← Product Code [DSI](/productcode/DSI) · K972498

# OBSERVER * CENTRAL STATION (K972498)

_Johnson & Johnson Medical, Inc. · DSI · Feb 17, 1998 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K972498

## Device Facts

- **Applicant:** Johnson & Johnson Medical, Inc.
- **Product Code:** [DSI](/productcode/DSI.md)
- **Decision Date:** Feb 17, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The OBSERVER Central Station with ST-segment measurement is a prescription device intended for use only by health care professionals. The device is used for the remote monitoring of physiological parameters, including non-invasive and invasive blood pressure; invasive hemodynamic and intracranial pressures; oxygen saturation via pulse oximetry; temperature; ECG ; and pulse/heart rate of adult, pediatric and neonatal patients. The device is located at a distance from the patient but within the same facility, including settings such as hospital and outpatient services, including general medical/surgical, critical care, intermediate/step-down care, emergency room, radiology, labor and delivery, operating and recovery room, cardiac catheterization lab, endoscopy and same-day surgery. Arrhythmia detection and ST-segment measurement are optional ECG features limited to the adult population. ST-segment measurement is contraindicated in paced patients. The ST-segment feature of the OBSERVER Central Station provides for the measurement of the ST-segment level and slope of the ECG waveform, alerting the clinician to STsegment changes. The significance of ST-segment changes must be determined by a clinician. The device is not designed, sold or intended for use except as indicated.

## Device Story

OBSERVER Central Station is a PC-based remote monitoring system for up to 8 patients. It receives physiological data (ECG, BP, SpO2, temperature, heart rate) from bedside monitors via hardwire (RS-232), wireless (900 MHz), or VHF telemetry. The system displays, records, and stores data for clinician review. The modification adds ST-segment measurement and alarm functionality for adult patients. Clinicians can adjust waveform points, alarm priorities, and limits. The system uses a Pentium PC with an SVGA monitor and thermal recorder. It aids clinical decision-making by providing real-time surveillance and historical waveform disclosure, allowing clinicians to identify and respond to physiological changes.

## Clinical Evidence

Bench testing only. The ST-enabled arrhythmia analysis software was validated against the AHA and MIT databases to measure QRS detection sensitivity, positive predictivity, and ventricular beat identification metrics. The algorithm was also tested against the European ST-T database to measure ST sensitivity, positive predictivity, and peak error. Software validation confirmed that ST-segment processing does not interfere with other system functions.

## Technological Characteristics

PC-based monitoring system using a Pentium processor and SVGA color monitor. Connectivity via RS-232, 900 MHz spread spectrum, or VHF (174-216 MHz). Features include a two-channel thermal recorder and optional printer. Software modification adds ST-segment measurement (level and slope) and alarm functionality. Methodology is consistent with the VitalCom VCOM Central Monitor.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- OBSERVER Central Station ([K933404](/device/K933404.md))
- VCOM Central Monitor ([K942147](/device/K942147.md))

## Reference Devices

- DINAMAP MPS Select Monitor ([K955113](/device/K955113.md))
- DINAMAP PLUS Monitor ([K943709](/device/K943709.md))
- DINAMAP PLUS Monitor ([K912188](/device/K912188.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K972498

## 510(k) Summary

(

## FEB | 7 1998

| Date           | November 18, 1997                                                                                                                                                                                                                                                                                            |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact        | Darlene T. Korab                                                                                                                                                                                                                                                                                             |
|                | Sr. Project Manager                                                                                                                                                                                                                                                                                          |
|                | Regulatory Affairs                                                                                                                                                                                                                                                                                           |
|                | Johnson & Johnson Medical, Inc.                                                                                                                                                                                                                                                                              |
|                | 4110 George Road                                                                                                                                                                                                                                                                                             |
|                | Tampa, FL 33634                                                                                                                                                                                                                                                                                              |
|                | Telephone: (813) 887-2663                                                                                                                                                                                                                                                                                    |
|                | Telefax: (813) 887-2263                                                                                                                                                                                                                                                                                      |
|                | E Mail: dkorab@crtus.jnj.com                                                                                                                                                                                                                                                                                 |
| Device Name    | • OBSERVER* Central Station                                                                                                                                                                                                                                                                                  |
|                | The subject of this premarket notification is the modification of the currently marketed Johnson & Johnson Medical, Inc. OBSERVER Central Station (K933404) to include ST-segment measurement.                                                                                                               |
| Common Name    | • Central Station                                                                                                                                                                                                                                                                                            |
| Classification | The classification name, 21 Code of Federal Regulations (CFR) Part and Paragraph number and classification of the OBSERVER Central Station with ST-segment measurement follow. The tier categorization based on the list (January 27, 1994) distributed by the Office of Device Evaluation is also included. |

Continued on next pagel

... ...

 .

{1}------------------------------------------------

| Classification<br>Name                                                       | 21 CFR Section | Class | Tier |
|------------------------------------------------------------------------------|----------------|-------|------|
| Monitor,<br>Electrocardio-<br>graphic                                        | 870.2360       | II    | 2    |
| Transmitters and<br>receivers,<br>physiological<br>signal,<br>radiofrequency | 870.2910       | II    | 2    |
| Detector and<br>alarm,<br>Arrhythmia                                         | 870.1025       | III   | 3    |

Classification name, 21 CFR section, class and tier are as follows: Classification, continued

#### Predicate Devices

The addition of ST-segment measurement to the currently marketed OBSERVER Central Station with arrhythmia analysis is substantially equivalent to the currently marketed OBSERVER Central Station with arrhythmia analysis which received market clearance February 9, 1994 via 510(k) K933404. Everything remains the same except for the addition of STsegment measurement. The added ST-segment measurement feature is substantially equivalent to the ST-segment measurement feature of the currently marketed VitalCom, Inc. VCOM Central Monitor which received market clearance on January 30, 1995 via VitalCom 510(k) K942147. The same firm - VitalCom, Inc. -- that is developing the ST-segment software for the OBSERVER Central Station developed the ST-segment software for the VitalCom, Inc. VCOM Central Monitor.

| Predicate Devices        | 510(k) Number |
|--------------------------|---------------|
| OBSERVER Central Station | K933404       |
| VCOM Central Monitor     | K942147       |

Continued on next page

{2}------------------------------------------------

ST-segment measurement is an optional feature of the OBSERVER Central Device Description Station with arrhythmia analysis. This feature allows for the measurement of the ST-segment level and slope and the issuance of alarms for ST elevations and depressions. The clinician may adjust certain waveform points and alarm priority and limits or rely on default settings for each patient to be analyzed.

> The OBSERVER Central Station (K933404) is a PC-based monitoring system designed to provide remote surveillance with alarms, trending and retrieval of wave-form and numeric physiological data for up to 8 patients who are monitored via various DINAMAP* Physiological Monitors or other appropriate bedside/patient monitors.

The OBSERVER Central Station displays, records and stores physiological data including ECG, non-invasive and invasive blood pressure, heart rate, temperature and pulse oximetry. The system is designed to be used with a hardwire interface using RS 232; wireless connectivity using 900 MHz spread spectrum or fixed frequency; or VHF (174-216 MHz, TV channels 7 through 13), including patient-worn telemetry. Monitors that may be networked with the OBSERVER Central Station for ECG include members of the Johnson & Johnson Medical Inc. DINAMAP* family of monitors, such as the DINAMAP MPS* Select* Monitor (K955113) and the DINAMAP* PLUS Monitor (K943709 and K912188), and patient-worn ECG VHF telemetry (VitalCom, Inc. K942147). The OBSERVER Central Station uses a Pentium PC with an SVGA, touch or non-touch, color monitor. Recordings can be made on either the built-in two-channel thermal recorder or with an optional HP LaserJet* Printer. Also optional is full disclosure (history) of all waveforms.

Continued on next page

*Trademark

{3}------------------------------------------------

Indications

The OBSERVER Central Station with ST-segment measurement is a prescription device intended for use only by health care professionals. The device is used for the remote monitoring of physiological parameters, including non-invasive and invasive blood pressure; invasive hemodynamic and intracranial pressures; oxygen saturation via pulse oximetry; temperature; ECG ; and pulse/heart rate of adult, pediatric and neonatal patients. The device is located at a distance from the patient but within the same facility, including settings such as hospital and outpatient services, including general medical/surgical, critical care, intermediate/step-down care, emergency room, radiology, labor and delivery, operating and recovery room, cardiac catheterization lab, endoscopy and same-day surgery.

Arrhythmia detection and ST-segment measurement are optional ECG features limited to the adult population. ST-segment measurement is contraindicated in paced patients. The ST-segment feature of the OBSERVER Central Station provides for the measurement of the ST-segment level and slope of the ECG waveform, alerting the clinician to ST-segment changes. The significance of ST-segment changes must be determined by a clinician. The device is not designed, sold or intended for use except as indicated.

#### Technological Characteristics

The OBSERVER Central Station with ST-segment measurement has the same technological characteristics as the predicate devices. The modification of the device software is an enhancement of the arrhythmia analysis feature to allow for the measurement of the ST segment with corresponding alarms. The methodology for ST-measurement is virtually the same as the VitalCom, Inc. VCOM Central Monitor.

Continued on next page

{4}------------------------------------------------

: 上一篇: :

វ្ 1

| Performance       | The ST-enabled arrhythmia analysis software was tested against the AHA and MIT databases to measure the algorithm's QRS detection sensitivity and positive predictivity and ventricular beat identification sensitivity, positive predictivity and false positive rate.                                                                                                     |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                   | The algorithm was also tested against the European ST-T database to measure the algorithm's ST sensitivity, positive predictivity and peak error.                                                                                                                                                                                                                           |
|                   | Software validation was performed to show that the OBSERVER Central Station correctly processes ST information, while not affecting other portions of the system.                                                                                                                                                                                                           |
| Conclusions       | In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this premarket notification, Johnson & Johnson Medical concludes that the modified device, the OBSERVER* Central Station with ST-segment measurement is safe, effective and substantially equivalent to the predicate devices as described herein. |
| Other Information | Johnson & Johnson Medical, Inc. will update and include in this summary any other information deemed reasonably necessary by the FDA.                                                                                                                                                                                                                                       |

. •

·

:

ーディ

: : :

:

:

{5}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 17 1998

Ms. Darlene T. Korab Johnson & Johnson Medical Inc. 4110 George Road Tampa, FL 33634

K972498 Re : OBSERVER* Central Station Regulatory Class: III (Three) Product Code: 74 DSI Dated: November 18, 1997 Received: November 19, 1997

Dear Ms. Korab:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to Comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Image /page/5/Picture/9 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Thomas J. Callshan

Thomas J. Callahan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

#### INDICATIONS FOR USE

Page 1 of 1

510(k) Number (if known): K972498

Device Name: OBSERVER* Central Station

Indications for Use:

(

The OBSERVER Central Station with ST-segment measurement is a prescription device intended for use only by health care professionals. The device is used for the remote monitoring of physiological parameters, including non-invasive and invasive blood pressure; invasive hemodynamic and intracranial pressures; oxygen saturation via pulse oximetry; temperature; ECG ; and pulse/heart rate of adult, pediatric and neonatal patients. The device is located at a distance from the patient but within the same facility, including settings such as hospital and outpatient services, including general medical/surgical, critical care, intermediate/step-down care, emergency room, radiology, labor and delivery, operating and recovery room, cardiac catheterization lab, endoscopy and same-day surgery.

Arrhythmia detection and ST-segment measurement are optional ECG features limited to the adult population. ST-segment measurement is contraindicated in paced patients. The ST-segment feature of the OBSERVER Central Station provides for the measurement of the ST-segment level and slope of the ECG waveform, alerting the clinician to STsegment changes. The significance of ST-segment changes must be determined by a clinician. The device is not designed, sold or intended for use except as indicated. _

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

==============================================================================================================================================================================

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ﺮ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤ (per 21 CFR 801.109

OR

Over-The Counter Use -------

(Optional Format 1-2-96)

*Trademark

---

**Source:** [https://fda-staging.innolitics.com/device/K972498](https://fda-staging.innolitics.com/device/K972498)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).