INNERDYNE RADIALLY EXPANDING VASCULAR ACCESS SYSTEM
K972684 · Innerdyne, Inc. · DYB · Oct 10, 1997 · Cardiovascular
Device Facts
Record ID
K972684
Device Name
INNERDYNE RADIALLY EXPANDING VASCULAR ACCESS SYSTEM
Applicant
Innerdyne, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Oct 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The InnerDyne, Inc., Radially Expanding Vascular Access System is intended to provide percutaneous access to the vasculature. The Radially Expanding Vascular Access System is indicated for the following usages: Percutaneous vascular access. The Radially Expanding Vascular Access System is intended for use during minimally invasive surgery for temporary dilation access to the vasculature for passage of instruments into the vasculature.
Device Story
System consists of radially expandable sleeve assembly, tapered dilator, and sheath with integral hemostasis valve. Operates by inserting expandable sleeve over guidewire into vessel; sleeve radially compresses to reduce diameter for insertion. Once in place, dilator/sheath assembly is inserted through sleeve lumen, causing sleeve to expand radially and dilate surrounding tissue. Dilator is removed, leaving sleeve and sheath in place to provide sealed port for instrument passage. Allows initial placement of small diameter sheath, which can be removed while leaving sleeve in vessel to accommodate larger dilator/sheath for larger instruments. Used in clinical settings by physicians for minimally invasive procedures. Benefits include controlled, sequential dilation of vascular access site.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and functional similarities to the predicate device.
Technological Characteristics
System comprises expandable sleeve, tapered dilator, and sheath with integral hemostasis valve. Components assembled via adhesives, welds, and mechanical interlocks using medical grade materials. Operates via mechanical radial expansion. Sterilization via alternative methods/locations consistent with InnerDyne Step™ family.
Indications for Use
Indicated for percutaneous vascular access in patients undergoing minimally invasive surgery requiring temporary dilation access to the vasculature for the passage of instruments. Configured for arterial or venous access.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
K972684
OCT 10 1997
## SUMMARY OF SAFETY & EFFECTIVENESS INFORMATION
- InnerDyne, Inc. Submitter: 5060 Amelia Earhart Dr Salt Lake City, Utah (801)-350-3600
Date Prepared: 11 July 1997
- Rick Gaykowski Contact: Corporate Vice President, Regulatory Affairs and Quality Assurance
Classification Name: Percutaneous Introducer Common/Usual Name: Percutaneous Introducer With Sheath Trade/Proprietary Name: InnerDyne Radially Expanding Vascular Access System
The Radially Expanding Vascular Access System consists of an expandable sleeve assembly: The tubular member of the expandable is configured so as to be radially compressed to reduce the outside diameter of the device prior to insertion. Upon use, the expandable sleeve assembly is inserted through the skin and surrounding tissue into the selected vessel over a previously placed guidewire. (An alternative placement method would have the needle placed within the lumen of the expandable sleeve, and the combined needle and sleeve inserted into the target vessel. The guidewire would be placed and the needle removed.) The dilator/sheath assembly, which consists of a tapered dilator and sheath with integral hemostasis valve, is inserted over the guidewire through the lumen of the expandable sleve which expands radially to accommodate the dilator/sheath. This process in turn, radially expands the surrounding tissue. Following dilator is removed, leaving the expandable sleeve and sheath with integral hemostasis valve in place to provide a sealed port for passage of instruments.
This system configuration allows the user to initially place a small diameter sheath for passage of small instruments. The sheath can then be removed from the lumen of the sleeve while leaving the sleve inserted in the vessel. A larger diameter dilator/sheath can then be inserted through the sleeve to create a larger port for passage of the larger instruments.
The device is assembled from medical grade materials under GMP and ISO conditions. Components are molded and machined by qualified suppliers. The components are assembled and secured by adhesives, welds, and mechanical interlocks.
The subject InnerDyne, Inc., Radially Expanding Vascular Access System device is substantially equivalent to the predicate InnerDyne, Inc., Radially Expanding Dilating Introducer device (K931489). The subject Radially Expanding Vascular Access System device is similar to the referenced predicate in size, function, product dimensions and indications for use.
The Radially Expanding Vascular Access System is intended for use during minimally invasive surgery for temporary dilation access to the vasculature for passage of instruments into the vasculature. The
{1}------------------------------------------------
device is configured to be used for arterial or venous access.
The basic design principles for the subject InnerDyne, Inc., Radially Expanding Vascular Access System device and the predicate Radially Expanding Dilating Introducer device (K931489) are similar, and remain essentially unchanged from information previously provided to the Agency. The product configuration, composition, and utilized materials are similar in each of the products. The principles of operation for the InnerDyne, Inc., Radially Expanding Vascular Access System device, and the predicate Radially Expanding Dilating Introducer device (K931489) are similar. That is, each of these products employs a similar insertion technique, indications for use, contraindications for use, warnings and precautions. The subject device differs from the referenced Radially Expanding Dilating Introducer predicate (931489) only with changes to the materials used in the construction of the expanding sleve portion and the presentation style of the product. Alternative locations and methods of sterilization are being included. The new material used in the construction of the expandable sleeve as well as the alternative sterilization locations and methods are similar to those used in the InnerDrne Step™ family of products ( K943253, K940232, K950632, and K950658).
From the foregoing, we conclude that the Radially Expanding Vascular Access System device is as safe and effective as currently marketed devices for the stated indications.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The logo is black and white.
OCT 1 0 1997
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Rick Gaykowski InnerDyne, Inc. 5060 Amelia Earhart Drive Salt Lake City, Utah 84116
Re: K972684 InnerDyne Radially Expanding Vascular Access System Regulatory Class: II (two) Product Code: 74 DYB Dated: July 11, 1997 Received: July 17, 1997
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Rick Gaykowski
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled,
"Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page I of I
## 510(k) Number (if known): K972684
Device Name: InnerDyne, Inc., Radially Expanding Vascular Access System
The InnerDyne, Inc., Radially Expanding Vascular Access System is intended Indications for Use: to provide percutaneous access to the vasculature. The Radially Expanding Vascular Access System is indicated for the following usages: Percutaneous vascular access. .
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|-------------------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Cardiovascular, Respiratory, and Neurological Devices | |
| 510(k) Number | K972684 |
| Prescription Use (Per 21 CFR 801.109) | OR | Over-the-Counter Use (Optional Format 1-2-96) |
|---------------------------------------|----|-----------------------------------------------|
|---------------------------------------|----|-----------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.