K973050 · Triangle Biomedical Sciences, Inc. · JOY · Nov 7, 1997 · Hematology
Device Facts
Record ID
K973050
Device Name
MICRO21 WITH WBC ESTIMATE
Applicant
Triangle Biomedical Sciences, Inc.
Product Code
JOY · Hematology
Decision Date
Nov 7, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5260
Device Class
Class 2
Indications for Use
Intelligent Medical Imaging, Inc.'s MICRO21 with WBC Estimate is a laboratory instrument for locating, digitally storing and displaying white blood cells to aid the technologist in performing the WBC Estimate. Examination and determination of the results must be performed by qualified individuals.
Device Story
The MICRO21 with WBC Estimate is an automated microscopic system used in clinical laboratories. It processes glass slides containing stained blood smears. The device locates WBCs, captures and stores digital images, and displays them on a monitor for technologist review. After the technologist verifies the classification of the cells, the device calculates a WBC estimate using a formula based on the number of confirmed WBCs and the number of low-power fields visited. The output is a reported range of WBCs/uL displayed on a final report. This system aids technologists in performing WBC estimates, replacing manual microscopic estimation methods. It benefits patients by providing an automated, standardized approach to WBC estimation, reducing manual labor and potential variability in cell counting.
Clinical Evidence
Bench testing only. Accuracy was evaluated using 86 blood smear samples with total WBC counts ranging from 100 to 36,100 cells/uL. The device's WBC estimate was correlated against automated cell counters (TOA Sysmex NE 9000, Coulter STKS, Technicon H2) and manual WBC estimation. Intra-instrument and within-specimen precision were assessed by processing 10 smears 10 times each. Three technologists performed image reviews to verify cell classification.
Technological Characteristics
Automated microscopic system; utilizes digital imaging and automated cell location. Operates on stained glass blood smear slides. Calculation is based on a fixed formula using confirmed WBC counts and low-power fields visited. System includes a color monitor for technologist review and reporting. No specific materials or software architecture (e.g., ML/DL) are detailed beyond the automated processing of digital images.
Indications for Use
Indicated for use as a laboratory instrument to assist technologists in performing white blood cell (WBC) estimates by locating, storing, and displaying digital images of WBCs from stained blood smears.
Regulatory Classification
Identification
An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)
{0}------------------------------------------------
# 510(k) Summary
Designated Point of Contact: Otto Stellner Title: Vice President, Regulatory Compliance Project Manager: Alyn J. Hansen Date: August 5, 1997
Trade Name: MICRO21 with WBC Estimate Classification Name: Automated Cell Locating Device Classification Number: 81JOY Class: II Regulation Number: 864.5260
## Description
The MICRO21 with WBC Estimate is a new intended use that follows the same process as the currently approved MICRO21 with White Blood Cell (WBC) Differential (Diff). The MICRO21 with WBC Diff, Ref. No. K925670/A, is an automated microscopic system that locates WBCs, stores digital images of the cells and displays the images in an organized manner to aid technologists in performing the WBC Diff procedure. The MICRO21 process is substantially equivalent to the manual microscopic process.
NOV - 7 19
The MICRO21 with WBC Estimate is an automated microscopic procedure that calculates an estimate of WBCs/uL using the information collected during the MICRO2/ with WBC Differential. Upon completion of the review process by a technologist, an algorithm calculates the WBC Estimate by using the number of classified WBCs and the number for low-power (200x magnification) fields visited. An estimated range for the WBCs/uL is calculated and reported. A summary of the MICRO21 with WBC Estimate process is as follows:
- Patient blood smears are made on glass slides using an automated blood smearing device. 1.
- Each slide is stained with Wright's stain. 2.
- The slides are barcoded, placed into a slide frame and inserted into the MICRO21 for processing. 3.
- The MICRO21 locates WBCs and stores digital images of the cells. 4.
- The technologist reviews the stored images on a color monitor and verifies the white cell classification. న్.
- The MICRO21 automatically calculates the WBC Estimate by dividing the classified WBCs by the 6. number of low-power fields visited.
- At the completion of the procedure, the WBC Estimate is displayed as a range and is available on the 7. Final Report display and printout.
#### Test Method 1, Accuracy
To confirm equivalence, 86 blood smear samples with Total WBC Counts ranging from 100 to 36,100 cells/JL were used. The study correlated the WBC Estimate on the MICRO21 with the total White Blood Cell Count generated by an automated cell counter such as the TOA Sysmex NE 9000, Coulter STKS or Technicon H2 and compared the MICRO21 with WBC Estimate to the Manual WBC Estimate. The automated cell counters were controlled and culibrated according to manufacturer's specifications at the various sites.
{1}------------------------------------------------
#### Test Method 2, Precision
To confirm the MICRO21 intra-instrument precision, one blood smear was prepared from each of 10 specimens. Each blood smear was processed 10 times in random order. The stored images were then presented to a technologist for review. The results were used to provide intra-instrument and within specimen precision.
# Test Method 3, Correlation, WBC Classification Review
The technologist reviewed the images of nucleated cells captured by the MICRO21 and verified the image classification. The number of confirmed white blood cells (real cells) after the review was completed was the number used in the MICRO21 calculation of WBC Estimate. Non-white blood cells were reclassified as smudges, ignore or nucleated red blood cells. Three technologists performed the MICRO21 reviews.
# Calculation of the MICRO21 with WBC Estimate
The method is based on the number of WBCs located within an area of optimal cell distribution and is correlated with the absolute WBC/yL of blood. The calculation of the MICRO2 / with WBC Estimate is based on the following formulation:
#### Real Cells x CF = MICRO21 WBC Estimate Fields
- Real Cells = the number of confirmed WBCs .
- Fields = the number of low-power (200x magnification) fields visited to capture the Real . Cells
- CF = Correction Factor, the numerical factor need to match MICRO2/ WBC Estimate with . Total WBC Count obtained using automated blood cell analyzer.
- MICRO2/WBC Estimate = WBC/uL of blood .
The Correction Factor was defined as the mean value of the 86 specimens tested.
#### Reporting the MICRO21 with WBC Estimate
The calculated MICRO21 WBC Estimate is displayed on the Report Screen and printed out in the Final Report. The reportable range is from 100 to ≥ 25,000 WBCs/pL and is displayed as follows:
MICRO21 WBC Estimate x , xxx - - y , yyy WBC/ uL
#### Manual WBC Estimate
The Manual WBC Estimate was obtained by reviewing 86 blood smears under 500 or 1,000x magnification and calculating an estimate based on white blood cells per 10 fields of view. Three technologists performed the manual reviews.
## Conclusion
Results are reported in the Summary of Results section and confirm that the MICRO21 with WBC Estimate is substantially equivalent to a Manual WBC Estimate. The results of this comparison confirm the safety and effectiveness of the MICRO21 WBC Estimate for the intended use as a laboratory instrument for locating, digitally storing and displaying white blood cells to aid the technologist in performing the WBC Estimate. Examination and determination of the results must be performed by qualified individuals.
> Page 4 of 22 4360 Northlake Blvd., Suite 214 · Palm Beach Gardens, FL 33410 Phone (561) 627-0344 · FAX (561) 627-0344
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with its wings spread, rendered in a simple, bold line drawing.
Public Health Service
I ... In U.........
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. R. Otto Stellner Vice President, Regulatory Compliance Intelligent Medical Imaging 4360 Northlake Boulevard, Suite 214 Palm Beach Gardens, Florida 33410
NOV - 7 1997
Re : K973050 Trade Name: MIRCO21™ with WBC Estimate Regulatory Class: II Product Code: JOY Dated: August 5, 1997 Received: Auqust 15, 1997
Dear Mr. Stellner:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{3}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Ky 10000
Page 1 of 1
510(k) Number :
Device Name: MICRO2I™ with WBC Estimate
Indications for Use:
#### For In Vitro Diagnostic Use
Intelligent Medical Imaging, Inc.'s MICRO21 with WBC Estimate is a laboratory instrument for locating, digitally storing and displaying white blood cells to aid the technologist in performing the WBC Estimate. Examination and determination of the results must be performed by qualified individuals.
INTELLIGENT MEDICAL IMAGING, INC. DOES NOT PROMOTE THE USE OF, PROVIDE SUPPORT FOR, MAKE CLAIMS OF EFFECTIVENESS OF OR ASSUME LIABILITY FOR THIS DEVICE IN ANY OTHER COMMERCIAL, EXPERIMENTAL OR INVESTIGATIONAL. APPLICATION.
Use of this device in any manner that is inconsistent with its intended use may be in violation of federal law.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number
Prescription Use
(per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.