The Exeter All Plastic Acetabular Component (previously cleared in K936132) is intended to be used for primary and secondary reconstruction of the bearing surface of the acetabulum as a result of painful and/or severly disabled hip joints due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis or revision of a failed acetabular component. The Howmedica® Bipolar Prosthesis, Centrax® (previously cleared in K855231) is intended to be used in conjunction with a femoral stem component for the reconstruction of a femoral head damaged by fresh fracture, non-union, aseptic necrosis of the head and neck, or by osteoarthritis or post-traumatic arthritis.
Device Story
Duration® Stabilized UHMWPE acetabular components are orthopedic implants designed for hip arthroplasty. The device utilizes an ultra-high molecular weight polyethylene (UHMWPE) material subjected to a proprietary 'Duration® Stabilized' process, which serves as an alternative to standard air-irradiated gamma sterilization. The device is intended for use by orthopedic surgeons in clinical settings for hip joint reconstruction. The primary clinical benefit is a reduction in volumetric wear compared to conventionally gamma-sterilized components, as demonstrated in in vitro hip joint simulator testing. The device is used in conjunction with femoral stem components and CoCr articulating counterfaces. Surgeons utilize these components to restore joint function and alleviate pain in patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data provided. Evidence consists of bench-only in vitro testing using a multiaxial hip joint wear simulator. Testing involved over 5 to 10 million cycles using CoCr heads and bovine calf serum lubricant. Results showed a 27% to 40% reduction in volumetric wear compared to conventionally gamma-sterilized controls. The document explicitly states that in vitro test results have not been shown to correlate with clinical wear mechanisms.
Technological Characteristics
Material: Duration® Stabilized UHMWPE. Components: Exeter All Plastic Acetabular Component and Centrax® Bipolar Prosthesis. Sterilization: Duration® Stabilized process (alternative to air-irradiated gamma). Testing: Multiaxial hip joint wear simulator, CoCr articulating counterfaces, bovine calf serum lubricant.
Indications for Use
Indicated for patients requiring hip reconstruction due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, revision of failed acetabular components, femoral head fracture, non-union, or aseptic necrosis of the femoral head and neck.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
510(k) Summary
K972792
Proprietary Name: Duration® Stabilized UHMWPE Acetabular Components
Common Name: UHMWPE Acetabular Components
21 CFR 888.3358 Classification Name and Reference:
Proposed Requlatory Class: Class II Device Product Code: JDI
:
For information contact: Frank Maas Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford, NJ 07070 Telephone: (201) 507-7875 Fax: (201) 507-6870 Date Summary Prepared: 7-25-97
This submission describes additional UHMWPE acetabular components that can be sterilized by the Duration® Stabilized process previously cleared in submission K934060, as an alternate to standard air irradiated gamma sterilization. The additional components are the Exeter All Plastic Acetabular Component (K936132) and the Howmedica® Bipolar Prosthesis, Centrax® (K855231). Corresponding wear claims, previously cleared in submission K 963612 and listed below, are also applicable to these additional components.
Howmedica's Duration® Stabilized UHMWPE, System 12®, size P4, neutral acetabular inserts, with an inner diameter of 32mm, a bearing thickness of 6.1 mm and aged for 6 months in an oxygen environment, showed a 36% reduction in volumetric wear versus the same cup conventionally gamma sterilized and aged for 6 months in an oxygen environment. Testing was performed in a multiaxial hip joint simulator for over 10 million cycles, using a 32mm CoCr Head articulating counterface and bovine calf serum as a lubricant. The results of in vitro tests have not been shown to correlate with clinical wear mechanisms.
!
4
{1}------------------------------------------------
Howmedica's Duration® Stabilized UHMWPE, a hemispherical cup of generic design for hip simulator testing, with an inner diameter of 32 mm, and a bearing thickness of 8.4 mm, showed a 40% reduction in volumetric wear versus the same cup conventionally gamma sterilized. Testing was performed in a multiaxial hip joint wear simulator for over 5 million cycles, using a 32 mm diameter CoCr Head articulating counterface and tow calcium" containing bovine calf serum as a lubricant. The results of in vitro tests have not been shown to correlate with clinical wear mechanisms.
Howmedica's Duration® Stabilized UHMWPE, System 12® Neutral Acetabular Insert, size P4. with an inner diameter of 32mm, and a bearing thickness of 6.1 mm, showed a 40% reduction in volumetric wear versus the same cup conventionally gamma sterilized. Testing was performed in a hip joint wear simulator for over 5 million cycles, using a 32 mm CoCr Head articulating counterface and high calcium" containing bovine calf serum as a lubricant. The results of these in vitro tests have not been shown to correlate with clinical wear mechanisms or in vitro third body wear mechanisms involving fragments of bone cement, UHMWPE , metal, ceramic, etc.
In an independent laboratory test on Howmedica's Duration® Stabilized UHMWPE, a hemispherical cup of generic design for hip simulator testing, with an inner diameter of 32 mm, a bearing thickness of 8.4 mm, and having undergone 11 days of heating in air to simulate 5 years "aging", showed a 27% reduction in volumetric wear versus the same cup conventionally gamma sterilized, having undergone 11 days of heating in air to simulate 5 years "aging". Testing was performed in a multiaxial hip joint wear simulator over 10 million cycles, using a 32 mm CoCr Head articulating counterface and bovine calf serum as a lubricant. The results of in vitro tests have not been shown to correlate with clinical wear mechanisms.
ル
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a human profile, composed of three curved lines. The logo is presented in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Frank Maas .Manager, Requlatory Affairs Howmedica, Inc. Pfizer Hospital Products Group 359 Veterans Boulevard Rutherford, New Jersey 07070
OCT 1 6 1997
Re: K972792 Duration Stabilized UHMWPE Exeter All Trade Name: Plastic Acetabular Component and Centrex® Bipolar Component Regulatory Class: II Product Codes: JDI and KWY Dated: July 25, 1997 Received: July 28, 1997
Dear Mr. Maas:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note:
{3}------------------------------------------------
Page 2 - Mr. Frank Maas
this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic enrough on or or on control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.qov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
ications for Use
510(k) Number (if known):
Device Name: Duration® Stabilized UHMWPE Acetabular Components
Indications for Use:
The Exeter All Plastic Acetabular Component (previously cleared in K936132) is intended to be used for primary and secondary reconstruction of the bearing surface of the acetabulum as a result of painful and/or severly disabled hip joints due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis or revision of a failed acetabular component.
The Howmedica® Bipolar Prosthesis, Centrax® (previously cleared in K855231) is intended to be used in conjunction with a femoral stem component for the reconstruction of a femoral head damaged by fresh fracture, non-union, aseptic necrosis of the head and neck, or by osteoarthritis or post-traumatic arthritis.
SEE NEXT PAGE FOR DURATION® WEAR CLAIMS
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X | OR | Over-The-Counter Use | |
|----------------------|---|----|----------------------|--|
| (Per 21 CFR 801.109) | | | | |
(Division Sign-Off)
D - General Restorative Devices
510(k) K972792 (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.