The Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter is recommended for use in percutaneous dilatation of the iliac, femoral, popliteal, tibial, renal, and tibioperoneal vessels in whose lumens are obstructed by atherosclerotic plaque. This catheter is not intended for use in the coronary arteries. Any other use than those indicated is not recommended.
Device Story
Smash™ PTA Catheter is a bilumen catheter with a controlled-compliance balloon at the distal tip; used for percutaneous dilatation of peripheral vessels (iliac, femoral, popliteal, tibial, renal, tibioperoneal) obstructed by atherosclerotic plaque. Physician operates device in clinical setting. Balloon inflated/deflated via side port to specific diameter/pressure; second lumen allows guidewire insertion, pressure monitoring, and contrast media infusion. Two marker bands aid placement. Device benefits patients by restoring vessel patency. Non-clinical performance testing (burst strength, compliance, fatigue, bond strength, trackability) confirms safety and efficacy.
Clinical Evidence
Bench testing only. Testing followed FDA guidance for Interventional Cardiology Devices (May 1994). Evaluated balloon burst strength, compliance, inflation/deflation performance, fatigue, bond strength, catheter profile, and body burst pressure. Additional evaluations included resistance through stenosis, pull-back resistance, and trackability. Biocompatibility testing was performed and found acceptable.
Technological Characteristics
Bilumen catheter; Nylon balloon and shaft; ABS manifold. Two gold marker bands. Rated burst pressure 7-15 atm. Compatible with 0.035-inch guidewires. Shaft length 60-120 cm; balloon diameter 3-12 mm. Mechanical device; no energy source or software.
Indications for Use
Indicated for percutaneous dilation of iliac, femoral, popliteal, tibial, renal, and tibioperoneal vessels obstructed by atherosclerotic plaque. Not for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Premarket Notification for - Smash™ PTA Catheter
# FEB 1 2 1998
## 16. 510(k) Summary
| Date Prepared | June 30, 1997 |
|--------------------------------|---------------------------------------------------------------------------------------|
| Submitter | |
| Address: | Schneider (USA) Inc<br>5905 Nathan Lane<br>Minneapolis, MN 55442 |
| Phone : | (612) 550-5500 |
| Fax : | (612) 550-5771 |
| Contact Person | |
| | Ronald W. Bennett<br>Senior Regulatory Affairs Specialist |
| Device Name and Classification | |
| Trade Name | Smash™ Percutaneous Transluminal Angioplasty<br>(PTA) Catheter |
| Common Name | Percutaneous Transluminal Angioplasty (PTA)<br>Catheter |
| Classification | Class II |
| Predicate Device | Match-35TM Percutaneous Transluminal<br>Angioplasty (PTA) Catheter - K913297, K926271 |
#### Device Description
The Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter consists of a bilumen catheter with a controlled-compliance balloon mounted at the distal tip. The balloon is designed to inflate to a known diameter and length at a specific pressure and is inflated and deflated via the side port. The second lumen allows access to the distal tip of the catheter for guidewire insertion, pressure monitoring and infusion of contrast media. Two marker bands are positioned under the balloon at the proximal and distal tapers to aid in accurate placement.
and K942154
### Intended Use
The Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter is recommended for use in percutaneous dilatation of the iliac, femoral, popliteal, tibial, renal, and tibioperoneal vessels in whose lumens are obstructed by atherosclerotic plaque. This catheter is not intended for use in the coronary arteries. Any other use than those indicated is not recommended.
{1}------------------------------------------------
# Technical Characteristic Comparison to Predicate
The Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter has the same intended use and basic construction as the predicate device, the Match-35™ Percutaneous Transluminal Angioplasty (PTA) Catheter. The following table compares the technical characteristics of the two devices:
| Feature | Match-35™ Percutaneous<br>Transluminal Angioplasty<br>(PTA) Catheter | Smash™ Percutaneous<br>Transluminal Angioplasty<br>(PTA) Catheter |
|----------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------|
| Configuration | Bilumen | Bilumen |
| Shaft Length (cm) | 40-200 | 60-120 |
| Shaft French Size | 5 | 5 |
| Balloon Length (mm) | 20-100 | 20-80 |
| Balloon Diameter (mm) | 3-12 | 3-12 |
| Rated Burst Pressure (atm) | 7-12 | 7-15 |
| Guidewire Size (inches) | 0.035 | 0.035 |
| Balloon Markers | 2 (Gold / Tantalum) | 2 (Gold) |
| Balloon Material | Polyethylene Terephthalate | Nylon |
| Manifold Material | Polycarbonate | ABS |
| Shaft Material | Polyester | Nylon |
## Performance Data
The Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter had non-clinical performance testing as did the predicate device, the Match-35TM Percutaneous Transluminal Angioplasty (PTA) Catheter.
Testing of the Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter generally followed the PTCA Balloon Catheters section in Part II of the "Guidance for the Submission of Research and Marketing Applications for Interventional Cardiology Devices", May 1994. The following types of tests were performed similar to those in the guidance:
Balloon Minimum Burst Strength Balloon Compliance (Distensibility) Balloon Inflation/Deflation Performance Balloon Fatigue (Repeated Balloon Inflation) Bond Strength Catheter Diameter and Balloon Profile Catheter Body Burst Pressure
{2}------------------------------------------------
The following additional evaluations were also performed by Schneider (USA):
Resistance through Stenosis Pull Back Resistance into Introducer Trackability Catheter Prep Time
The biocompatibility of the catheter was also tested and found acceptable.
#### Summary
In summary, Schneider (USA) Inc has demonstrated the Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter is substantially equivalent to the Match-35TM Percutaneous Transluminal Angioplasty (PTA) Catheter based on design, test results, and indications for use and is therefore acceptable for commercialization.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image is a black and white seal for the U.S. Department of Health and Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle.
FEB 1 2 1998
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Ronald W. Bennett Senior Regulatory Affairs Specialist Schneider (USA) Inc. Pfizer Hospital Products Group 5905 Nathan Lane Minneapolis, MN 55442
K972512 Re: Smash™ Percutaneous Transluminal Angioplasty (PTA) Regulatory Class: II (two) Product Code: 74 LIT Dated: January 16, 1998 Received: January 20, 1998
Dear Mr. Bennett:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: : General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-Also, please note the regulation entitled, "Misbranding by 4639. reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahon
Thomas J. Cállahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Premarket Notification for Smash™ PTA Catheter
Page of
510(k) Number (if known):
Device Name:
Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter
Indications for Use:
The Smash™ Percutaneous Transluminal Angioplasty (PTA) Catheter is recommended for use in percutaneous dilation of the iliac, femoral, popliteal, tibial, renal, and tibioperoneal vessels in whose lumens are obstructed by atherosclerotic plaque. This catheter is not intended for use in the coronary arteries. Any other use than those indicated is not recommended.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--|--|--------------------------------------------------------|
|--|--|--------------------------------------------------------|
| | | <table><tr><td>(Division Sign-Off)</td></tr><tr><td>Division of Cardiovascular, Respiratory,</td></tr><tr><td>and Neurological Devices</td></tr></table> | (Division Sign-Off) | Division of Cardiovascular, Respiratory, | and Neurological Devices |
|------------------------------------------|--|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|------------------------------------------|--------------------------|
| (Division Sign-Off) | | | | | |
| Division of Cardiovascular, Respiratory, | | | | | |
| and Neurological Devices | | | | | |
| 510(k) Number | K972512 |
|---------------|---------|
|---------------|---------|
| Prescription Use (Per 21 CFR 801.109) | OR | Over-The-Counter Use |
|---------------------------------------|----|----------------------|
|---------------------------------------|----|----------------------|
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.