K973013 · Vas-Cath, Inc. · LIT · Jun 19, 1998 · Cardiovascular
Device Facts
Record ID
K973013
Device Name
OPTI-PLAST (CENTURION) 5.5F ANGIOPLASTY CATHETER
Applicant
Vas-Cath, Inc.
Product Code
LIT · Cardiovascular
Decision Date
Jun 19, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Centurion Peripheral Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries, post-deployment dilatation of peripheral vascular stents and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The catheter is not for use in coronary arteries.
Device Story
Opti-Plast Centurion 5.5F PTA Catheter is a dual-lumen balloon dilatation catheter; one lumen for 0.035" guidewire insertion, one for balloon inflation/deflation with contrast media. Features two radiopaque gold marker bands for positioning. Used by physicians in clinical settings for peripheral vascular interventions. Device is inserted percutaneously; balloon is inflated to labeled diameter/length at 10 atm operating pressure to dilate stenotic vessels or stents. Output is mechanical force applied to vessel walls. Benefits include restoration of blood flow in peripheral arteries and dialysis fistulae. Performance verified via bench testing including burst strength, fatigue, and trackability.
Clinical Evidence
Bench testing only. Evaluated balloon burst strength, distensibility, inflation/deflation time, fatigue, joint/material strength, profile, sheath compatibility, flow rate, trackability, tip torque, puncture resistance, and scratch resistance. Testing included cycling to rated burst pressure and burst testing within a stent.
Indicated for patients requiring percutaneous transluminal angioplasty (PTA) of femoral, iliac, and renal arteries, post-deployment dilatation of peripheral vascular stents, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
<973013
C. R. Bard, Inc. 8195 Industrial Blvd. Covington, GA 30209
# JUN 1 9 1998
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
C. R. Bard, Inc., Radiology Division Submitter Information: A. 13183 Harland Dr. Covington, GA 30014
| Contact Person: | Donna J. Wilson |
|----------------------------------------------|----------------------------------------------------------------|
| Contact Person's Address: | 8195 Industrial Blvd.<br>Covington, GA 30014 |
| Contact Person's Telephone: | 770-784-6135 |
| Contact Person's FAX: | 770-784-6419 |
| Date of Preparation: | June 17, 1998 |
| B. Device Name: | Opti-Plast® Centurion 5.5F PTA Catheter |
| C. Predicate Device: | Opti-Plast® 5.5F PTA Catheter (#K933483) |
| Predicate Device for<br>Indications for Use: | Medi-Tech® Blue-Max™ Balloon Dilatation<br>Catheter (#K934191) |
#### D. Device Description:
The Opti-Plast Centurion 5.5F PTA Catheter is a dual lumen catheter with a balloon mounted on its distal end. One lumen accommodates the insertion guidewire and the other provides a channel for inflation and deflation of the balloon with contrast media. The catheter is designed to be used in conjunction with a 0.035" diameter guidewire. There are two radiopaque marker bands secured on the catheter body to indicate the position of the balloon within the vasculature. The balloon inflates to the labeled diameter and length at a 10 atmosphere Operating Pressure. The Rated Burst Pressure of the device ranges from 17-20 atmospheres, depending on balloon diameter (see attached table).
{1}------------------------------------------------
#### E. Indications for Use:
Centurion Peripheral Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries, post-deployment dilatation of peripheral vascular stents and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The catheter is not for use in coronary arteries.
### F. Technological Characteristics Summary:
Summarized in the attached table.
#### G. Performance Data:
Bench testing was conducted for the following characteristics: balloon burst strength, balloon distensibility, balloon inflation/deflation time, balloon fatigue, joints and material strength, catheter balloon profile, introducer sheath compatibility, catheter flow rate and maximum injection pressure, deflatability and trackability, tip torque performance, puncture resistance ("toughness"), balloon scratch resistance, balloon fatigue (cycling to rated burst pressure) within a stent and balloon burst within a stent. The test results indicate that the Opti-Plast Centurion 5.5F PTA Catheter is substantially equivalent to the stated predicate device, that there are no new safety or effectiveness issues, and that the device can be utilized for its stated indication.
{2}------------------------------------------------
## Technological Characteristics Summary
.
| Features | Proposed Vas-Cath Opti-Plast®<br>Centurion 5.5F PTA Catheter | Vas-Cath Opti-Plast® 5.5F PTA<br>Catheter (K933483) | Medi-tech® Blue-Max™ Balloon<br>Dilatation Catheter (K934191) |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indicated Use | PTA of the femoral, iliac and renal<br>arteries, post-deployment dilatation<br>of peripheral vascular stents and for<br>the treatment of obstructive lesions of native<br>or synthetic arteriovenous dialysis fistulae. | PTA of vessels of the peripheral<br>vascular system | PTA of the Iliac, Femoral and Renal Arteries<br>and for the treatment of obstructive lesions of<br>native or synthetic arteriovenous dialysis fistulae |
| Balloon Material | Nylon | Nylon | Poly-5TM |
| Inflated Balloon Diameters | 4 mm - 10 mm | 4 mm - 10 mm | 4 mm - 10 mm |
| Inflated Balloon Length | 2 cm - 4 cm | 2 cm - 10 cm | 2 cm - 10 cm |
| Recommended Operating<br>Pressure | 10 atmospheres | 6 atmospheres - 8 atmospheres | NA |
| Rated Burst Pressure | 17 atmospheres - 20 atmospheres | 8 atmospheres - 12 atmospheres | 17 atmospheres |
| Material for catheter shaft,<br>tip, collar, and extensions | Nylon Co-polymer | Nylon Co-polymer | NA |
| Coating on shaft | Glissando | None | Medi-Glide |
| Radiopaque shaft | Yes | Yes | NA |
| Catheter Shaft Diameter | 5.5 French | 5.5 French | 5.8 French |
| Shaft Length | 50 cm - 140 cm | 50 cm - 120 cm | 40 cm - 120 cm |
| Shaft configuration | Double lumen | Double lumen | Double lumen |
| Marker Bands | Yes, Gold | Yes, Tantalum | Yes, material NA |
| Tip Length | 3 mm | 5 mm | 5 mm |
| Guidewire Capability | 0.035 inch diameter | 0.035 inch diameter | 0.035 inch diameter |
NA - Not available from promotional literature
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 9 1998
Ms. Donna J. Wilson Director, Requlatory Affairs C.R. Bard, Inc. 8195 Industrial Boulevard Covington, GA 30209
Re: K973013 Opti-Plast® Centurion 5.5F PTA Catheter Requlatory Class: II (two) Product Code: 74 LIT Dated: March 20, 1998 March 23, 1998 Received:
Dear Ms. Wilson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
### Page 2 - Ms. Donna J. Wilson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name: Bard® Opti-Plast® Centurion 5.5 F PTA Catheter
Indications for Use:
Centurion Peripheral Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries, post-deployment dilatation of peripheral vascular stents and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The catheter is not for use in coronary arteries.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
V Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use_
(Optional Format 1-2-96)
i
Division Sign Off
(Division Sign-Off) Division of Cardiovascular, Resphalory, and Neurological Devices x 974013 510(k) Number -----
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.