The Partnership Revision Femoral Stems consist of a family of Titanium femoral stems with Titanium plasma spray coating that are intended to be used with Howmedica V40™ femoral heads, Howmedica Unipolar and Bipolar components, and Howmedica acetabular components in primary and revision total hip arthroplasty. These femoral stems are designed to be press fit into the proximal femur. They do not achieve fixation by biological ingrowth.
Device Story
Partnership Revision Femoral Components are titanium femoral stems with titanium plasma spray coating; intended for use in primary and revision total hip arthroplasty. Stems are press-fit into proximal femur; do not achieve biological ingrowth fixation. Used with Howmedica V40 femoral heads, unipolar/bipolar components, and acetabular components. Device provides structural support for hip joint replacement. Surgeons perform implantation in clinical/OR settings. Clinical benefit includes restoration of hip function in patients with degenerative joint disease. FDA requires postmarket surveillance due to potential for coating delamination, spalling, or metal debris generation associated with plasma spray coatings compared to sintered coatings; surveillance monitors long-term survivorship and revision rates over nine years.
Clinical Evidence
No clinical data provided in the submission. Bench testing only; fatigue load carrying capacity testing demonstrated performance exceeding minimum ISO load requirements.
Technological Characteristics
Materials: Ti-6Al-4V (ASTM F 136) or TMZF alloy (Ti-11.5 Mo-6Zr-2 Fe, draft ASTM F 1813). Coating: CP titanium plasma spray (ASTM F 1580). Design: Press-fit femoral stem. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for patients undergoing primary or revision total hip arthroplasty requiring a femoral stem component. Designed for press-fit fixation in the proximal femur; not indicated for biological ingrowth fixation.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
K9 7 2x93
NOV - 3 1997
## 510(k) Summary
Proprietary Name: Partnership Revision Femoral Components
Common Name: Hip Prosthesis
Classification Name and Reference: 21 CFR 888.3353 Hip Joint Metal/Ceramic/Polymer semi-constrained cemented or nonporous uncemented prosthesis.
Proposed Regulatory Class: Class II
Device Product Code: LZO
For information contact: Frank Maas Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford, NJ 07070 Telephone: (201) 507-7875 (201) 507-6870 Fax: Date Summary Prepared: 8-1-97
The Partnership Revision Femoral Components consist of a family of Titanium femoral stems with Titanium plasma spray coating. The stems are intended to be used with Howmedica V40™ femoral heads, Howmedica Unipolar and Bipolar components, and Howmedica acetabular components in primary and revision total hip arthroplasty. These femoral stems are designed to be press fit into the proximal femur. They do not achieve fixation by biological ingrowth.
These femoral components will be made available in two substrate alloys: 1) Ti-6Al-4V, which meets the requirements of ASTM specification F 136; or 2) TMZF alloy (Ti-11.5 Mo-6Zr-2 Fe). The TMZF alloy is the subject of draft ASTM specification F 1813. The Titanium plasma spray coating is CP titanium, which conforms to ASTM specification F 1580.
The substantial equivalence of the Partnership Revision Femoral stems is based on an equivalence in intended use, materials, design, and relative indications and contraindications to Howmedica's Meridian® Femoral Stem (K940307), Osteolock™ CL Femoral Stem. TMZF (K944592). Osteolock™ CL Femoral Stem, Ti-6Al-4V (K941141), Osteonics' Restoration Monolithic II Slotted Hip Stems (K951671), and Zimmer's Multilock Hip Prosthesis (K921308).
Testing has demonstrated that the fatigue load carrying capacity of the Partnership Revision Femoral stems exceeds the minimum ISO load requirements.
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized image of three wavy lines, which are meant to represent the human form.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 3 1997
Mr. Frank Maas Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard 07070 Rutherford, New Jersey
Re : K972893 Partnership Revision Femoral Components Requlatory Class: 2 Product Codes: LZO and LPH Dated: August 4, 1997 Received: August 5, 1997
Dear Mr. Maas:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affectinq your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does
{2}------------------------------------------------
Page 2 - Mr. Frank Maas
not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
Under Section 522(a) of the act, manufacturers of certain types of devices identified by the Act or designated by FDA are required to conduct postmarket surveillance studies. FDA has identified under Section 522(a)(1)(C) the device cleared for marketing by this letter as requiring postmarket surveillance. The rationale for this decision is contained in the enclosed attachment.
Within thirty (30) days of first introduction or delivery for introduction of this device into interstate commerce you are required to submit to FDA certification of the date of introduction into interstate commerce, a detailed protocol which describes the postmarket surveillance study, and a detailed profile of the study's principal investigator that clearly establishes the qualifications and experience of the individual to conduct the proposed study. For your information, general guidance on preparing a protocol for a postmarket surveillance study is attached.
Submit five (5) copies to:
Center for Devices and Radiological Health Postmarket Surveillance Studies Document Center Room 3083 (HFZ-544) 1350 Piccard Drive Rockville, Maryland 20850
Within sixty (60) days of receipt of your protocol, FDA will either approve or disapprove it and notify you of the Agency's action in writing. You should not begin your postmarket surveillance study of this device until the protocol has been Data generated under an unapproved protocol may not approved. satisfy your obligation under section 522. Please note that you must continue to collect and report data needed to maintain compliance with Medical Device Reporting regulations (21 CFR 803) .
Failure to certify accurately the date of initial introduction of your device into interstate commerce, to submit timely an acceptable protocol, or to undertake and complete an FDA approved postmarket surveillance study consistent with the protocol will be considered violations of section 522. In accordance with the Medical Device Amendments of 1992, failure of a manufacturer to meet its obligations under section 522 is a prohibited act under section 301 (q) (1) (C) of the Act (21 U.S.C. 331 (q) (1) (C) . Further, under section 502 (t) (3) of the
{3}------------------------------------------------
Page 3 - Mr. Frank Maas
act (21 U.S.C. 352(t)(3)), a device is misbranded if there is a failure or refusal to comply with any requirement under section 522 of the act. Violations of sections 301 or 502 may lead to regulatory actions including seizure of your product, injunction, prosecution, or civil money penalties.
If you have questions concerning postmarket surveillance study requirements, contact the Postmarket Surveillance Studies Branch at (301) 594-0639.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address
"http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
f
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosures
{4}------------------------------------------------
## RATIONALE FOR POSTMARKET SURVEILLANCE REQUIREMENTS FOR PLASMA SPRAYED POROUS COATED HIP SYSTEMS
The Center for Devices and Radiological Health (CDRH) is requiring that manufacturers of plasma sprayed porous coated hip prostheses cleared through the 510(k) process whose submissions lack nine year cloared checagi one trating the long-term safety and effectiveness of the device, conduct postmarket surveillance, pursuant to section 522(a) (1) (C) of the Food, Drug, and Cosmetic Act (Act). This surveillance is required for the reasons discussed below.
Long-term (i.e., nine years and greater) clinical investigations of Long cerm (1707) firm Joof sintered porous coatings made from cobaltchrome beads or titanium fiber mesh have provided a reasonable assurance of the safety and effectiveness of hip prostheses with these coatings, (See 58 FR 3227, Jan. 8, 1993). In bench testing, plasma sprayed porous coatings have demonstrated lower mechanical strength and, in particular, lower resistance to abrasion than that ynoted for sintered cobalt-chrome bead or sintered titanium fiber mesh coatings. Thus, while short-term (i.e., less than four years) clinical data uncovers no difference in the clinical failure rates between plasma spray porous coated hip prostheses and hip systems with sintered porous coatings, because of the lack of long-term clinical safety and effectiveness information, CDRH believes that hips covered by plasma sprayed porous coatings should be monitored to insure that they are not prematurely failing due to metal debris, coating spalling, or coating delamination. These concerns, if realized, could result in significant incapacitation and require revision surgery.
The Food and Drug Administration (FDA) will determine if your protocol will result in the collection of useful data or other information necessary to protect the public health and to provide safety and effectiveness information for the device. We anticipate that an acceptable protocol will provide an actuarial survivorship analysis over a nine year period by monitoring what proportion of hips are revised in a representative sample of several hundred patients in whom the device was implanted, without bone cement, for rehabilitation of a hip damaqed by noninflammatory degenerative joint disease.
Stratification of the data may be necessary according to revision of the acetabular component, the femoral component, or both components. It is appropriate to further judge clinical success by obtaining patient self-assessment of how well the hip prosthesis is functioning. You will also need to collect data on patient deaths, losses to follow-up, and baseline variables which could potentially confound the survivorship data (e.g., device configuration of stem and cup models, diagnosis, and patient condition prior to initial implantation as judged by Harris Hip Score). Such data should function as an early warning system for late occurring effectiveness and safety problems.
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Partnership Revision Femoral Stems
Indications for Use:
The Partnership Revision Femoral Stems consist of a family of Titanium femoral stems with Titanium plasma spray coating that are intended to be used with Howmedica V40™ femoral heads, Howmedica Unipolar and Bipolar components, and Howmedica acetabular components in primary and revision total hip arthroplasty. These femoral stems are designed to be press fit into the proximal femur. They do not achieve fixation by biological ingrowth.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
(Division Sign-Off) Division of General Restorative Devices 510(k) Number_K972893
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.