K972710 · Hematronix, Inc. · JJW · Aug 15, 1997 · Clinical Chemistry
Device Facts
Record ID
K972710
Device Name
QUANTIFY(901)
Applicant
Hematronix, Inc.
Product Code
JJW · Clinical Chemistry
Decision Date
Aug 15, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Indications for Use
The Hematronix qUAntify™ Control System is a liquid, ready-touse, urine dipstick control. No reconstitution or dilution is required. The control is prepared from human source material and predetermined chemicals. Controls evaluate the quality of day-to-day test performance. The qUAntify Control System is used to determine accuracy and precision in the measurement of the physical and chemical tests commonly employed in semiquantitative urine dipstick Alternative tests such as the Glucose (Clinites), testing. Bilirubin (Ictotest) and Ketone (Acetest) are also assayed. This control can be used to quality control the specified urine hCG tests. Two distinct levels are included to verify the recovery of normal and elevated patient values.
Device Story
qUAntify is a liquid, ready-to-use urine dipstick control system; requires no reconstitution or dilution. Prepared from human source material and predetermined chemicals. Used in clinical laboratory settings to evaluate day-to-day test performance; monitors accuracy and precision of semiquantitative urine dipstick tests (e.g., specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, hemoglobin, hCG). Two distinct levels (normal and abnormal) verify recovery of patient values. Healthcare providers use control results to validate instrument performance and ensure reliable patient testing. Contains sodium azide (0.05%) and human blood products; donors screened for HIV, HCV, and HBsAg.
Clinical Evidence
Bench testing only. Evaluated on multiple instruments for 11 analytes (specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, hemoglobin, hCG). Performance compared to Kenlor Liquid Urine Control package insert ranges. Results demonstrated substantial equivalence in precision and recovery for both normal and abnormal levels.
Technological Characteristics
Liquid, ready-to-use urine control. Matrix: human source materials and predetermined chemicals. Contains 0.05% sodium azide. Stability: 2 years at 2-8°C (unopened), 30 days at room temp (opened). No software or electronic components.
Indications for Use
Indicated for use as a quality control material to monitor the accuracy and precision of semiquantitative urine dipstick testing, including specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, hemoglobin, and hCG tests. Intended for laboratory use to verify recovery of normal and elevated values.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
{0}------------------------------------------------
K972710
### AUG 1 5 1997
Page Al of A3
### 510(k) Summary
For any questions contact: James D. Lapicola 524 Stone Road Benicia, CA 94510 (707) 746-7833 ext 208 FAX: (707) 746-7837
Date Prepared: July 18, 1997
#### Gentlemen:
Following is a summary of the basis for a substantially equivalent determination:
| Hematronix Product | |
|---------------------|-----------------------------------------------|
| Trade Name | qUAntify |
| Common Name | Clinical Chemistry<br>Urine Control |
| Classification Name | Quality Control Material<br>per 21CFR862.1660 |
Substantially Equivalent Device: Kenlor Liquid Ürine 510 ( k ) Control see #K890577
- is currently "manufactured for" Please note: qUAntify HEMATRONIX, INC. by Kenlor Industries. It is intention to manufacture the material our ourselves.
### Device Description/INTENDED USE
The Hematronix qUAntify™ Control System is a liquid, ready-touse, urine dipstick control. No reconstitution or dilution is required. The control is prepared from human source material and predetermined chemicals.
Controls evaluate the quality of day-to-day test performance. The qUAntify Control System is used to determine accuracy and precision in the measurement of the physical and chemical tests commonly employed in semiquantitative urine dipstick Alternative tests such as the Glucose (Clinites), testing. Bilirubin (Ictotest) and Ketone (Acetest) are also assayed. This control can be used to quality control the specified urine hCG tests. Two distinct levels are included to verify the recovery of normal and elevated patient values.
{1}------------------------------------------------
I
# Technological Characteristic Comparison
Both Kenlor's Liquid Urine Control and Hematronix' qUAntify are prepared from human source materials and predetermined chemicals. Two distinct levels are provided to verify the recovery of normal and elevated patient values.
Equivalency between qUAntify and Kenlor Liquid Urine Control
| | qUAntify<br>(Applicant) | Liquid Urine Control<br>(Kenlor) |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | To monitor the:<br>precision system<br>for the analytes<br>listed. | To monitor the<br>precision system<br>for the analytes<br>listed. |
| Measurement<br>Technique: | Semiquantitative<br>testing and assays | Semiquantitative<br>testing and assays |
| Analytes: | Specific Gravity<br>pH<br>Leukocytes<br>Nitrite<br>Protein<br>Glucose<br>Ketones<br>Urobilinogen<br>Bilirubin<br>Hemoglobin<br>HCG | Specific Gravity<br>pH<br>Leukocytes<br>Nitrite<br>Protein<br>Glucose<br>Ketones<br>Urobilinogen<br>Bilirubin<br>Hemoglobin<br>HCG |
| Number of<br>Common Analytes: | 11 | 11 |
| Matrix: | Human Source Materials<br>and predetermined<br>chemicals | Human Source Materials<br>and predetermined<br>chemicals |
| Stability: | Unopened Vial<br>2 years at 2°-8°C<br><br>Opened Vial<br>30 days at room temp. | Unopened Vial<br>2 years at 2°-8°C<br><br>Opened Vial<br>30 days at room temp. |
{2}------------------------------------------------
## Non Clinical Performance Data - Conclusions
qUAntify was evaluated on several instruments for all 11 The ranges observed were compared to published analytes. analy of the same analytes in package inserts provided with Kenlor's Liquid Urine Control. The ranges were substantially The data is summmaried in the following table. equivalent.
| qUAntify<br>mfg by HEMATRONIX, INC. | Liquid Urine Control<br>mfg by Kenlor | |
|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Level I (Normal): | | |
| Specific Gravity<br>pH | 1.020 ±.005<br>6 ±.5 | 1.010 ±.005<br>5.5 ±.5 |
| Level II (Abnormal): | | |
| Specific Gravity<br>pH<br>Leukocytes<br>Nitrite<br>Protein<br>Glucose<br>Ketones<br>Urobilinogen<br>Bilirubin<br>Hemoglobin<br>HCG | 1.010-1.015<br>8<br>++<br>+<br>100<br>250-500<br>++<br>1-8<br>++<br>250<br>+ | 1.005-1.015<br>7-8<br>+<br>+<br>100<br>250-350<br>++<br>1-4<br>+++<br>250<br>+ |
Equivalency Ranges as measures of Precision.
## Safety and Effectiveness Conclusions:
- qUAntify and Kenlor's MSDS list: 1) sodium azide Safety acts as a constituent (0.05%); and 2) Human blood blood products which is potentially and a biohazardous material. Each blood donor, however, is tested by FDA approved methods for HIV, HCV and HBs Aq.
- As seen in the Non Clinical Performance Effectiveness -Data section above, the qUAntify control system will evaluate the guality of dayto-day test performance when used to determine accuracy and precision in the measurement of the physical and chemical commonly employed in semitests quantitative urine dipstick testing.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image contains the text 'Public Health Service'. The text is in a simple, sans-serif font and is horizontally aligned. The words are stacked on top of each other, with 'Public' on the top line, 'Health' on the second line, and 'Service' on the bottom line.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. James D. Lapicola Executive Vice President Hematronix, Inc. 524 Stone Road Benicia, CA 94510
AUG 1 5 1997
K972710 Re : qUAntify Requlatory Class: I Product Code: JJW Dated: July 18, 1997 Received: July 21, 1997
Dear Mr. Lapicola:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
{4}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to the regulation Chercrea, "Miobranding by rolusios of the general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
:10(k) Number (if known):
)evice Name: qUAntify
ndications For Use:
The Hematronix qUAntify TM Control System is a liquid, ready-to-use urine dipstick No reconstitution or dilution is required. The control is prepared from control. numan source material and predetermined chemicals.
jontrols evaluate the quality of day-to-day test performance. The qUAntify Control ystem is used to determine accuracy and precision in the measurement of the physical and chemical tests commonly employed in semiquantitative urine dipstick testing. Ilternative tests such as the Glucose (Clinites), Bilirubin (Ictotest) and Ketone (Acetest) are also assayed. This control can be used to quality control the specified rine hCG tests. Two distinct levels are included to verify the recovery of normal nd elevated patient values.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
-Off)
Clinical Laboratory Devices
Number K972710
escription Use er 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.