DEPUY AMK 3 PEG PATELLA

K972643 · Depuy, Inc. · JWH · Jan 21, 1998 · Orthopedic

Device Facts

Record IDK972643
Device NameDEPUY AMK 3 PEG PATELLA
ApplicantDepuy, Inc.
Product CodeJWH · Orthopedic
Decision DateJan 21, 1998
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DePuy AMK 3 Peg Patella is intended for use with the DePuy AMK Total Knee System for cemented, tricompartmental total knee arthroplasty, utilizing the AMK system's existing femoral components, tibial trays, and tibial inserts. The AMK system is intended to replace severely disabled or painful joints resulting from osteoarthritis or post-traumatic arthritis, joints with correctable varus or valgus deformity and joints with failed previous surgeries where pain, deformity or dysfunction persist.

Device Story

DePuy AMK 3 Peg Patella is an all-polyethylene orthopedic implant; features domed articular surface and 3-peg fixation surface. Designed for use with DePuy AMK Total Knee System components (femoral, tibial trays, inserts) in cemented, tricompartmental total knee arthroplasty. Implanted by orthopedic surgeons in clinical/hospital settings to replace painful or disabled knee joints. Fixation achieved via acrylic bone cement. Device restores joint function and reduces pain for patients with osteoarthritis, post-traumatic arthritis, or deformity.

Clinical Evidence

Bench testing only. No clinical data provided. Testing demonstrated that the fixation strength of the AMK 3 Peg Patella is similar to the Johnson & Johnson PFC Oval Sombrero Patella and the Zimmer NexGen All Polyethylene Patella.

Technological Characteristics

All-polyethylene patellar component; domed articular surface; 3-peg fixation surface. Intended for cemented use only. No electronic, software, or energy-based components.

Indications for Use

Indicated for patients requiring cemented, tricompartmental total knee arthroplasty due to severe joint disability or pain from osteoarthritis, post-traumatic arthritis, correctable varus or valgus deformity, or failed previous surgeries.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### SUMMARY OF SAFETY AND EFFECTIVENESS | NAME OF FIRM: | DePuy Inc. | |-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------| | | 700 Orthopaedic Drive | | | Warsaw, Indiana 46581-0988 | | 510(k) CONTACT: | Cheryl Hastings | | | Manager, Clinical Affairs | | TRADE NAME: | DePuy AMK 3 Peg Patella | | COMMON NAME: | All Polyethylene Patella | | CLASSIFICATION: | 888.3560 Prosthesis, Knee, PatelloFemorotibial, Semiconstrained, Cemented, Polymer, Metal, Polymer | | DEVICE PRODUCT CODE: | 87 JWH | | SUBSTANTIALLY EQUIVALENT DEVICES: | <ul><li>DePuy AMK All Polyethylene Patella</li><li>DePuy LCS 3 Peg Patella</li><li>Johnson &amp; Johnson PFC Oval Sombrero Patella</li></ul> | - Zimmer NexGen All Polyethylene Patella જુન ### DEVICE DESCRIPTION AND INTENDED USE: The DePuy AMK 3 Peg Patella is an all polyethylene patella component with a domed articular surface and a 3 peg fixation surface. It is intended for use with the DePuy AMK Total Knee System for cemented, tricompartmental total knee arthroplasty, utilizing the AMK system's existing femoral components, tibial trays, and tibial inserts. The AMK system is intended to replace severely disabled or painful joints resulting from osteoarthritis or post-traumatic arthritis, ioints with correctable varus or valgus deformity and joints with failed previous surgeries where pain, deformity or dysfunction persist. #### BASIS OF SUBSTANTIAL EQUIVALENCE: The DePuy AMK 3 Peg Patella is identical to the AMK All Polyethylene Patella which has been previously cleared by FDA except that the fixation surface has been changed from two circular arcs to 3 pegs. The material, intended use and design of the articular surface of the patella are the same as those of the original AMK All Polyethylene Patella. The 3 peg fixation surface is similar to the fixation surface used on the DePuy LCS All Polyethylene 3 Peg Patella. Testing indicates that the fixation strength of the AMK 3 Peg Patella is similar to those of the Johnson & Johnson PFC Oval Sombrero Patella and the Zimmer NexGen All Polyethylene Patella. ## 000002 {1}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 2 1 1998 Ms. Cheryl Hastings Manager, Regulatory Submissions DePuy, Inc. P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988 K972643 Re : AMK 3 Peq Patella Requlatory Class: II Product Code: JWH Dated: December 17, 1997 Received: December 19, 1997 Dear Ms. Hastings: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on this device being equivalent only to similar devices labeled and intended to be fixed within bone with acrylic "bone You may, therefore, market your device subject to cement . " the general controls provisions of the Act and the following limitations: - This device may not be labeled or promoted for non-1. cemented use. - 2. All labeling for this device, including package label and labeling included within the package, must prominently state that the device is intended for cemented use only. {2}------------------------------------------------ ### Page 2 - Ms. Cheryl Hastings - Any non-cemented fixation of this device is considered 3. investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulation under 21 CFR, Part 812. All users of the device for noncemented fixation must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) to conduct the investigation. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General requlation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be {3}------------------------------------------------ Page 3 - Ms. Cheryl Hastings obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known) _k972643 Device Name AMK 3 Peg Patella Indications for Use: The DePuy AMK 3 Peg Patella is intended for use with the DePuy AMK Total Knee System for cemented, tricompartmental total knee arthroplasty, utilizing the AMK system's existing femoral components, tibial trays, and tibial inserts. The AMK system is intended to replace severely disabled or painful joints resulting from osteoarthritis or post-traumatic arthritis, joints with correctable varus or valgus deformity and joints with failed previous surgeries where pain, deformity or dysfunction persist. Concurrence of CDRH, Office of Device Evaluation Acolla (Div Divisi of General Restorative Devices 510(k) Number. K97264 Prescription Use >> OR Over-The Counter Use (Per 21 CFR 801.109) # 000001
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...