post-insertion. The system includes specialized tools for graft repositioning and device removal. Used by surgeons in an operating room setting. The device provides secure graft-to-bone fixation, facilitating graft incorporation and stability during the healing process.
Clinical Evidence
Bench testing only. Comparative testing against the predicate device evaluated the strength of the implant-implant interface and the implant-graft-host bone construct (static and dynamic). Post-test evaluations assessed the host bone-implant interface, graft-implant interface, and graft incorporation. Results demonstrated that the ACL Clip-In performed at a minimum equivalent to or better than the predicate device in all tested parameters.
Technological Characteristics
Two-component implant: Base and Eyelet. Materials: Ti-6Al-4V ELI (ASTM F-136) and Nitinol (Ni-Ti alloy). Base diameters: 8mm, 10mm, 12mm. Sensing/Actuation: Mechanical locking via flexible wire. Delivery: Endoscopic, manual insertion via hex driver. Sterilization: Not specified.
Indications for Use
Indicated for fixation of ligament and tendon grafts in cruciate ligament reconstruction surgeries.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K973323
## 510 (K) SUMMARY
# DEC - 1 1997
- Date of Summary Preparation 1 Sept. 2, 1997
### Submitter's Name & Address 2 Innovasive Devices, Inc. 734 Forest Street, Marlborough, MA 01752 Telephone: 508-460-8229 508-460-8997 Fax:
- 3 Contact Person MR. Eric Bannon Vice President, QA/RA
#### 4 Manufacturer's Contact Person
Mr. T. Wade Fallin Exec. Vice President Innovasive Devices, Inc. 180 South 600 West Logan, UT 84321 Telephone: 801-753-7675 Fax: 801-753-7698
- 5 Proprietary Device Name ACL Clip-In
- 6 Common Name Endosteal Fixation Device
#### 7 Classification Name
Smooth or Threaded Metallic Bone Fixation Fastener. Single/multiple component bone fixation appliances and accessories
- 8 Classification Reference 21CFR 888.3030
- 9 Device Product Code 87 MBI
- 10 Regulatory Class Class II
{1}------------------------------------------------
#### 11 Materials
The following material is used in the manufacturing of the Base and Eyelets. Ti-6A1-4V ELI per ASTM F-136. Nitinol Wire (Ni-Ti Alloy)
#### Indications for Use 12
Innovasive Devices, Inc., ACL Clip-In is indicated for use in the fixation of ligament and tendon grafts in cruciate ligament reconstruction surgeries.
#### Description 13
### Intended Use
The Innovasive Devices, Inc., ACL Clip-In is intended to be a fixation device in the repair of tendons and ligaments in Cruciate Reconstructive surgery.
### Technological Characteristics
The ACL Clip-In device is a two component implant used in the fixation of ligament and tendon grafts in cruciate ligament reconstruction surgeries. The two device components are the Base and the Eyelet which are assembled together in the bone during the surgical procedure.
The ACL Clip-In device consists of an outer threaded Base which is designed to be endoscopically delivered through a prepared tibio-femoral tunnel and manually inserted into the antero-lateral femur using a hex driver. The proximal end of the Base has a flexible wire assembled to it.
The Eyelet has a smooth transverse opening to accept the soft tissue without damaging it. The proximal end of the Eyelet has a circular groove which is of a larger diameter of the flexible wire in the Base. The flexible wire provides for a spring like property as the Eyelet passes through the Base. When the proximal end of the Eyelet engages with the flexible wire, it locks the Eyelet in place.
With regard to implant size, the Base comes in 8mm, 10mm and 12mm diameters. The Eyelets are size specific and can be used only with the specified base component. The components will be labeled to indicate the same.
#### Substantially Equivalent Predicate Devices 14
| FRA B<br>BLE | | | | | | |
|-----------------------------|---------------------|-----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|--|--|
| .<br>1 8 ---<br>100 100 100 | -40 D --------<br>S | יינו<br>C | .<br>1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1991 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -<br>THE CA | 00 | | |
| | | | | | | |
| MANUFACTURER'S<br>NAME | DEVICE NAME | 510(k) DOCUMENT NO. | 510(k)<br>CLEARED |
|----------------------------------------|--------------------------|---------------------|-------------------|
| Innovasive Devices,<br>Marlborough, MA | 8mm Ligament<br>fastener | K970316 | Yes |
{2}------------------------------------------------
#### Substantial Equivalence 15
### 15.1 Technological Characteristics
### 15.1.1 Similarities
- (a) The ACL Clip-In device shares very similar technological characteristics with the predicate device. Both devices have:
- 1) a bone fixation element, and
- 2) a linkage element that connects the graft to the bone fixation element.
- (b) The bone fixation element is an externally threaded, internally cannulated Base.
- (c) The linkage element is an Eyelet with a transverse opening for receiving the graft and a proximal pin which locks into the Base.
### 15.1.2 Technological Differences
The ACL Clip-in uses different materials that the predicate device.
#### 15.2 Rationale for Differences
- (a) The subject device uses a well proven material which provides for greater strength and biocompatability.
- (b) Another potential clinical advantage is afforded by the connection means between the bone fixation element and the linkage element. This connection provides an axial lock, but will allow for rotational movement. Thus, the surgeon can adjust graft position along the axis of the bone tunnel after the linkage element, with graft attached, is connected to the bone fixation element. This is accomplished by engaging the Clip-In removal tool on the distal end of the Eyelet and turning it to reposition the graft in a distal to inferior direction.
- (c) A further potential clinical advantage is the ability to disengage the linkage element from the bone fixation element without damaging the linkage element or the graft by use of a simple removal punch. This may be an advantage in the event that the surgeon may need to reconfigure the linkage-graft construction. This removal tool engages the proximal end of the assembled bone fixation element and linkage element, and provides release with a rotationally directed compression force. Also provided is a implant removal tool, which helps unthread the assembly (base + eyelet) via the tibio-femoral tunnel.
#### 16 Performance Data
No performance standards exist at this time for the regulatory class under which the ACL Clip-in is proposed for classification. However, in support of this submission, Innovasive
{3}------------------------------------------------
Devices, Inc. has conducted the voluntary device testing in order to assess the effects of the new technological characteristics on the modes of failure of the device.
The effects of different materials on the modes of failure of the subject device were assessed in comparative testing with a selected predicate device.
The testing performed was:
- Strength of implant-implant interface .
- Strength of implant-graft-host bone construct .
- a. Static
- b. Dynamic
Post-test evaluation included:
- Host Bone- Implant Interface ●
- Graft Implant Interface ●
- Graft Incorporation
With regard to testing, the subject device demonstrated at a minimum equivalent or better performance compared to the selected predicate device in each of the three tests outlined above. With regard to post test evalutaion, the subject device demonstrate equivalent or better performance compared to the selected predicate device in post dynamic test evaluations of the host bone-implant interface and the graft-implant interface. Provisions for graft incorporation are essentially identical between the subject and predicate devices, as all provide transverse support of the soft tissue loop and all provide the same amount of graft-host bone surface area contact to facilitate graft incorporation.
#### 17 Conclusion
Based on the design concept, use of standard material, feature comparisons to a selected predicate device, the device and predicate device testing, Innovasive Devices, Inc. believes that sufficient evidence exists to conclude that the ACL Clip-In Device is substantially equivalent to existing legally marketed endosteal fixation devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 1 1997
Mr. Eric Bannon ·Vice President, Quality Assurance/Regulatory Affairs Innovasive Devices, Inc. 734 Forest Street 01752 Marlborough, Massachusetts
K973323 Re: ACL Clip-In Trade Name: Regulatory Class: II Product Code: MBI and HWC Dated: September 2, 1997 Received: September 4, 1997
Dear Mr. Bannon:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
### Page 2 - Mr. Eric Bannon
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.
Cella M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE 3.
#### 3.1 Statement of Indications for Use
Innovasive Devices, Inc. (IDI) - ACL Clip-In is indicated for use in the fixation of ligament and tendon grafts in cruciate ligament reconstruction surgeries.
**Prescription Use**
**(Per 21 CFR 801.109)**
Acolla
Divis 510k) I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.