PALAUREX 78

K972954 · Aurex Precious Metal Industries (Pty)Ltd · EJT · Sep 5, 1997 · Dental

Device Facts

Record IDK972954
Device NamePALAUREX 78
ApplicantAurex Precious Metal Industries (Pty)Ltd
Product CodeEJT · Dental
Decision DateSep 5, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3060
Device ClassClass 2

Indications for Use

Dental casting alloy used in combination with dental ceramics for fabrication of metallo-ceramic restoration.

Device Story

Palaurex 78 is a low gold, palladium-based dental casting alloy. It is intended for use in the fabrication of metallo-ceramic restorations. The alloy is processed by dental technicians in a laboratory setting to create dental prosthetics. The device provides a structural metal framework for porcelain veneering. By utilizing high noble metal content, the alloy is designed to provide biocompatibility and mechanical stability for dental restorations, aiding in the functional and aesthetic rehabilitation of patients requiring dental crowns or bridges.

Clinical Evidence

No clinical data. Bench testing only, conducted in accordance with ANSI/ADA 5 and ISO 9693 standards to evaluate physical and mechanical properties including melting point, hardness, yield strength, elongation, CTE, and density.

Technological Characteristics

Low gold, palladium-based casting alloy. Composition: Au (2%), Pd (77.8%), Cu (13%), Ga (7%), Ru (0.2%). Physical properties: Melting range 1150-1190°C, Vickers hardness 330-380, Yield strength 780-850 MPa, CTE 14.3 x 10^-6/°C. Tested per ANSI/ADA 5 and ISO 9693.

Indications for Use

Indicated for use in the fabrication of metallo-ceramic dental restorations in combination with dental ceramics.

Regulatory Classification

Identification

A noble metal alloy is a device composed primarily of noble metals, such as gold, palladium, platinum, or silver, that is intended for use in the fabrication of cast or porcelain-fused-to-metal crown and bridge restorations.

Special Controls

*Classification.* Class II (special controls). The special control for these devices is FDA's “Class II Special Controls Guidance Document: Dental Noble Metal Alloys.” The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. See § 872.1(e) for availability of guidance information.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo for "AUREX". The logo consists of a stylized graphic above the word "AUREX". The graphic appears to be composed of three overlapping triangles, creating a complex geometric shape. The logo is black and white. P.O. Box 509 Edenvale 1610 Republic of South Africa Telephone : +27 (11) 609-8640 Telefax : +27 (11) 452-3918 SEP - 5 1997 -18- ## 510 (k) Summary Trade name: Common name: Classification name: Class ification number: PALAUREX 78 Dental casting alloy Gold based alloys and precious metal alloys for clinical use EJT Legally marketed device: Jelenko Freedom Plus Description of the device: Low gold palladium based casting alloy Intended use of the device: Type IV restoration with porcelain veneering ## Summary of the technological characteristics Test methods applied: as in ANSI/ADA 5 and ISO 9693 Comparison of composition: | ALLOY | | COMPOSITION (weight%) | | | | | | |-------|--------------|-----------------------|------|----|----|----|-----| | | Name | Au | Pd | Cu | In | Ga | Ru | | Legal | Freedom Plus | 2 | 78.4 | 8 | 6 | 5 | x | | New | Palaurex 78 | 2 | 77.8 | 13 | | 7 | 0.2 | x is less than 1 % Comparison of physical and mechanical properties | ALLOY | Melting point range (°C) | | Hardness (Vickers 5/30) | | Yield strength (MPa) | | Elongation (%) | | CTE (x10-6/°C) | Density (g/cm3) | |--------------------|--------------------------|--------|-------------------------|------|----------------------|------|----------------|------|----------------|-----------------| | Name | solid. | liquid | soft | hard | soft | hard | soft | hard | | | | Legal Freedom Plus | 1170 | 1300 | 265 | | 565 | | 30 | | 13.9 | 10.8 | | New Palaurex 78 | 1150 | 1190 | 330 | 380 | 780 | 850 | 12 | 2 | 14.3 | 10.6 | ## Discussion The noble content metal is high (80%) in both alloy, with almost identical breakdown to the respective elements. Palaurex 78 has two base metal elements, Cu and Ga: Freedom Plus has In as well. ## Conclusion On the basis of the identical major elements and high noble metal content similar biological effect may be expected from both alloys. TÜV CERT Aurex Precious Metal Industries (Pty) Limited (Reg.No. 88/04891/07) Mone .. Samore and Rotinoro of Denning Motors for Dontal Toustlars {1}------------------------------------------------ Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the eagle in a circular fashion. SEP - 5 1997 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Jerome D. Davis Managing Director Aurex Precious Metal Industries (PTY)-LTD. P.O. Box 509 ... Edenvale 1610 Republic of South Africa K972954 Re: Palaurex 78 Trade Name: Requlatory Class: II EJT Product Code: Dated: June 13, 1997 Received: June 20, 1997 Dear Mr. Davis: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: {2}------------------------------------------------ Page 2 - Mr. Davis - - this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours Timothy A. Ulatowski Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): D Device Name: PALAUREX 78 Indications For Use: Dental casting alloy used in combination with dental ceramics for fabrication of metallo-ceramic restoration. Image /page/3/Picture/5 description: The image shows a solid black circle. The circle is slightly irregular in shape, with a few minor imperfections along its perimeter. The background is white, providing a stark contrast to the black circle. The image is simple and minimalist, focusing solely on the shape and color of the circle. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Supper Rioner (Division Sign-Off) Division of Dental, Infection Control, and General Hospita! Devices 510(k) Numbe Prescription Use Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...