K973318 · Ortivus Us, Inc. · MSX · Feb 19, 1998 · Cardiovascular
Device Facts
Record ID
K973318
Device Name
MOBIMED SYSTEM, (PEGASUS AND POLARIS)
Applicant
Ortivus Us, Inc.
Product Code
MSX · Cardiovascular
Decision Date
Feb 19, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Ortivus AB Mobimed™ is a cardiac monitoring system consisting of a portable and stationary unit which communicate with each other using the Mobitex radio-based cellular telecommunications network. The cardiac data includes ECG, SpO2, NIBP, GCS, and patient information. The system is intended for use by trained personnel in emergency care, specifically emergency ambulance and pre-hospital care.
Device Story
Mobimed™ System comprises portable 'Pegasus' units (ambulance-mounted or handheld) and stationary 'Polaris' units (hospital-based). System acquires and telemonitors patient physiological data via Mobitex radio-based cellular network. Inputs include 12-lead ECG, VCG, SpO2, manually entered NIBP, and GCS scores. Pegasus unit displays alarms and data locally; transmits data to Polaris for hospital-side monitoring and review. Enables continuous patient monitoring during transport. Healthcare providers use transmitted data to assess patient status and inform clinical decision-making during pre-hospital care. Benefits include improved remote patient assessment and data continuity from field to hospital.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC-601-1), ECG performance (IEC 601-2.27), SpO2 performance (ISO 9919), power source testing, environmental testing (shock, humidity, vibration, temperature), EMC (CISPR 11, IEC 61000-4-2,3,4,5), and communication link validation.
Technological Characteristics
System consists of portable Pegasus and stationary Polaris units. Connectivity via Mobitex radio-based cellular network. Power via internal rechargeable battery, 12V automotive system, or AC line. SpO2 module is Nellcor MP204P. Data transmission includes ECG, SpO2, NIBP (manual entry), and GCS. No automatic NIBP measurement. Standards: IEC-601-1, IEC 601-2.27, ISO 9919, CISPR 11, IEC 61000-4.
Indications for Use
Indicated for use by trained personnel in emergency ambulance and pre-hospital care settings for cardiac monitoring of patients. Transmits ECG, SpO2, NIBP, GCS, and patient information.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Predicate Devices
LifePak® Cardiac Monitor/LifeNet RS100 Receiving Station (K912189)
{0}------------------------------------------------
K413318
K973318
## SECTION 2. SUMMARY AND CERTIFICATION
## 510(K) SUMMARY A.
FEB 1 9 1998
## Summary of Safety and Effectiveness
In accordance with 21 CFR 807.92, the following information constitutes the Ortivus AB summary for the Mobimed™ System.
| SUBMITTER'S NAME: | Ortivus AB |
|---------------------|------------------------------------------|
| ADDRESS: | Enhagsslingan 5<br>S-183 25 Täby, Sweden |
| CONTACT PERSON: | Jörgen Eklund, Quality Manager |
| TELEPHONE NUMBER: | 011-46-8-446 45 10 |
| FAX NUMBER: | 011-46-8-446 45 19 |
| DATE OF SUBMISSION: | 2 September 1997 |
1. Identification of device
Proprietary Name: Mobimed™ System, (Pegasus and Polaris) Common Name: Monitor, Cardiac; Radio-frequency physiological signal transmitter and receiver; Oximeter. Classification Status: Class II per regulations 870.2300 and 870.2910 and 870.2700 Product Codes: 74DRT, 74DRG, and 74DQA
- 2. Equivalent devices
Ortivus believes the Mobimed™ System is substantially equivalent to the LifePak® Cardiac Monitor/LifeNet RS100 Receiving Station manufactured by Physio-Control Corporation (K912189) The SpO2 module is the Nellcor MP204P, same as used in the NPB-75 made by Spegas (K964239)
- 3. Description of the Device
The Mobimed™ System consists of one or several stationary units called Polaris (typically hospital located) receiving patient cardiac information from mobile units called Pegasus. A mobile unit can be permanently mounted in an ambulance or operate as a portable unit. The Mobimed™ equipment can telemonitor (using radiobased cellular telecommunications network) a patient where he or she is at the moment. This makes it possible to monitor the patient and acquire data during transportation to the hospital. The system can acquire and transmit the following information:
- VCG or 12-Lead ECG (automatically and manually acquired reports). ●
- One channel ECG or VCG reports .
- Monitoring ECG (not transmitted)
- . Trends (automatically and manually acquired).
- Patient records/forms
{1}------------------------------------------------
- Alarms (presented on the Pegasus and not transmitted to Polaris) .
- Messages .
- 4. Intended use
The Ortivus AB Mobimed™ is a cardiac monitoring system consisting of a portable and stationary unit which communicate with each other using the Mobitex radio-based cellular telecommunications network. The cardiac data includes ECG, SpO2, NIBP, GCS, and patient information. The system is intended for use by trained personnel in emergency care, specifically emergency ambulance and pre-hospital care.
- 5. Technological characteristics, comparison to predicate device.
Like the predicate devices, the Mobimed™ System is intended to be used in an emergency care situation to communicate cardiac information from a remote site such as an ambulance to a hospital. Standard ECGs can be transmitted, but in addition, SpO2, NIBP and GCS information can also be transmitted. The NIBP (Non-invasive blood pressure) values are entered manually into the Pegasus. This means that the values must be obtained by some other equipment, for example a standard blood pressure cuff. There is no automatic NIBP measurement equipment. The GCS (Glasgow Coma Scale) is a scoring system for the patient's status. GCS is an internationally well spread scoring system for estimating the degree of consciousness of the patient. The GCS value is based upon the observation of three fundamental physiological parameters; eye movements, motor response, and verbal response. Each parameter gets a numerical value based on standardized questions, and the sum of the three individual values is the GCS.
| Characteristic | Predicate device,<br>Lifepak 11 | Mobimed™ System |
|----------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Single lead ECG | Yes | No |
| 12 lead ECG | Yes | Yes |
| Frank lead (VCG)ECG | No | Yes |
| SpO₂ | No | Yes |
| Interpretive ECG | Yes | No |
| Digital telecommunications | Yes | Yes |
| Local ECG display | Paper and Monitor | Monitor only |
| Receiving station display | Paper only | Monitor and paper |
| Power sources | Internal rechargeable battery<br>or AC line (charging source) | Internal rechargeable battery<br>or 12 volt automotive<br>system (charging source)or<br>AC line |
| Labeling | Labels on devices, an<br>operating manual and a<br>physicians guide | Labels on devices, a<br>Pegasus User's Manual, and<br>a Polaris User's Manual |
Comnarison table
{2}------------------------------------------------
- 6. Discussion of performance testing.
An extensive collection of tests has been conducted and successfully completed, including electrical safety (to IEC-601-1), ECG performance (using appropriate sections of IEC 601-2.27), SpO2 performance (Using ISO 9919), power source (battery and charger), environmental (shock, humidity, vibration, temperature), EMC (using CISPR 11 and IEC 61000-4-2,3,4, and 5) and communications to and from the hospital based unit.
- 7. Conclusion
Based on extensive performance testing and a comparison to the predicate device, it is the conclusion of Ortivus AB that the Mobimed™ System is substantially equivalent to devices already on the marked (cleared by the 510(k) process) and presents no new concerns about safety and effectiveness. In fact, the use of cellular digital technology enhances communication capability over previous direct radio frequency communications systems.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
FEB 1 9 1998
Ms. Constance G. Bundy Ortivus, AB Enhagsslingan 5 SE-183 25 Täby, Sweden
Re: K973318 Trade Name: Ortivus AB Mobimed™ System Requlatory Class: II (two) Product Code: 74 MSX January 21, 1998 -Dated: February 3, 1998 Received:
Dear Ms. Bundy:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions . Failure to comply with the GMP regulation may result in" In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Ms. Constance G. Bundy
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE B.
510(k) Number_ K 97 3318
Device Name: Mobimed™ System
Indications for Use: The Ortivus AB Mobimed™ is a cardiac monitoring system consisting of a portable and stationary unit which communicate with each other using the Mobitex radio-based cellular telecommunications network. The cardiac data includes ECG, SpO2, NIBP, GCS, and patient information. The system is intended for use by ........ trained personnel in emergency care, specifically emergency ambulance and pre-hospital care.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE) M. f (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number_ Prescription Use OR Over the Counter Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.