# THE C-FIT & DC-FIT ACETABULAR CUP SYSTEM (K972995)

_Corin USA · JDI · Nov 10, 1997 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K972995

## Device Facts

- **Applicant:** Corin USA
- **Product Code:** [JDI](/productcode/JDI.md)
- **Decision Date:** Nov 10, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3350
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Relief of pain and restoration of hip function following the effects of osteo, rheumatoid and post-trauma disease effects, avascular necrosis, and total hip revision. The C-Fit and DC-Fit plasma sprayed acetabular components are intended for use ONLY with bone cement.

## Device Story

Acetabular cup system for hip arthroplasty; used to resurface acetabulum. Components include cobalt chrome alloy cups with plasma spray coating for mechanical interlock with PMMA bone cement; used with UHMWPE inserts, femoral stems, and modular heads. System provides mechanical fixation in cemented hip procedures. Benefits include pain relief and restored hip function for patients with degenerative joint disease or revision needs. Used by orthopedic surgeons in clinical settings.

## Clinical Evidence

No clinical trials conducted for this submission. Evidence relies on mechanical bench testing and historical clinical use data (10 years for C-Fit, 2 years for DC-Fit) outside the USA with no reported significant postoperative problems.

## Technological Characteristics

Materials: Cobalt chrome alloy. Surface: Plasma sprayed coating for mechanical interlock with PMMA bone cement. Form factor: Modular acetabular cups, 42mm-70mm diameter range. Fixation: Cemented, screw-compatible. Sterilization: Not specified.

## Regulatory Identification

A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Mallory Head Acetabular Cup range

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "510(k) SUMMARY" in bold, underlined font. Above this text is a handwritten string of characters that appears to be "K972995". The handwriting is in black ink and is somewhat stylized.

NOV 10 1997

| Name of Company:    | Corin Medical Ltd                                                                         |
|---------------------|-------------------------------------------------------------------------------------------|
|                     | The Corinium Centre                                                                       |
|                     | Cirencester                                                                               |
|                     | Gloucestershire                                                                           |
|                     | GL7 1YJ                                                                                   |
|                     | England                                                                                   |
| Name of Device:     | C-Fit and DC-Fit Acetabular Cup System                                                    |
| Device Description: | Acetabular Cup System for use with previously cleared femoral stems<br>and modular heads. |

The acetabular cups are available in a range of diameters and are plasma sprayed.

The devices are used to resurface the acetabulum and are used in combination with UHMWPE inserts, femoral stems and modular heads to reinstate function following the degenerative effects of osteo and rheumatoid arthritis, post-trauma disease effects, avascular necrosis and septic or aseptic total hip revision.

The acetabular cups are manufactured from cobalt chrome alloy. The cobalt chrome plasma spray coating provides an increased surface roughness to act as a key for mechanical interlock with PMMA bone cement.

## Summary of substantial equivalence and safety and effectiveness

Corin claim that the range of C-Fit and DC-Fit Cups are substantially equivalent to the Mallory Head Acetabular Cup range which has been promoted in the USA by Biomet Inc since 1990. The reasons for this are summarised below.

- Both are modular cups manufactured from cobalt chrome alloy and utilise UHMWPE a) liners.
- b) Both are coated with a plasma spray of the same alloy.
- Both are designed for cemented use and for use with screws. c)
- Both are available in a range of sizes from 42mm 70mm and can accommodate a d) range of femoral head sizes.

This submission is supported by extensive mechanical test data confirming the devices' safety and effectiveness. The plasma coated C-Fit and DC-Fit Acetabular Cups have been in clinical use outside the USA for ten years and two years respectively and no significant postoperative problems have been reported.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NON 10 1997

Mr. Craiq Corrance ·President Corin U.S.A. 10500 University Center Drive, Suite 130 Tampa, Florida 33612

Re : K972995 The C-Fit and DC-Fit Acetabular Cup System Trade Name: Regulatory Class: II Product Code: JDI Dated: July 21, 1997 ¯ Received: August 12, 1997

Dear Mr. Corrance:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you miqht have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{2}------------------------------------------------

Page 2 - Mr. Craig Corrance

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contribution of the contribution of th

**510(k) Number (if known)**

The C-Fit & DC-Fit Acetabular Cup System______________________________________________________________________________________________________________________________________ Device Name: _________________________________________________________________________________________________________________________________________________________________

## INDICATIONS FOR USE

Relief of pain and restoration of hip function following the effects of osteo, rheumatoid and Netier of pain and restoration of mip state effects, avascular necrosis, and total hip revision.

The C-Fit and DC-Fit plasma sprayed acetabular components are intended for use QNLY with bone cement.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation

| Prescription Use                        | X       |
|-----------------------------------------|---------|
| (per 21 CFR 801.109)                    |         |
| OR                                      |         |
| Over The-Counter Use                    |         |
| (Division Sign-Off)                     |         |
| Division of General Restorative Devices |         |
| 510(k) Number                           | K972995 |

---

**Source:** [https://fda-staging.innolitics.com/device/K972995](https://fda-staging.innolitics.com/device/K972995)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com