MODIFIED OLBERT BALLOON CATHETER SYSTEM, NOPROFILE OLBERT BALLOON CATHETER SYSTEM
K973113 · Boston Scientific Corp · FGE · Nov 10, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K973113
Device Name
MODIFIED OLBERT BALLOON CATHETER SYSTEM, NOPROFILE OLBERT BALLOON CATHETER SYSTEM
Applicant
Boston Scientific Corp
Product Code
FGE · Gastroenterology, Urology
Decision Date
Nov 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Modified Olbert@ Balloon Catheter products are recommended for biliary tract stenosis/sphincter stenosis and gastrointestinal tract dilatation through a flexible endoscope. The recommended application is printed on the label. The recommended application on the label may refer to any combination of transpapillary, pyloric, colonic, anastomotic, and/or esophageal dilatation.
Device Story
Modified Olbert Balloon Catheter is a balloon dilatation catheter used for biliary and gastrointestinal tract strictures. Device consists of a polyurethane balloon mounted on a catheter shaft; operated by a physician through a flexible endoscope. Principle of operation involves inflation of the balloon to mechanically dilate stenotic regions. Output is the physical expansion of the stricture, intended to improve patency. Used in clinical settings; physician-operated. Benefits include minimally invasive treatment of strictures. Device is provided sterile for single use.
Clinical Evidence
Bench testing only. Biocompatibility assessment performed on patient- and fluid-contact materials with satisfactory results. No clinical data presented.
Technological Characteristics
Polyurethane balloon; catheter shaft length 180 cm; diameter 5.8-7.8 Fr. Balloon diameter 4-14 mm; length 2.5-4.0 cm; max pressure 9-12 atm. Sterilized via ethylene oxide. Packaging: styrene/butadiene copolymer (SBS) tray with Tyvek lid and pouch.
Indications for Use
Indicated for patients requiring biliary tract stenosis/sphincter stenosis or gastrointestinal tract dilatation via flexible endoscope. Contraindicated for surgically intervened bile ducts, totally impassable strictures, or strictures caused by external mechanical obstruction (e.g., sutures, metallic clips).
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
1-of 3
# SECTION 9 510(K) SUMMARY
## NOV 1 0 1997
## FOI RELEASABLE
Pursuant to § 13(1)(3)(A) of the Food, Drug, and Cosmetic Act, Boston Scientific Corporation is required Pursualit to your results of the result, Dreg, and secures.
to submit with this Premarket Notification " ...dequate summary of any information respecting setty and to submit with this frometion will be made available upon request of any person." Boston encentific Corporation chooses to submit a summary of information respecting safety and effectiveness.
| > Common/Usual Names: | Balloon Catheter | | |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------|------------------|----------------------|
| > TRADE/PROPRIETARY NAME: | Olbert Balloon Catheter System®, NoProfile® Olbert Balloon<br>Catheter System® | | |
| > Classification Name &<br>DEVICE CLASSIFICATION: | Class II | | |
| | Name | Number | 21 CFR Ref. |
| | Esophageal Dilator<br>Endoscope & Accessories | 78 KNQ<br>78 KOG | 876.5365<br>878.1500 |
| > DEVICE PANEL/BRANCH: | Gastroenterology-Urology (GU)<br>Gastro-Renal (GRDB) | | |
| > Owner/Operator: | Boston Scientific Corporation<br>One Boston Scientific Place<br>Natick, MA 01760<br>Owner/Operator No. 9912058 | | |
| > Contact Person: | Daniel J. Dillon, Senior Regulatory Affairs Specialist | | |
## INDICATIONS FOR USE
The Modified Olbert@ Balloon Catheter products are recommended for biliary tract stenosis/sphincter Inc Mounted Cloud Collection through a flexible endoscope. The recommended application is printed on the label.
#### CONTRAINDICATIONS
Not recommended for cascs in which the bile duct has been surgically intervened, a stricture which is totally impassable, or for dilatation of strictures due to external mechanical obstruction, i.e., suure and metallic clips.
{1}------------------------------------------------
## POTENTIAL COMPLICATIONS
Potential complications with biliary tract dilation include incre-hepatic damage, transient bleeding, post-operative cholangitis, and stricture of a primary hepatic duct. Potential complication from GI tract dilation include, but may not be limited to perforation , hemorthage, hematoma, septicemia./infection, allergic reaction to contrast medium.
# DESCRIPTIVE AND TECHNOLOGICAL CHARACTERISTICS OF PROPOSED AND PREDICATE DEVICES
Boston Scientific Corporation believes that the Modified Olbert® Balloon Catheter is substantially equivalent to the currently-marketed predicate devices. Table 9-1 compares the descriptive characteristics of these products.
## PERFORMANCE CHARACTERISTICS
Laboratory testing regarding characteristics was performed on Modified Olbert@ Balloon Catheter to verify its safety and performance. A biocompatibility assessment was performed on the patient- and fluidcontact materials of the Modified Olbert® Balloon Catherer with satisfactory results.
## PACKAGING, STERILIZATION, AND PYROGENICITY
The Modified Olbert® Balloon Catheter will be packaged a styrene/butadiene copolymer (SBS) tray with a Tyvek lid and a Tyvek outer pouch. The Modified Olbert® Balloon Catheter will be sterilized using ryvek no and a 1 your of your procol for ethylene oxide sterilization. Pyrogenicity testing will be performed on a periodic basis to monitor bacterial endotoxin levels.
## CONCLUSION
Boston Scientific Corporation believes that Modified Olbert® Balloon Catherer is substantially equivalent Doston Belonal o Ocedicate devices. Table 9-1 compares the descriptive characteristics of these to the carrently manologies in Table 9-1, the Modified Olbert® Balloon Catherer is equivalent in its indications for use, while being very similar in design and materials. In addition, Boston Scientific metemons for are, and biocompatibility information. The information. The information presented provides assurance that the Modified Obert® Balloon Catheter will meet the minimum requirements that are considered acceptable for its intended use.
{2}------------------------------------------------
| T ABLE 9-1: COMPARISON OF MODIFIED OLBERT® BALLOON CATHETER TO PREDICATE DEVICES |
|----------------------------------------------------------------------------------|
| |
| Modified Olbert®<br>Balloon Catheter<br>(This 510(k)) | NoProfile® with<br>Modified End Connector<br>(510(k) No. K934411) | Modified NoProfile®<br>(510(k) No. K945576) | NoProfile® with<br>Vectran Fibers<br>(510(k) No. K952887) | Olbert Balloon Catheter<br>for Biliary Dilatation<br>(510(k) No. K851647) | Vector TTS™ Balloon<br>Dilatation Catheter<br>(510(k) No. K961438) |
|-------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------|-----------------------------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------------------------|
|-------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------|-----------------------------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------------------------|
| USE | | | | | | |
|------------|------------------------------------------------------|---------|---------|---------|---------|--------------------------------|
| Indication | Dilatation of Biliary and<br>Other Non-GI Strictures | «- Same | «- Same | «- Same | «- Same | Dilatation of GI<br>Strictures |
| BALLOON | | | | | | |
|------------------------------------|--------------|------------|------------|------------|------------|---------------------------|
| Diameter | 4-14 mm | 3-10 mm | 3-10 mm | 3-12 mm | 4-14 mm | 6-25 mm |
| Length | 2.5-4.0 cm | 2.0-4.0 cm | 2-10 cm | 2-10 cm | 2.5-4.0 cm | 5.5-8.0 cm |
| Maximum<br>Recommended<br>Pressure | 9-12 atm | 10-12 atm | 10-12 atm | 12 atm | 9-12 atm | 2-6 atm |
| Material | Polyurethane | «- Same | «- Same | «- Same | «- Same | Ethylene Vinyl<br>Acetate |
| CATHETER SHAFT | | | | | | |
| Length | 180 cm | 180 | 40-180 cm | 40-180 cm | 40-180 cm | 180-240 cm |
| Diameter | 5.8-7.8 Fr | 4.8-7.8 Fr | 4.8-7.8 Fr | 4.8-7.8 Fr | 6-7.5 Fr | 7 Fr., tapers to 5 Fr |
8
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The text is in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2015
Boston Scientific Corporation Daniel J. Dillon Senior Regulatory Affairs Specialist One Boston Scientific Place Natick, MA 01760-1537
Re: K973113 Trade/Device Name: Olbert Balloon Catheter System® Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE, KNO Dated (Date on orig SE Itr): August 18, 1997 Received (Date on orig SE Itr): August 20, 1997
Dear Daniel J. Dillon,
This letter corrects our substantially equivalent letter of November 10, 1997.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{4}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical dovice reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D.
Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## SECTION 1 INDICATIONS FOR USE
510(k) Number: K973113
Device Name: Modified Olbert® Balloon Catheter
Indication for Use:
The Modified Olbert@ Balloon Catheter products are recommended for biliary tract stenosis/sphincter stenosis and gastrointestinal tract dilatation through a flexible endoscope. The recommended application is printed on the label. *
* The recommended application on the label may refer to any combination of transpapillary, pyloric, colonic, anastomotic, and/or esophageal dilatation.
Robert Ratliff/
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number 73117
Prescription Use _ (Per 21 CFR 801.109)
Over-the-Counter Use_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.