K972726 · Medical Packaging Corp. · JSM · Oct 8, 1997 · Microbiology
Device Facts
Record ID
K972726
Device Name
SNAP SWAB CULTURE COLLECTION AND TRANSPORT SYSTEM
Applicant
Medical Packaging Corp.
Product Code
JSM · Microbiology
Decision Date
Oct 8, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2390
Device Class
Class 1
Indications for Use
Medical Packaging Corporation's SNAP SWAB™ Culture Collection and Transport System, with Modified Stuart's, Modified Amies, and Cary-Blair Transport Media is intended for use as a disposable, sterile culture Collection device for use in the collection, preservation, and transportation of microbiological specimens.
Device Story
Disposable, sterile culture collection and transport system. Consists of two plastic parts: top cap containing transport media (0.3-0.6 mL) and bottom tube. Available with single or dual Rayon® tipped swabs on plastic shafts, or Rayon® Mini-Tip swab on aluminum shaft for male urethral or nasopharyngeal specimens. Operation: user collects specimen, inserts swab into tube, and fits bottom into top. User bends top 45° to break hollow swab shaft, allowing media to flow through shaft into bottom tube to moisten specimen. Maintains specimen viability for up to 72 hours at room temperature. Used in clinical settings for transport to laboratory for microbiological analysis.
Indicated for collection, preservation, and transport of bacteriological specimens to the laboratory for evaluation. Intended to preserve specimen integrity and minimize overgrowth of organisms during transport.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
CULTURE-PAK™ Collection and Transport System, Modified Stuart's Media (K932337/S1)
CULTURE-PAK™ Collection and Transport System, Modified Stuart's Media (K936078)
CULTURE-PAK™ Collection and Transport System, Modified Amies Transport Media (K964637)
CULTURE-PAK™ Collection and Transport System, Clinical Transport Media (K970597)
Submission Summary (Full Text)
{0}------------------------------------------------
K972726
## SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of safety and effectiveness information is being submitted in accordance with the requirements of The Safe Medical Devices Act of 1990 (SMDA 1990) and 21 CFR Part 807.92.
DATE OF SUMMARY PREPARATION: 18 July, 1997
| COMPANY: | Medical Packaging Corporation<br>941 Avenida Acaso<br>Camarillo, CA 93012-8700<br>Phone: 805-388-2383<br>Fax: 805-388-5531 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Patricia V. Willis |
| DEVICE NAME: | SNAP SWABTM<br>Culture Collection and Transport System |
| DEVICE CLASSIFICATION: | Class I, General Controls |
| COMMON NAME: | Microbiological specimen Culture<br>Collection and Transport device |
| PREDICATE DEVICE(S): | CULTURE-PAKTM Collection and Transport<br>System, Modified Stuart's Media<br>(K932337/S1 and K936078 ), and<br>CULTURE-PAKTM Collection and Transport<br>System, Modified Amies Transport Media,<br>K964637, CULTURE-PAKTM Collection<br>and Transport System, Clinical Transport<br>Media, K970597, Medical Packaging<br>Corporation, Camarillo, CA |
#### Intended Use
Medical Packaging Corporation's SNAP SWAB™ Culture Collection and Transport System, with Modified Stuart's, Modified Amies, and Cary-Blair Transport Media is intended for use as a disposable, sterile culture Collection device for use in the collection, preservation, and transportation of microbiological specimens.
{1}------------------------------------------------
### Substantial Equivalence
Medical Packaging Corporation's SNAP SWAB™ Culture Collection and Transport System, available with Modified Stuart's, Modified Amies, and Cary-Blair Transport Media, is substantially equivalent to Medical Packaging Corporation's CULTURE-PAK™ Collection and Transport System, Modified Stuart's Media (K936078) and ( K932337/S1), CULTURE-PAK ™ Collection and Transport System, Modified Amies Transport Media, K964637, and CULTURE-PAK™ Collection and Transport System, Clinical Transport Media (K970597). Listed devices are intended to collect, preserve, and transport clinical specimens to the laboratory for microbiological analysis. Devices are composed of two plastic parts:
- The top plastic part of the single swab device holds 0.4 mL of transport media . and a plastic shaft, Rayon® tipped swab. The top plastic part of the dual swab device holds 0.6 mL of transport media and two plastic shaft, Rayon® tipped swabs , and the top plastic part of the device with the Rayon® Mini-Tip swab holds 0.3 mL of transport media. The aluminum shaft, Rayon® Mini-Tip swab is for collection of either male urethral or nasopharyngeal specimens.
- The bottom plastic part of the device protects the swab(s) { hereafter referred . to as: Culture Collection device ) and fits tightly into the top plastic part. The SNAP SWAB™ Culture Collection and Transport System is packaged sterile. Sterilization is conducted at SteriGenics, International, Inc., and validated by North American Science Associates, Inc. for all configurations.
The methods for use are the same for the CULTURE-PAK ™ Collection and Transport System, Modified Stuart's Media (K936078) and ( K932337/S1), CULTURE-PAK ™ Collection and Transport System, Modified Amies Transport Media. K964637, and CULTURE-PAK ™ Collection and Transport System, with Clinical Transport Media, (K970597). The two plastic parts are pulled apart, exposing the swab(s). After the culture is collected, the bottom is refitted into the top. The media seal is broken and the fluid is forced into the bottom extrusion to keep the Culture Collection device moist for up to 72 hours at room temperature.
The primary similarities and differences between Medical Packaging Corporation's CULTURE-PAK ™ Collection and Transport System. Modified Stuart's Transport Media, CULTURE-PAK ™ Collection and Transport System, Modified Amies Media, CULTURE-PAK ™ Collection and Transport System, Clinical Transport Media, Cary-Blair Transport Media option, and the SNAP SWAB™ Culture Collection and Transport System are the following: (1) The release of media from the cap through the hollow shaft of the SNAP SWAB™ is accomplished by breaking the
{2}------------------------------------------------
### 510(k) Notification . 18 July, 1997 Medical Packaging Corporation SNAP SWAB™ Culture Collection and Transport System
swab shaft within the cap rather than breaking the plastic snap valve. The action required for breaking the swab and the snap valve are identical. Once the shaft is broken, the media flow through the shaft is identical to the CULTURE-PAK™ device. (2) The SNAP SWAB™ culture tube is individually labeled and is packaged 50 to a labeled foil pouch.
The SNAP SWAB™ Culture Collection and Transport System and the CULTURE-PAK™ Collection and Transport System are have the same intended use. Like the CULTURE-PAK™, the SNAP SWAB™ sealed hollow shaft is seated in the plug that holds the media in the top part of the device. The top is bent 45 ° to break the shaft allowing the media to flow through the exposed open shaft orifice into the bottom of the tube. The SNAP SWAB™ Culture Collection and Transport System, will also be available with two plastic shaft, Rayon® tipped swabs, and with an aluminum shaft Rayon® Mini-Tip ( Universal ) rayon swab.
All used materials should be treated as potentially infectious and bio-hazardous. Proper handling and disposal methods should be employed.
Danicia Uilles Date: 7-18-97
Patricia V. Willis Director, Customer Service and Regulatory Affairs
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling an eagle or bird in flight. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the bird symbol.
OCT - 8 1997
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Patricia V. Willis Director, Regulatory Affairs/ Customer Service Medical Packaging Corp. 941 Avendia Acasco Camarillo, CA 93012
Re: K972726
> Trade Name: Snap Swab Culture Collection and Transport System Regulatory Class: I Product Code: JSM Dated: July 18, 1997 Received: July 21, 1997
Dear Ms. Willis:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in-vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
ાત્રત્ય
510(k) Submission #K972726 SNAP SWAB Submission Revision . September 30, 1997 Medical Packaging Corporation
# INDICATIONS FOR USE:
Medical Packaging Corporation's SNAP SWAB™ Culture Collection and Transport System is intended to be used for the collection, preservation, and transport of bacteriological specimens to the laboratory for evaluation. The media is intended to preserve the integrity of the specimen as well as minimize overgrowth of other organisms during transport.
ាប់
C
Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K972726
Prescription Use ✓
OR
Over-The Counter Use *_*_
(Per 21 CFR 801.109)
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.