K972933 · J-Mac System, Inc. · LLZ · Oct 27, 1997 · Radiology
Device Facts
Record ID
K972933
Device Name
VOX/BASE FOR WINDOWS
Applicant
J-Mac System, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 27, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The device is intended to display various image data used in a hospital, clinic, physician's office and other health care settings. It handles data from CR, CT, DS, MR, NM, US, ES, RG and other modalities. The J-MAC system is an accessory to diagnostic imaging devices, it is not a diagnostic device. The J-MAC VOX/BASE System User Manual provides warnings and precautions regarding intended use, and detailed information on system operation.
Device Story
J-MAC VOX/BASE is a standalone PACS software for Windows; enables medical professionals to view, retrieve, store, import, process, and transmit medical images via networks or phone lines. Input: medical images from CR, CT, DS, MR, NM, US, ES, RG modalities. Operation: four modules (Query, Receive, View, Manager) utilize WinSock library and TCP/IP protocol for DICOM service communication. View module provides basic image processing (flip, rotate, invert), screen layout tools, and image management (zoom, select, hide, scroll). Output: displayed images on workstation monitor. Used in hospitals, clinics, and physician offices. Healthcare providers use output for image review and management; device acts as an accessory to diagnostic imaging devices, not a diagnostic device itself.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Standalone software application for Microsoft Windows (95, NT 3.51, 4.0+). Modular architecture (Query, Receive, View, Manager). Connectivity via TCP/IP network protocol using WinSock library. Compliant with DICOM 3.0 and ACR-NEMA 2.0 standards. Image processing capabilities include basic manipulation (flip, rotate, invert) and display management.
Indications for Use
Indicated for medical professionals in hospitals, clinics, and physician offices to view, retrieve, store, import, process, and transmit medical images from CR, CT, DS, MR, NM, US, ES, and RG modalities.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
J-MAC VOX/ BASE System 510(k) Notification
# OCT 27 1997
#### 510(k) Summary
## K972933
#### J-MAC VOX/BASE System for Windows
Common classification name: Digital Image Communications System (PACS)
J-MAC Systems, Inc. Oda Building 10-1 Kita 9 Nishi 3 Kita-ku, Sapporo, 060, Japan Phone: 81-11-726-6262 81-11-726-6260 Fax:
Contact Person:
Sheila W. Pickering Ph.D. 320 Fairway Drive Half Moon Bay, California 94019 Telephone/Fax 415 726 3223
### A. Legally Marketed Predicate Device
The J-MAC VOX/BASE for Windows is substantially equivalent to the RadWorks Medical Imaging Software (K962699) manufactured by Applicare Imaging, Inc, as well as other PACS systems. The J-MAC device is substantially equivalent to the RadWorks predicate device with regard to device features and specifications, as well as intended use. Both devices are stand alone PACS software systems with similar operating requirements, and both comply with DICOM 3.0 and ACR-NEMA 2.0. Both are Windows based applications. Both products are intended for use with Computed Tomography (CT), Magnetic Resonance Imaging (MRJ), Computed Radiography (CR), Digital Subtraction Angiography (DS), Nuclear Medicine (NM), Ultrasound (US), Endoscopy (ES), Radiographic imaging (RG), and other modalities.
,
### B. Device Description
The J-MAC VOX/BASE for Windows is a stand alone software product which allows medical professionals to view, retrieve, store, import, process and transmit medical images over networks or phone lines. This type of product is commonly referred to as a Picture Archival and Communication System (PACS).
The software consists of four modules: Query, Receive, View and Manager. The function of each is as follows:
- 1. Query This module implements the Query and Retrieve DICOM Services as a Service Class User. It calls functions in the WinSock library to connect to a remote Image Server using the TCP/IP network protocol.
{1}------------------------------------------------
- 2. Receive This module implements the Storage DICOM Services as a Service Class provider. It calls functions in the WinSock library to accept connections from a remote Service Class User using the TCP/IP network protocol.
- 3. View this module contains tools for
- basic image processing (flipping, rotating, inverting) .
- screen layout (cell dimensions and arrangement, colors, font) .
- image management (magnifying lens, zoom, select, hide, scroll) ●
- 4. Manager This module provides a convenient way of knowing which images are already stored locally.
## C. Intended Use
The device is intended to display various image data used in a hospital, clinic, physician's office and other health care settings. It handles data from CR, CT, DS, MR, NM, US, ES, RG and other modalities. The J-MAC system is an accessory to diagnostic imaging devices, it is not a diagnostic device. The J-MAC VOX/BASE System User Manual provides warnings and precautions regarding intended use, and detailed information on system operation.
{2}------------------------------------------------
ﻳﻮ
J-MAC VOX/ BASE System 510(k) Notification
## D. Substantial Equivalence
| | Predicate Device | Submission Device | |
|------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------|
| Product Name | RadWorks Medical Imaging Software<br>K962699 | J-MAC VOX/Base/Windows<br>Medical Imaging Software | Substantially<br>Equivalent |
| Intended Users | Medical professionals working in image<br>analysis and management | Medical professionals working in image<br>analysis and management | yes |
| Site of Use | Hospitals, clinics, remote health care<br>facilities | Hospitals, clinics, remote health care<br>facilities | yes |
| Languages | English | Japanese and English | yes |
| Software structure | Modular design | Modular design (4 modules: Query, Receive, View, Manager) | yes |
| Compliance with<br>standards | Complies with DICOM 3.0 and<br>ACR-NEMA 1.0 or 2.0 or proprietary | Complies with DICOM 3.0 and<br>ACR-NEMA 1.0 or 2.0 or proprietary | yes |
| Operating System | Microsoft Windows NT | Microsoft Windows 95, Windows NT3 3.51, 4.0 or later | yes |
◆
【
ﻤﺴﺴﺴﺴﺴﺴ
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings outstretched, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 27 1997
Sheila W. Pickering, Ph.D. J-MAC Systems, Inc. c/o Wilson Sonsini Goodrich & Rosati 320 Fairway Drive Half Moon Bay, CA 94019
Re:
K972933
VOX/BASE™ for Windows Dated: August 5, 1997 Received: August 8, 1997 Regulatory class: Unclassified Procode: 90 LLZ
Dear Dr. Pickering:
We have reviewed your Section 5100k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranion.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
h.T. Kiau Yui
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a series of handwritten characters. The characters appear to be numbers and possibly letters. The characters are written in black ink on a white background and are somewhat stylized.
J-MACVON BASF. 510(k) Notification
#### FDA Submission Cover Sheet
510(k) Number (if known): Not applicable
Device Name J-MAC VOX/BASE for Windows
Indications For Use:
The J-MAC VOX/BASE for Windows is a stand alone software product which allows medical professionals to view, retrieve, store, import, process and transmit medical images over networks or phone lines. This type of product is commonly referred to as a Picture Archival and Communication System (PACS). It is intended to display various image data used in a hospital, clinic, physician's office and other health care settings. It handles data from CR, CT, DS, MR, NM, US, ES, RG and other modalities.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrance Of CDRH, Office Of Device Evaluation (ODE)
| Prescription Use | <span style="text-decoration: overline;">✓</span> | OR | Over-The-Counter Use | <span style="text-decoration: overline;"></span> |
|---------------------|---------------------------------------------------|----|----------------------|--------------------------------------------------|
| (Per 21CFR 801.109) | | | | |
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
| 510(k) Number | K972933 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.