K972832 · Depuy, Inc. · FYA · Oct 21, 1997 · General, Plastic Surgery
Device Facts
Record ID
K972832
Device Name
PROVISION SURGICAL HELMET SYSTEM
Applicant
Depuy, Inc.
Product Code
FYA · General, Plastic Surgery
Decision Date
Oct 21, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Pro Vision Surgical Helmet System consists of disposable surgical apparel and re-usable headgear, battery holder and battery pack. The Pro Vision Surgical Helmet System is intended to be worn by healthcare workers in the operating room during surgical procedures to protect both the healthcare worker and the surgical patient from the transfer of microorganisms, body fluids and particulate material.
Device Story
System comprises disposable surgical apparel (hoods/gowns) and reusable headgear with battery-powered air delivery; used by healthcare workers in operating rooms. Air delivery system pulls air through filter media and directs it to wearer's face/neck area; provides barrier protection against fluid-borne pathogens while maintaining breathability. Hytrel elastomer-coated material provides fluid resistance. Protects both wearer and patient from cross-contamination of microorganisms and particulates.
Clinical Evidence
Bench testing only. Material testing performed per ASTM ES-21 (resistance to synthetic blood) and ASTM ES-22 (resistance to viral penetration).
Technological Characteristics
Components: Reusable helmet, battery pack (nickel metal hydride), disposable hoods/gowns. Materials: Hytrel elastomer-coated fabric, molded polycarbonate face shield. Air delivery: Active filtration system. Standards: ASTM ES-21, ASTM ES-22.
Indications for Use
Indicated for healthcare workers in the operating room to protect the worker and patient from transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
DePuy Sterile View Surgical Exhaust System
Sterile View Disposable Barrier Hood/Gowns
DePuy Disposable Filter/Hood
Stackhouse FreedomAire Surgical Helmet System
Stryker Steri Shield Turbo III
Submission Summary (Full Text)
{0}------------------------------------------------
K972832
### SUMMARY OF SAFETY AND EFFECTIVENESS
OCT 21 1997
| NAME OF FIRM: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, In 46581-0988 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) CONTACT: | Cheryl K. Hastings<br>Manager, Clinical Affairs |
| TRADE NAME: | ProVision Surgical Helmet System |
| CLASSIFICATION: | 878.4040 Surgical apparel |
| DEVICE PRODUCT CODE: | 79 FYA |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES: | ⌀ DePuy Sterile View Surgical Exhaust System<br>⌀ Sterile View Disposable Barrier Hood/Gowns<br>⌀ DePuy Disposable Filter/Hood<br>⌀ Stackhouse FreedomAire Surgical Helmet System<br>⌀ Stryker Steri Shield Turbo III |
#### INTENDED USE AND DEVICE DESCRIPTION:
The ProVision Surgical Helmet System consists of disposable surgical apparel and re-usable headgear, battery holder and battery pack. The Pro Vision Surgical Helmet System is intended to be worn by healthcare workers in the operating room during surgical procedures to protect both the healthcare worker and the surgical patient from the transfer of microorganisms, body fluids and particulate material.
The disposable, sterile surgical apparel of the ProVision system include a barrier hood/gown with Hytrel® elastomer, a barrier hood/gown toga with Hytrel, a barrier hood with Hytrel and a paper hood. A molded, polycarbonate face shield is sealed to the hood material. The Hytrel coated material is fluid resistant but vapor transmissive. The gown, toga and hood with Hytrel therefore protect the wearer from fluid borne pathogens, while maintaining "breathability". The barrier hood/gown and barrier hood/gown toga each come in three sizes. Testing has shown that the Hytrel coated material of the barrier hood/gown and barrier hood/gown toga passes ASTM standard test ES-21 for resistance to synthetic blood and ES-22 for resistance to penetration by blood borne pathogens using viral penetration as a test system.
The non-sterile, re-usable component of the Pro Vision system is a two piece unit comprised of a helmet and a battery pack. The helmet is a headgear supporting an air delivery system. The air delivery system pulls air from outside the hood or hood/gown, through a filter media, and directs
# 000004
{1}------------------------------------------------
it to the wearer's face/neck area. The air delivery system is powered by a nickel metal hydride battery pack.
#### BASIS OF SUBSTANTIAL EQUIVALENCE:
The ProVision Surgical Helmet System is similar in design and identical in use to the DePuy Sterile View Surgical Exhaust System combined with the Sterile View Barrier Hood/Gowns. The system has been modified to make it more comfortable and compact, lighter, cooler and quieter while maintaining adequate filtration efficiency of the ProVision Surgical Helmet System is comparable to that of the DePuy Disposable Filter Hood.
The Pro Vision Surgical Helmet System is also similar in design and intended use to the Stackhouse FreedomAire Surgical Helmet System and the Stryker Steri Shield Turbo III.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
OCT 21 1997
Ms. Cheryl K. Hastings Manager, Clinical Affairs DePuy, Incorporated 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
K972832 Re : Provision Surgical Helmet System Trade Name: Requlatory Class: II Product Code: FYA July 29, 1997 Dated: Received: July 30, 1997
Dear Ms. Hastings:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{3}------------------------------------------------
Page 2 - Ms. Hastings
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be -------------------------------------------------------------------------------------------------------------------obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Timothy A. Ulatows Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
. Enclosure
{4}------------------------------------------------
510(k) Number (if known) Қ972832
ProVision Surgical Helmet System Device Name
Indications for Use:
The Pro Vision Surgical Helmet System consists of disposable surgical apparel and re-usable headgear, battery holder and battery pack. The Pro Vision Surgical Helmet System is intended to be worn by healthcare workers in the operating room during surgical protect both the healthcare worker and the surgical patient from the transfer of microorganisms, body fluids and particulate material.
Concurrence of CDRH, Office of Device Evaluation
Division Sign Off
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital 510(k) Number
Prescription Use OR
(Per 21 CFR 801.109)
Over-The-Counter Use
## 0000001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.