K972525 · Michigan Instruments, Inc. · DRM · Oct 2, 1997 · Cardiovascular
Device Facts
Record ID
K972525
Device Name
THUMPER
Applicant
Michigan Instruments, Inc.
Product Code
DRM · Cardiovascular
Decision Date
Oct 2, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Thumper® Model 1007 is intended for use on persons in immediate need of manual cardiopulmonary resuscitation (CPR) as described in the American Heart Association, Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiac Care, 1992. The Thumper® Model 1007 type of medical device is referred to in the AHA CPR Guideline as an "automatic mechanical chest compressor." It is used as an adjunct to manual CPR for use by trained personnel to optimize compression and reduce rescuer fatigue in prolonged resuscitation efforts.
Device Story
Thumper Model 1007 is a pneumatically powered mechanical cardiopulmonary resuscitator; provides automated chest compressions and ventilation to patients in clinical death. Device consists of gas-powered piston assembly, massager pad, and backboard. Operates by delivering compressions at 90 cycles/minute with 50% duty cycle and 5:1 compression-to-ventilation ratio, mimicking manual CPR. Includes built-in constant-flow, time-cycled ventilator delivering 100% oxygen. Used by trained personnel as an adjunct to manual CPR to optimize compression consistency and reduce rescuer fatigue during prolonged resuscitation. Healthcare providers use the device to support circulatory and respiratory function; output directly replaces manual chest compressions and mouth-to-mouth breathing. Benefits include standardized CPR delivery and reduced physical strain on rescuers.
Clinical Evidence
Bench testing only. Comparative testing performed on Thumper Models 1004, 1005, and 1007 using a calibrated spring load (MII model T106) and a calibrated test lung (MII Pneu® View Model 2601i). Endpoints included piston compression rate (cycles/min), depth of compression (inches), and compression duration duty cycle (%). Results demonstrated compliance with AHA manual CPR standards and comparable performance to predicate devices.
Technological Characteristics
Pneumatically powered mechanical chest compressor. Components: gas-powered piston, massager pad, backboard, and integrated constant-flow, time-cycled ventilator. Delivers 100% oxygen. Dimensions: smaller and lighter than predicates. Operation: 90 compressions/minute, 50% duty cycle, 5:1 compression-to-ventilation ratio. No software or electronic algorithms; purely mechanical/pneumatic control system.
Indications for Use
Indicated for adult patients in clinical death (lack of spontaneous breathing and pulse) requiring cardiopulmonary resuscitation (CPR).
Regulatory Classification
Identification
An external cardiac compressor is an externally applied prescription device that is electrically, pneumatically, or manually powered and is used to compress the chest periodically in the region of the heart to provide blood flow during cardiac arrest. External cardiac compressor devices are used as an adjunct to manual cardiopulmonary resuscitation (CPR) when effective manual CPR is not possible (e.g., during patient transport or extended CPR when fatigue may prohibit the delivery of effective/consistent compressions to the victim, or when insufficient EMS personnel are available to provide effective CPR).
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Nonclinical performance testing under simulated physiological conditions must demonstrate the reliability of the delivery of specific compression depth and rate over the intended duration of use.
(2) Labeling must include the following:
(i) The clinical training necessary for the safe use of this device;
(ii) Adjunctive use only indication prominently displayed on labels physically placed on the device and in any device manuals or other labeling;
(iii) Information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations,
*e.g.,* adult, pediatric and/or infant); and(iv) Information on the time necessary to deploy the device as demonstrated in the performance testing.
(3) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(4) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user, including an evaluation for the time necessary to deploy the device.
(5) For devices containing software, software verification, validation, and hazard analysis must be performed.
(6) Components of the device that come into human contact must be demonstrated to be biocompatible.
{0}------------------------------------------------
K972525
Image /page/0/Picture/1 description: The image shows the logo for Michigan Instruments INC. The logo consists of a stylized globe with three arrowheads pointing to the right. To the right of the globe is the text "Michigan Instruments" in a bold, sans-serif font, with "INC." in a smaller font below the word "Instruments".
OCT - 2 1997
4717 Talon Court. S.E. · Grand Rapids, MI 49512-5409 Phone (616) 554-9696 · Fax (616) 554-3067 1-800-530-9939
## July 2, 1997
## 510(k) Summary - Thumper® Model 1007
- James D. Maatman, President Submitter -Michigan Instruments. Inc. 4717 Talon Court, SE Grand Rapids, MI 49512-5409 Tel: 616-554-9696 Fax: 616-554-3067
- Trade Name Thumper® Model 1007 .
- . Common name - Mechanical Cardiopulmonary Resuscitator
- . Classification name - External cardiac compressor, product code 870.5200
- Substantially equivalent devices -.
Thumper® Model 1004:
a "preamendment device."
Thumper® Model 1005:
510(k) Clearance dated May 15, 1985), reference number K851139.
### Description of Thumper® Model 1007
The Thumper® Model 1007 is a pneumatically powered external cardiac compressor used on a patient in a state of clinical death and in immediate need of respiratory and circulatory support. The Thumper® Model 1007 uses a gas powered piston assembly with a massager pad and associated backboard to perform the function of a rescuer pressing on the patient's chest with his or her hands. A built-in ventilator to replaces the rescuer's mouth to mouth breathing for the patient. The action of the piston and the action of the associated ventilator have been designed to perform mechanical CPR according to contemporary American Heart Association (AHA) CPR guidelines for manual CPR. The Thumper® Model 1007 delivers standard CPR in a 5:1 compression-ventilation ratio with a compression duration that is 50% of the cycle length at a rate of 90 compressions per minute.
{1}------------------------------------------------
## 510(k) Summary - Thumper® Model 1007
- Intended use The Thumper® Model 1007 is intended for use on persons in immediate need of manual cardiopulmonary resuscitation (CPR) as described in the American Heart Association, Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiac Care, 1992. The Thumper® Model 1007 type of medical device is referred to in the AHA CPR Guideline as an "automatic mechanical chest compressor." It is used as an adjunct to manual CPR for use by trained personnel to optimize compression and reduce rescuer fatigue in prolonged resuscitation efforts.
- Technical characteristics The Thumper® Model 1007 is very similar and performs the same clinical function as the predicate devices. Thumper® Model 1004 and Thumper® Model 1005. The main distinguishing characteristics of the Thumper® Model 1007 are in response to Thumper® Model 1004 and Thumper® Model 1005 clinical user input and requests for product changes:
- The Model 1007 is smaller and lighter than the predicate devices. .
- The complexity and number of the Model 1007 controls has been reduced .
- . The Model 1007 controls have improved ergonomics and ease of use.
- . The built-in ventilator of the Model 1007 is of the constant flow, timecycled operation as opposed to the previous pressure limited time-cycled operation.
- . The Model 1007 built-in ventilator delivers 100% oxygen as compared to approximately 80% delivered oxygen of the Model 1004 and Model 1005
- Bench Test The Thumper® Model 1004. Thumper® Model 1005. and Thumper® Model 1007 were comparatively tested. An exemplar model of each unit was first qualified by passing its respective quality assurance test. Then, in turn, each exemplar was tested using a MII calibrated spring load. This unit (MII model T106) was designed and built based on a clinical study of human chest elastic properties during CPR. The T106 was instrumented to produce piston deflection waveforms for comparison. Simultaneously with the operation of the Thumper® piston, the associated built-in ventilators for each exemplar were used to ventilate a calibrated test lung, MII Pneu® View Model 2601i, set to lung compliance and resistance values typical of an adult CPR situation. The resultant piston compression rate (cycles/min), depth of compression (inches), compression duration duty cycle (%) demonstrate compliance with AHA standards for manual CPR. The ventilators for each exemplar demonstrated comparable ventilation of the test lung.
{2}------------------------------------------------
# 510(k) Summary - Thumper® Model 1007
- Conclusions from Bench Tests The Thumper Model 1007 provides compressive piston action similar to the Thumper Model 1004 and Thumper Model 1005. The compression action of the Thumper Model 1007 at 90 compressive cycles per minute with a compression duty cycle of 50% against the T 106 spring test load complies with present AHA manual CPR standards. The Thumper® Model 1007 built-in ventilator ventilated the test lung in the same manner as the Thumper® Model 1004 and Thumper® Model 1005. All three models of the Thumper® comply with 2-rescuer manual CPR ventilation by providing ventilation on a 5:1 compression to ventilation ratio. The bench tests indicate the Thumper® Model 1007 is substantially equivalent in design and performance to both the predicate devices, Thumper® Model 1004 and Thumper® Model 1005.
James D. Moatman
James D. Maatman, President Michigan Instruments, Inc.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with a single snake winding around a staff, and it is positioned within a circular frame. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Public Health Service
OCT - 2 1997
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
Mr. James D. Maatman President Michigan Instruments, Inc. 4717 Talon Court, S.E. Grand Rapids, Michigan 49512-5409
Re: K972525 Thumper® Model 1007 Requlatory Class: III (Three) Product Code: DRM September 25, 1997 Dated: Received: September 26, 1997
Dear Mr. James D. Maatman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{4}------------------------------------------------
Page 2 - Mr. James D. Maatman
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
510(K) Number: K972525
Device Name: Thumper® Cardiopulmonary Resuscitator
Indications for Use:
This device is used to perform Cardiopulmonary Resuscitation (CPR) on adult patients and only adult patients in cases of clinical death as defined by a lack of spontaneous breathing and pulse.
## PLEASE DO NOT WRITE BELOW THIS LINE
Concurrence of CDRH, Office of Device Evaluation (ODE)
Seon G. Lemkelee
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number_k 97
Prescription Use
(Per 21 C.F.R. § 801.109)
Over-the-Counter Use ___________
פיונים - 71377/300 - 0468564.01
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.