← Product Code [CGA](/productcode/CGA) · K972473

# ONE TOUCH II HOSPITAL BLOOD GLUCOSE MONITORING SYSTEM DATA DOCK (K972473)

_Lifescan, Inc. · CGA · Dec 16, 1997 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K972473

## Device Facts

- **Applicant:** Lifescan, Inc.
- **Product Code:** [CGA](/productcode/CGA.md)
- **Decision Date:** Dec 16, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, and the DataDock and other optional accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring blood glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal blood glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.

## Device Story

DataDock is an optional accessory for the ONE TOUCH® II Hospital Blood Glucose Monitoring System; functions as a docking station for the meter. Improves ease-of-use and enhances data management capabilities for the existing system. Used in hospital, clinic, and home settings by healthcare providers or patients. Facilitates data transfer and management; output used by clinicians to monitor blood glucose levels and inform clinical decision-making. Benefits include streamlined data handling and improved workflow for glucose monitoring.

## Clinical Evidence

No clinical data provided; device is an accessory to a previously cleared system.

## Technological Characteristics

Docking station accessory for blood glucose meter; provides data management and ease-of-use features. No changes to the core glucose oxidase sensing technology of the primary meter.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K972473

# 510(k) Summary

DEC I S 1997

| Submitter           | LifeScan, Inc.<br>1000 Gibraltar Drive<br>Milpitas, CA 95037<br>John E. Hughes |
|---------------------|--------------------------------------------------------------------------------|
| Date                | July 1, 1997                                                                   |
| Proprietary Name    | ONE TOUCH® II Hospital Blood Glucose<br>Monitoring System<br>DataDock          |
| Common Name         | Blood Glucose Monitor                                                          |
| Classification Name | 75CGA  Glucose oxidase, glucose                                                |

## Device Description

The DataDock is an optional accessory for the ONE TOUCH® II Hospital Meter. When the ONE TOUCH® II Hospital Meter is inserted into the DataDock, it provides features which improve ease-of-use and enhance data management capability.

## Intended Use

The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, and the DataDock and other optional accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring blood glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal blood glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.

### Substantial Equivalence

The DataDock is an optional accessory modifying the ONE TOUCH® II Hospital Blood Glucose Monitoring System currently in commercial distribution. The modification neither significantly changes the safety and effectiveness nor makes any change to its intended use of the device.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 6 1997

John E. Hughes Manager, Regulatory Affairs LifeScan Inc. 1000 Gibraltar Drive Milpitas, California 95035-6312

K972473 Re : ONE TOUCH II DataDock Hospital Blood Glucose Monitoring System Regulatory Class: II Product Code: CGA October 1, 1997 Dated: October 3, 1997 Received:

Dear Mr. Hughes:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:

this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Sitman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

K972473

# 510(k) Summary

DEC I S 1997

| Submitter           | LifeScan, Inc.<br>1000 Gibraltar Drive<br>Milpitas, CA 95037<br>John E. Hughes |
|---------------------|--------------------------------------------------------------------------------|
| Date                | July 1, 1997                                                                   |
| Proprietary Name    | ONE TOUCH® II Hospital Blood Glucose<br>Monitoring System<br>DataDock          |
| Common Name         | Blood Glucose Monitor                                                          |
| Classification Name | 75CGA  Glucose oxidase, glucose                                                |

## Device Description

The DataDock is an optional accessory for the ONE TOUCH® II Hospital Meter. When the ONE TOUCH® II Hospital Meter is inserted into the DataDock, it provides features which improve ease-of-use and enhance data management capability.

## Intended Use

The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, and the DataDock and other optional accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring blood glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal blood glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.

### Substantial Equivalence

The DataDock is an optional accessory modifying the ONE TOUCH® II Hospital Blood Glucose Monitoring System currently in commercial distribution. The modification neither significantly changes the safety and effectiveness nor makes any change to its intended use of the device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 6 1997

John E. Hughes Manager, Regulatory Affairs LifeScan Inc. 1000 Gibraltar Drive Milpitas, California 95035-6312

K972473 Re : ONE TOUCH II DataDock Hospital Blood Glucose Monitoring System Regulatory Class: II Product Code: CGA October 1, 1997 Dated: October 3, 1997 Received:

Dear Mr. Hughes:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note:

this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Sitman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

### Indications for Use Form

510(k) Number (if known)

972473

Device Name:

ONE TOUCH® II Hospital Blood Glucose Monitoring System DataDock

Indications for Use

The ONE TOUCH® II HOSPITAL Blood Glucose Monitoring System, consisting of the ONE TOUCH® II Hospital Meter, ONE TOUCH® Test Strips, control solutions, data management software, the optional DataDock and other accessories is intended for quantitative measurement of glucose in whole blood in hospital, clinic, and home settings. It is intended for monitoring glucose levels in neonatal or non-neonatal whole blood samples. Non-neonatal glucose results can be reported in either whole blood equivalents or plasma/serum equivalents.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(per 21 CFR 801.109) ✓

OR

Over-The-Counter Use (Optional Format 1-2-96)

-
(Division Sign-Off)
Division of Clincal Laboratory Devices,
510(k) Number. 972473 134

---

**Source:** [https://fda-staging.innolitics.com/device/K972473](https://fda-staging.innolitics.com/device/K972473)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).