K973284 · Roche Diagnostic Systems, Inc. · CHH · Dec 18, 1997 · Clinical Chemistry
Device Facts
Record ID
K973284
Device Name
COBAS INTEGRA HDL-CHOLESTEROL DIRECT
Applicant
Roche Diagnostic Systems, Inc.
Product Code
CHH · Clinical Chemistry
Decision Date
Dec 18, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1175
Device Class
Class 1
Indications for Use
The cassette COBAS INTEGRA HDL - Cholesterol Direct (HDL-D) contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative determination of HDL - cholesterol direct concentration in serum and plasma.
Device Story
COBAS INTEGRA HDL-D Reagent Cassette is an in vitro diagnostic system for use on the COBAS INTEGRA Analyzer; enables direct, automated quantitative determination of HDL cholesterol in serum and plasma without sample pretreatment. Principle of operation: polyanions and detergent selectively solubilize HDL cholesterol while rendering LDL, VLDL, and chylomicrons detergent-resistant; solubilized cholesterol undergoes sequential enzymatic oxidation by cholesterol esterase and cholesterol oxidase; hydrogen peroxide produced reacts with chromogens to form a colored dye. Absorbance measured at 552 nm is proportional to HDL cholesterol concentration. Used in clinical laboratories by trained personnel; streamlines workflow by eliminating centrifugation and manual precipitation steps required by older methods. Provides clinicians with rapid lipid profile data to assess CAD risk.
Clinical Evidence
Bench testing only. Performance evaluated using 258 samples. Comparison to predicate showed a correlation coefficient of 0.901 and linear regression of 0.77x + 0.33 mmol/L. Precision (total % CV) reported as 2.2%. Sensitivity is 8.2 x 10^-2 ΔA per mmol/L.
Technological Characteristics
Reagent cassette contains MOPS, polyanions, 4-amino-antipyrine, MgCl, N,N-bis-m-toluidine, cholesterol esterase, cholesterol oxidase, and horseradish peroxidase. Liquid reagent format. Operates via selective inhibition colorimetric assay. Designed for use on COBAS INTEGRA analyzer (absorbance-based measurement). On-board stability: 12 weeks. Assay range: 0.01 - 4.0 mmol/L.
Indications for Use
Indicated for the quantitative determination of HDL cholesterol in human serum and plasma for patients requiring lipid profile assessment, particularly those at risk for atherosclerotic Coronary Artery Disease (CAD).
Regulatory Classification
Identification
A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Roche company logo. The logo consists of the word "Roche" in bold, black letters, centered within a black hexagon. The hexagon is outlined in black and provides a border around the company name.
K973284
Dec 18, 1997
# 510(k) Summary
# Roche COBAS INTEGRA® HDL-D Reagent Cassette
In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.
#### I. Identification of 510(k) Sponsor:
Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
510(k) Submission dated August 28, 1997
Contact: Maria Feijoo Manager, Regulatory Affairs Phone: (908) 253-7310 Fax: (908) 253-7547
{1}------------------------------------------------
#### II. Device Name:
The device name, including both the trade/proprietary name and the classification name are provided in Table 1.
## Table 1
| Proprietary Name | Classification Name | Product Code | Regulation Number |
|-------------------------------------------|-------------------------|--------------|-------------------|
| COBAS INTEGRA HDL -<br>Cholesterol Direct | Cholesterol test system | CHH | 862.1175 |
{2}------------------------------------------------
## Identification of the legally marketed device to which the 510(k) sponsor claims III. equivalence:
The COBAS INTEGRA HDL-D Reagent Cassette is substantially equivalent to the currently marketed COBAS INTEGRA HDL reagent system (K951595).
#### Description of the Device/Statement of Intended Use: IV.
The COBAS INTEGRA HDL-D reagent cassette is intended for use with the COBAS INTEGRA Analyzer. The COBAS INTEGRA Analyzer and COBAS INTEGRA Reagent cassettes together provide an integrated system for in vitro diagnostic testing. The COBAS INTEGRA Analyzer along with 107 other Roche COBAS INTEGRA Reagent Cassettes were previously cleared on September 8, 1995 (K951595); January 25, 1996 (K954992); July 23, 1996 (K961824); October 31.1996 (K963292); January 21, 1997 (K964457) and August 12, 1997 (K972250).
The COBAS INTEGRA Analyzer utilizes three measuring principles, i.e., absorbance, fluorescence polarization and ion-selective electrodes. The analyzer has a throughput of up to 600 tests per hour with STAT samples prioritized and tested immediately. Random sample access, robotics and a user interface optimize time management and streamline workflow. The COBAS INTEGRA can store up to 68 COBAS INTEGRA Reagent Cassettes on board, 24 hours a day at 2-8°C. The COBAS INTEGRA Reagent Cassettes are compact and preparation-free with the added convenience of long term on-board stability. Barcode readers are used to identify newly loaded reagent cassettes, samples for patient identification, and rack inserts and to read calibration and control data from the cassette label. COBAS INTEGRA tests include chemistry, drugs of abuse, immunology, ion selective electrodes, therapeutic drug monitoring, and hematology reagents. For additional information on the COBAS INTEGRA Analyzer and its constituent modules, please refer to the Operator's Manual in Volumes 1 through 2, pages 92-703, of the original 510(k) submission (K951595).
Through this submission, it is the intention of Roche Diagnostic Systems to gain clearance for the COBAS INTEGRA HDL-D Cassette.
The COBAS INTEGRA HDL-D reagent cassette is an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative determination of High Density Lipoprotein cholesterol direct (HDL-D) in serum and plasma.
Low HDL levels are recognized as a strong and common risk factor for atherosclerotic Coronary Artery Disease (CAD). Therefore, determination of HDL cholesterol is routinely offered as part of a lipid profile. Usually VLDL and LDL are selectively precipitated from serum or plasma samples followed by determination of cholesterol in the HDL - containing supernatant. These techniques require a centrifugation step to remove the precipitated lipoproteins and thus cannot be fully automated. The COBAS INTEGRA HDL - Cholesterol Direct, however, allows for the
{3}------------------------------------------------
direct specific determination of HDL cholesterol in the presence of LDL. VLDL and chylomicrons without any sample pretreatment. Therefore, lending itself to automated routine analysis.
The principle of the COBAS INTEGRA HDL-D reagent cassette is based on the absorption of synthetic polyanions to the surface of lipoproteins. LDL, VLDL, and chylomicrons are transformed into a detergent-resistant form. whereas HDL is not. Combined action of polyanions and detergent solubilizes cholesterol from HDL, but not from LDL, VLDL, and chylomicrons. Solubilized cholesterol is oxidized by the sequential enzymatic action of cholesterol esterase and cholesterol oxidase. The hydrogen peroxide produced in this reaction is reacted with chromogens to form a colored dye. The increase in absorbance at 552 nm is directly proportional to the HDL cholesterol concentration of the sample.
### V. Summary of the technological characteristics of the new device in comparison to those of the predicate.
The COBAS INTEGRA HDL-D reagent cassette is similar to the COBAS INTEGRA HDL reagent system in that HDL cholesterol is quantitated using an enzymatic, colorimetric method (CHOD/PAP).
The differences are that the current COBAS INTEGRA HDL application requires pretreatment of the sample with a separating reagent prior to analysis whereas the COBAS INTEGRA HDL-D reagent cassette measures the HDL cholesterol concentration directly in the sample.
The current COBAS INTEGRA HDL reagent application is used in conjunction with the Roche HDL Cholesterol Precipitating Reagent and the COBAS INTEGRA Cholesterol reagent cassette (K951595). The Roche HDL Cholesterol Precipitating Reagent uses phosphotungstic acid and magnesium ions to precipitate the chylomicrons, VLDL, and LDL. After centrifugation, HDL remaining in the supernatant is quantitated by its cholesterol content using an enzymatic. colorimetric method (CHOD/PAP).
A summary of the similarities, differences and performance characteristics between the COBAS INTEGRA HDL-D reagent cassette and the COBAS INTEGRA HDL reagent system are listed in Table 2.
{4}------------------------------------------------
# Table 2
| | | <b>COBAS INTEGRA HDL-D reagent</b><br>cassette | <b>COBAS INTEGRA HDL reagent</b><br>system | | |
|---------------------------------|--|--------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|----------------------|
| Intended Use | | quantitative determination of HDL<br>cholesterol | quantitative determination of HDL<br>cholesterol | | |
| Methodology | | Selective inhibition colorimetric<br>assay (no pretreatment required) | Phosphotungstic acid pretreatment<br>used with colorimetric method<br>(CHOD/PAP) | | |
| Sample type | | Serum and plasma | Serum and plasma | | |
| Calibrator | | Roche HDL Calibrator Direct<br>lyophilized human serum<br>Assigned value ~1.58 mmol/L<br>(61 mg/dL) | Roche Calibrator (human)<br>Assigned value ~4.08 mmol/L (158 mg/dL) (undiluted) | | |
| Controls | | Roche Control Serum N and P<br>(human) | Roche Control Serum N and P<br>(human) | | |
| Reagent type | | Liquid | Granulate | | |
| Reagent (active<br>ingredients) | | R1: MOPS<br>Polyanions<br>4 - Amino-antipyrine<br>Mg Cl<br>R2: N,N-bis-m-toluidine<br>Cholesterol esterase<br>Cholesterol oxidase<br>Horseradish | Pretreatment: Phosphotungstic acid<br>R: HEPES<br>Sodium cholate<br>4 - Chlorophenol<br>4 - Amino-antipyrine<br>Cholesterol esterase<br>Cholesterol oxidase<br>Horseradish | | |
| On-board Stability | | 12 weeks | 8 weeks | | |
| Performance<br>Characteristics: | | | | | |
| Assay range | | 0.01 - 4.0 mmol/L<br>(0.386 - 155 mg/dL) | 0 - 5.0 mmol/L<br>(0 - 193 mg/dL) | | |
| | | | | | |
| Precision: | | Level 1 | Level 2 | Level 1 | Level 2 |
| Mean (mmol/L) | | 0.60 mmol/L<br>(23.2 mg/dL) | 1.41 mmol/L<br>(54.6 mg/dL) | 0.82<br>(31.7 mg/dL) | 1.42<br>(54.9 mg/dL) |
| % CV (within run) | | 1.4 | 1.1 | 1.2 | 0.85 |
| % CV (total) | | 2.2 | 2.3 | 2.7 | 5.5 |
| Sensitivity | | $8.2 X 10^{-2}$ ΔA per mmol/L<br>( $2.1 X 10^{-3}$ ΔA per mg/dL) | $6.4 X 10^{-2}$ ΔA per mmol/L<br>( $1.7 X 10^{-3}$ ΔA per mg/dL) | | |
| Accuracy: | | | | | |
| Sample size (n) | | 258 | 232 | | |
| Corr. Coefficient | | 0.901 | 0.998 | | |
| Linear regression | | $0.77x + 0.33$ mmol/L<br>( $0.77x + 12.8$ mg/dL) | $0.99x - 0.05$ mmol/L<br>( $0.99x - 1.93$ mg/dL) | | |
{5}------------------------------------------------
## Brief discussion of the clinical and nonclinical tests relied on for a determination of VI. substantial equivalence:
Table 2 shows the clinical and nonclinical performance characteristics of the COBAS INTEGRA HDL-D Reagent Cassette. This information demonstrates that the performance of this device is substantially equivalent to the currently marketed COBAS INTEGRA HDL reagent system.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three tail feathers, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The logo is black and white.
Public Health Service
DEC 18 1997
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Maria Feijoo . Manager, Regulatory Affairs Roche Diagnostic Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
Re : K973284 COBAS INTEGRA HDL-D Requlatory Class: I Product Code: LBR, CHH Dated: November 25, 1997 Received: November 26, 1997
Dear Ms. Feijoo:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and -prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਕੇ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note:
this response to your premarket notification submission does not affect any obligation you might have under sections 531 --through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{7}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Litman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{8}------------------------------------------------
Page__________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known)
Device Name: Roche COBAS INTEGRA HDL-D Reagent Cassette
Indications for Use:
The cassette COBAS INTEGRA HDL - Cholesterol Direct (HDL-D) contains an in vitro diagnostic reagent system intended for use on COBAS INTEGRA for the quantitative determination of HDL - cholesterol direct concentration in serum and plasma.
L. Mitchell Chappell for A.W. Montgomery
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) ---
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.