# SUPER-DENT INTERMEDIATE RESTORATIVE MATERIAL (K972935)

_S.S. White Mfg., Ltd. · EMA · Jan 29, 1998 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K972935

## Device Facts

- **Applicant:** S.S. White Mfg., Ltd.
- **Product Code:** [EMA](/productcode/EMA.md)
- **Decision Date:** Jan 29, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

These materials are intended for use in dentistry as a base or cavity liner under permanent restorations and for temporary restorations to teeth.

## Device Story

Super-Dent Intermediate Restorative Material is a dental cement consisting of a powder and liquid component. When mixed, these components form a setting mass used by dental professionals in a clinical setting. The material functions as a base or cavity liner beneath permanent restorations or as a temporary restorative material to protect tooth cavities. The device relies on a chemical reaction between zinc oxide and eugenol, modified with polymers to improve strength and abrasion resistance. The primary clinical benefit is the ease of mixing, which facilitates a consistent, homogeneous material for reliable performance during dental procedures.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Zinc oxide and eugenol (4-Allyl-2-methoxyphenol) based cement. Chemically modified and reinforced with polymers for enhanced strength and abrasion resistance. Supplied as a powder and liquid for manual mixing.

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- Temporary Dental Cement ([K895487](/device/K895487.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a stylized, bold, and slanted text that reads "DWITE". The letters are thick and appear to be in a sans-serif font. The overall impression is one of a logo or a heading, possibly for a brand or organization named DWITE.

# K972935 TAN 29, 1998 Page 18 OF 31

510 (k) 80

S.S. White Manufacturing IJAME :

Unit 9, Madleaze Estate, ADDRESS : Bristol Road, Gloucester, GL1 SSG England

01452 307171 PHONE No :

01452 307187 FAX No. :

Hugh Savell CONTACT : nugh Baver

DATE PREPARED: 4" August, 1997

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo or wordmark that reads "SWTE" in bold, stylized letters. The letters are black and appear to be slightly slanted or italicized, giving the design a dynamic feel. The font is sans-serif and the letters are closely spaced, creating a compact and impactful visual.

ప్రా

### PREMARKET NOTIFICATION

510(k) SUMMARY (continued)

| TRADE NAME:          | Super-Dent Intermediate Restorative<br>Material   |
|----------------------|---------------------------------------------------|
| COMMON NAME:         | Temporary Dental Cement                           |
| CLASSIFICATION NAME: | Cement, Dental                                    |
| EQUIVALENT TO:       | Temporary Dental Cement; 510(k) number<br>K895487 |

### DESCRIPTION :

These materials consist of a powder and liquid which, when mixed together, form a setting mass suitable for their intended use.

# INTENDED USE:

These materials are intended for use in dentistry as a base or cavity liner under permanent restorations and for temporary restorations to teeth.

# TECHNOLOGICAL CHARACTERISTICS:

These materials rely on the reaction between zinc oxide and eugenol (4-Allyl-2-methoxyphenol) to form a setting mass. They are chemically modified to control the setting reaction and are reenforced with polymers to give enhanced strength and a surface more resistant to abrasion.

{2}------------------------------------------------

SWTE

JAN 2 9 19

Page 20 OF 31

**510 (k) SUMMARY (continued)**

## NON-CLINICAL DATA:

The setting characteristics of these materials are essentially identical at the stipulated mixing ratios. The compressive strength of the subject device is slightly lower and the disintegration slightly higher than the predicate device.

The subject device is subjectively very much easier to mix which is an important advantage in surgery use.

# CLINICAL DATA:

There is none applicable.

# CONCLUSIONS :

The data summarised above indicates that the subject device is substantially equivalent to the predicate device.

The slightly lower performance of the subject device is considered to be out-weighed by its easier mixing. This ensures a consistently homogenous mix and thereby more reliable performance in clinical use.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows a logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around a stylized graphic. The graphic consists of three curved lines that resemble a bird in flight. The logo is black and white and appears to be of a government organization.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Mr. Hugh Savell Technical and Compliance Manager S.S. White Manufacturing, Ltd. 9 Madleaze Estate, Bristol Road Gloucester, Gl1 5SG. England

Re : K972935 Super-Dent Intermediate Restorative Material Trade Name: Requlatory Class: II Product Code: EMA Dated: December 16, 1997 December 22, 1997 Received:

Dear Mr. Savell:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does

{4}------------------------------------------------

Page 2 - Mr. Savell

not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Tim M. Wlatucki

Timothy A. Ulatowski Directpr Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

ರಿ

Enclosure

{5}------------------------------------------------

3

| 510(k) Number (if known): | K972935                             |
|---------------------------|-------------------------------------|
| Device Name:              | SUPER-DENT INTERMEDIATE RESTORATIVE |
| Indications For Use:      | MATERIAL                            |

This material is intended for use in dentistry as a base or cavity liner under permanent restorations and for temporary restorations to teeth as an interim measure to protect cavities before a permanent restoration can be completed.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

|               | Susan Runver                                                        |
|---------------|---------------------------------------------------------------------|
|               | (Division Sign-Off)                                                 |
|               | Division of Dental, Infection Control, and General Hospital Devices |
| 510(k) Number | K972935                                                             |

| Prescription Use     |  | OR | Over-The-Counter Use     |
|----------------------|--|----|--------------------------|
| (Per 21 CFR 801.109) |  |    | (Optional Format 1-2-96) |

---

**Source:** [https://fda-staging.innolitics.com/device/K972935](https://fda-staging.innolitics.com/device/K972935)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com