KURZ precious metal tympanostomy tubes are intended to provide a means of aerating and draining the middle ear in patients with otitis media and to reduce the likelihood of recurrent postoperative middle ear infections for the duration in which the tube remains in situ. The MINIMAL Ventilation Tube (Benz) has been designed for 1. temporary ventilation of the middle ear in secretory otitis media in adult patients for whom paracentesis alone is not sufficient, yet retention time of conventional ventilation tubes would be too long. 2. It can also effectively be applied in cases of narrow meatus, atrophic drum and post-tympanoplasty tube dysfunctions. Due to its tiny size (dia. 0.6 mm), insertion after tympanoplasty considerably reduces risk of permanent tympanum perforation. Following surface anaesthesia, the tube is pushed through the tympanic membrane with alligator forceps without prior paracentesis.
Device Story
Tympanostomy tubes; bobbin-shaped or narrow-tube designs; materials include gold-platinum alloy, gold-plated silver, titanium, and gold-plated stainless steel. Implants placed in tympanic membrane via myringotomy to aerate/drain middle ear; reduce recurrent infection risk. MINIMAL tube features sharpened end for insertion without prior paracentesis; held by tissue tension. Operated by physicians using alligator forceps under surface anesthesia. Output is physical conduit for pressure equalization and fluid drainage. Benefits include reduced risk of permanent tympanum perforation and improved retention in specific anatomical cases.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: gold-platinum alloy, gold-plated silver, titanium, gold-plated stainless steel (1.4301). Form factors: bobbin-shaped, narrow tube with sharpened end, cone-shaped external flange. Features: optional wire for placement/removal, unilateral/bilateral eye extensions. Single patient use; sterile packaging.
Indications for Use
Indicated for patients with otitis media requiring middle ear aeration and drainage. MINIMAL Ventilation Tube (Benz) indicated for adult patients with secretory otitis media, narrow meatus, atrophic drum, or post-tympanoplasty tube dysfunction.
Regulatory Classification
Identification
A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.
Predicate Devices
Hubbard Airplane (Xomed-Treace)
Tiny Tytan Vent Tubes (Xomed-Treace)
Reuter Bobbin
Collar Button
Spoon Bobbin
Shah Type Tube
Beveled Bobbin
Shepard Grommets
Submission Summary (Full Text)
{0}------------------------------------------------
KURZ
Heinz Kurz GmbH Medizintechnik
Hausanschrift: Tübinger Straße 3 D-72144 Dußlingen
Telefon (0 70 72) 91 79-0 Telefax (07072) 91 79-7
Heinz Kurz GmbH · Medizintechnik · Postfach 39 · D-72142 Dußlingen
K973226
NOV - 4 1997
# 6. 510(k) SUMMARY
As required by Section 807.92(c)
- Submitter 1807.92(a){ 1)} (1) Dagmar S. Mäser Business Support International Amstel 320-1 Amsterdam, 1017AP The Netherlands
Tel. 011 - 31 - 20 - 428 9591 -428 9429 Fax
bsi@xs4all.nl e-mail
Identification of Devices 1807.92(a)(2)} (2) Tübingen Ventilation Tube Trade Names a. Tympanic Ventilation Tube b. DIABOLO Tympanic Ventilation Tubes C. MINIMAL Typanic Ventilation Tube (Benz) d. Tympanostomy Tube; Ventilation Tube Common Name Tube, Tympanostomy Classification Name Product Code 77 ETD Class II Regulatory Number 874.3880 Classification Panel ENT Branch (HFZ-470)
#### Identification of Predicate Devices 1807.92(a)(3)] (3)
The KURZ ventilation tubes are substantially equivalent to a wide variety of ventilation tube designs produced by XOMED and XOMED-TREACE, Jacksonville, FL, USA and by Smith & Nephew Richards, Bartlett, TN, USA. In terms of indication, the short-term MINIMAL Ventilation Tube (Benz) is substantially equivalent to the Hubbard Airplane and the Tiny Tytan Vent Tubes, manufactured by Xomed-Treace.
These devices have been found substantially equivalent through the 510(k) premarket notification process.
6-1
Update: 10/06/97
Geschäftsführer: Heinz Kurz Trauce Kurz-Buczki USt - Id. Nr. DE 81 1570328
Volksbank Steinlach-Wiesaz eG 181 7 640 618 541 Konto-Nr. 70 150 001 Baden-Württembergische Bank Tübingen (BLZ 641 200 30) Konto-Nr. 120 140 2300 Amtsgericht Tübingen HRB 1424
{1}------------------------------------------------
# Heinz Kurz GmbH Medizintechnik
510(k) Summary (Cont'd)
.
:
P
#### Description of Devices 1807.92(a)(4)] (4)
- Typical bobbin-shaped design made of gold-platinum, Tübingen a. gold-plated silver, and titanium; with or without a wire for secure placement and removal. Internal diameter of Size 2 is min. 0.2 mm wider than that of most predicate devices ( SE: Reuter Bobbin, Collar Button).
#### Tympanic Ventilation Tube b.
- Bobbin shape with unilateral eye extension of 1 Eye . inner flange; made of gold-platinum alloy (SE: Spoon Bobbin, Shah Type Tube)
- Bobbin shape with bi-lateral eye extensions of inner 2 Eyes flange; made of gold-platinum alloy (SE: Most T-Tubes)
#### DIABOLO Ventilation Tube C.
External flange is cone-shaped to reduce the length of the internal hollow cylinder; made of gold-platinum alloy (SE: Beveled Bobbin; Shepard Grommets)
#### MINIMAL Ventilation Tube (Benz) d.
Narrow tube with sharpened end for insertion without prior paracentesis. Material: gold-plated stainless steel (1.4301).
The devices are implants for single patient use. They are sold in an indivi dual sterile package in a sealed carton.
#### Intended Use of Devices [807.92(a)(5)] (5)
KURZ precious metal tympanostomy tubes are intended to provide a means of aerating and draining the middle ear in patients with otitis media and to reduce the likelihood of recurrent postoperative middle ear infections for the duration in which the tube remains in situ.
The MINIMAL Ventilation Tube (Benz) has been designed for
- temporary ventilation of the middle ear in secretory otitis media in adult ﺎ . patients for whom paracentesis alone is not sufficient, yet retention time of conventional ventilation tubes would be too long.
- 2. It can also effectively be applied in cases of narrow meatus, atrophic drum and post-tympanoplasty tube dysfunctions. Due to its tiny size (dia. 0.6 mm), insertion after tympanoplasty considerably reduces risk of permanent tympanum perforation.
Following surface anaesthesia, the tube is pushed through the tympanic membrane with alligator forceps without prior paracentesis.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a bold, sans-serif font. The words are stacked on top of each other. The text is black against a white background.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dagmar S. Maser Business Support International Heinz Kurz GmbH Medisintechnik Amstel 320-1 1017 AP Amsterdam The Netherlands
Re:
NOV - 4 1997
K973226 Tympanostomy Tubes; Tüububgen Ventilation Tubes (Standard) Tympanic Ventilation Tube (with one eye); Tympanic Ventilation Tube ( with 2 eyes); DIABLO Ventilation Tube and MINIMAL Ventilation Tube (Benz) Dated: July 20, 1997 Received: August 27, 1997 Regulatory class: II 21 CFR 874.3880/Procode: 77 ETD
Dear Mr. Maser:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
William Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Heinz Kurz GmbH Medizintechnik
| 510(k) Number | K973226 |
|---------------|-------------------------------------------------------------|
| Device Names | Tübingen, Tympanic (1 or 2 eyes), DIABOLO Ventilation Tubes |
# INDICATIONS FOR USE:
Tympanic tubes are inserted into the eardrum (myringotomy) for drainage and temporary ventilation of the middle ear subsequent to acute otitis media.
Tympanic tubes with one or two eyes are designed to increase the retention time in the eardrum.
The substantially shortened passage of the DIABOLO reduces the length of the internal hollow cylinder and the smooth surface of the gold seems to diminish the risk of luminal occlusion.
## Placement:
The tubes are placed in the tympanic membrane to provide the means for drainage of any fluid buildup in the middle ear while creating an avenue for the passage of air to equalize pressure on either side of the drum.
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Saylor
(Division Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
Prescription Use (Per CFR 801 109) Over-The-Counter Use _
OR
{4}------------------------------------------------
Heinz Kurz GmbH Medizintechnik
973226 (10/06/97) 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Device Names MINIMAL Tympanic Ventilation Tube (Benz)
### INDICATIONS FOR USE:
- Temporary ventilation of the middle ear in serous otitis media in adult patients. 1.
- In cases of narrow meatus, atrophic drum or post tympanoplasty tube dysfunction. 2. Due to its tiny size (Diameter 0.6 mm), insertion after tympanoplasty considerably reduces the risk of permanent perforation of the tympanum.
This tympanic ventilation tube, constructed of gold-plated stainless steel (1.4301), has been designed for temporary ventilation in special cases in which paracentesis alone is not sufficient, yet retention time of conventional tubes would be too long. The tube features a sharpened end that can be pushed through the tympanic membrane with alligator forceps without prior paracentesis.
If the effusion is somewhat thicker than expected, an additional small paracentesis beside the tube is recommended, via which the mucous tympanum can easily be aspirated due to the pressure balance which is achieved through the ventilation tube.
Should secretion encrustation occur, the MINIMAL Ventilation Tube can be replaced without the need for surface anaesthesia and without pain to the patient as it has no internal flange but is kept in position through tissue tension.
### Placement:
Following surface anaesthesia, the tube is pushed through the tympanic membrane with alligator forceps without prior paracentesis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED);
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
| 510(k) Number | K973226 |
|---------------|---------|
|---------------|---------|
| Prescription Use | OR | Over-The-Counter Use |
|-------------------|----|----------------------|
| (Per CFR 801 109) | | |
KRZ-INDM.ini
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.