The IMTEC Bone Tac System is designed to stabilize a Barrier Membrane onto a region of cortical plate. This may be used in craniofacial, maxillofacial or mandibular bone. Considerations such as quality of bone, bone type, functional loads exerted, general patient health and others should be carefully evaluated prior to use.
Device Story
IMTEC Bone Tac is a titanium alloy fixation device used to stabilize barrier membranes against cortical bone during surgical procedures. The system includes bone tacs, a seating instrument, and a mallet. The clinician places the tac through the barrier membrane into the bone using the seating instrument and a mallet strike. In dense cortical bone, a pilot drill hole may be required. The device is used in clinical settings by surgeons. It provides mechanical stabilization of the membrane to facilitate healing. Post-procedure, radiographs document tac position. Tacs are intended for single use and may be removed via surgical scalpel or elevator if necessary.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and design similarities to predicate devices.
Technological Characteristics
Materials: Titanium 6Al-4V ELI (tacs), 303 Stainless Steel (instrument handles), Polycarbonate (holder cover). Principle: Mechanical bone fixation via barbed metallic tack. Form factor: Low-profile head with stabilization barb. Sterilization: Non-sterile (requires user sterilization). Connectivity: None.
Indications for Use
Indicated for stabilization of barrier membranes on cortical bone plates in craniofacial, maxillofacial, or mandibular regions. Contraindicated in patients with vascular impairment, systemic/local infection, diabetes, radiation therapy, steroid therapy, anticoagulant therapy, clotting disorders, or other systemic/metabolic limitations compromising healing.
Regulatory Classification
Identification
An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.
Predicate Devices
Interpore International, IMZ Bone Tack System (K952167)
Interpore International, IMZ Membrane Tack System (K960945)
Submission Summary (Full Text)
{0}------------------------------------------------
# K973180
#### 510 (k) SUMMARY
#### SUBMITTED BY:
M. K. Patterson, Jr., PhD Sr. Vice President Regulatory Affairs IMTEC Corporation 2401 North Commerce Ardmore, Oklahoma 73401 (405) 223-4456
F.D.A. Registration Number: 1645158 Owner / Operator Number: 9003407
Date Submitted: August 20, 1997
### CLASSIFICATION/COMMON OR USUAL NAME/DEVICE NAME:
Classification Name: Smooth or threaded metallic bone fixation fastener (ref: 21 CFR 888.3040); Product Code: DZL Common/Usual Name: Bone Tack, Bone Screw or Bone Fixation Device. Proprietary Name: IMTEC Bone Tac
#### PREDICATE DEVICE:
Interpore International, IMZ Bone Tack System, K952167 Interpore International, IMZ Membrane Tack System, K960945
#### DEVICE DESCRIPTION:
The IMTEC Bone Tac is fabricated of titanium alloy and has a thin low profile head and a barb at the tip for stabilization. It is designed to stabilize guided tissue membranes during the healing process by providing an attachment mechanism to the adjacent bone at the surgical site. The device will be marketed non-sterile.
{1}------------------------------------------------
## INDICATIONS FOR USE:
The IMTEC Bone Tac System is designed to stabilize a Barrier Membrane onto a region of cortical plate. This may be used in craniofacial, maxillofacial or mandibular bone. Considerations such as quality of bone, bone type, functional loads exerted, general patient health and others should be carefully evaluated prior to use.
## PRINCIPLES OF OPERATION:
1. An aseptic technique is required for predicable success.
2. The tacs and instrumentation are provided NON-STERILE and must be appropriately sterilized prior to use.
3. Remove the cap from the seating instrument. Firmly press the seating instrument onto the head of the desired length tac in the tac holder. This will purchase the tac to be carried to the placement site. Seat the tac through the barrier material only with the tip of the seating instrument perpendicular to the bone surface or the tac may be damaged. Placement should be accomplished with a single firm tap on the base of the seating instrument with the mallet. To disengage the tac, roll the seating instrument to one side. If the tac is not completely seated after the first mallet strike, place the protective cap on the seating instrument and complete the process by tapping on the seating instrument with the cap positioned against the tac head. The seating instrument may be damaged if it is repeatedly struck with the tac in the "held" position.
4. Repeat this process until the barrier is firmly anchored. Always store the seating instrument with the cap in place. Otherwise, it may become distorted and not able to purchase the tac head.
5. A pilot drill hole may be required in areas of dense cortical bone. The twist drill required for this process is delicate and must be used with a cautious and light pressures to prevent fracture. A pilot hole should be placed through laminar bone when it is being used as the barrier.
6. Suture the soft tissue in the usual manner when the surgical procedure is complete. Every effort should be made to assure primary closure of all areas.
7. Radiographs should be taken to document the number and position of the tacs. This information should also be noted in the patient's record.
8. In instances where tac removal is required, after normal surgical exposure the tacs may be dislodged by using a surgical scalpel, periosteal elevator or other similar thin flat instrument. The tacs should be removed, inventoried and discarded. REUSE IS NOT RECOMMENDED. The surgical site is now resutured.
{2}------------------------------------------------
# CONTRAINDICATIONS:
General patient evaluation is critical prior to any surgical procedure. Contraindications include, but are not limited to, vascular impairment, systemic or local infection, diabetes, radiation, steroid, or anticoagulant therapy, clotting disorders or other systemic or metabolic limitations which would compromise healing.
## COMPLICATIONS:
Complications include those associated with any osseous surgical procedure and an esthetic usage. Type and severity of complications may indicate tac removal at the discretion of the clinician.
## MATERIALS OF CONSTRUCTION:
| Bone Tac | Titanium 6Al-4V ELI |
|------------------------------------|-------------------------------------|
| Bone Tac Instrument | |
| Handle | 303 Stainless Steel |
| Tip and Cap | Titanium 6Al-4V |
| Autoclavable Tac Holder | |
| Stainless Steel w/Titanium Inserts | 303 Stainless Steel/Titanium 6Al-4V |
| Cover | Polycarbonate (Autoclavable) |
| Bone Tac Mallet | |
| Handle and Head | Stainless Steel |
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
The design, material, configuration,method of sterilization and other technological characteristics of IMTEC Bone Tac are similar to the currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 8 1997
M.K. Patterson, Jr., Ph.D. Sr. Vice President Regulatory Affairs Imtec Corporation 2401 North Commerce Ardmore, Oklahoma 73401
Re: K973180 Trade Name: Imtec Bone Tac Requlatory Class: İI Product Code: DZL Dated: August 20, 1997 Received: August 25, 1997
Dear Dr. Patterson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, " through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{4}------------------------------------------------
Pageg 2 - Dr. Patterson
through 542 of the Act for devices under the Electronic chrough 942 or the not ro- rouisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
K. Clatton
thy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
. Enclosure
{5}------------------------------------------------
510(k) Number (if known):
IMTEC BONE TAC SYSTEM Device Name:
Indications For Usc:
i
The IMTEC Bone Tac System is designed to stabilize a Barrier Membrane onto a region of cortical plate. This may be used in craniofacial, maxillofacial or mandibular bone. Considerations such as quality of bone, bone type, functional loads exerted, general patient health and others should be carefully evaluated prior to use.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off) |
|-----------------------------------------|
| Division of General Restorative Devices |
| 510(k) Number | K973180 |
|---------------|---------|
|---------------|---------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.