HEARTSTRAMELECTRODE ADAPTER(EA-XX SERIES)/HEARTSTRAM DEFIBRILLATION PADS WITH PREATTACHED ELECTRODE ADAPTER(DP-EAXX)
Applicant
Heartstream, Inc.
Product Code
MKJ · Cardiovascular
Decision Date
Sep 25, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
Heartstream Electrode Adapter: For use with Heartstream Defibrillators or Manual Defibrillators with a Quick-Combo® Connection System. For Defibrillation, Monitoring and Pacing. Heartstream External Defibrillation Pads with Preattached Electrode Adapter: For use with Heartstream Defibrillators or Manual Defibrillators with a Quick-Combo® Connection System. For Defibrillation, Monitoring and Pacing.
Device Story
Heartstream electrode adapters serve as interface between Heartstream electrodes and various manual defibrillators; facilitate external cardiovascular pacing, monitoring, and defibrillation shocks up to 360J. Device consists of rigid thermoplastic connectors joined by insulated conductors/cables; available as reusable stand-alone accessories or preattached to single-use electrode sets. Operated by clinicians in clinical settings. Adapters ensure secure electrical connection between defibrillator socket and electrode proximal end; maintain signal integrity for pacing/monitoring/defibrillation. Benefit includes enabling compatibility between Heartstream electrodes and diverse manual defibrillator systems, ensuring effective energy delivery and patient safety.
Clinical Evidence
Bench testing only. Compliance with IEC 601-1, IEC 601-2-4, IEC 601-2-25, and ANSI/AAMI DF39-1993 demonstrated mechanical and electrical compatibility, effective energy delivery up to 360J, and capability for external transcutaneous cardiac pacing.
Technological Characteristics
Rigid thermoplastic connectors; insulated conductors/cables. Designed for electrical/mechanical compatibility with manual defibrillators. Complies with IEC 601-1 (safety), IEC 601-2-4 (defibrillators), IEC 601-2-25 (ECG), and ANSI/AAMI DF39-1993 (AEDs).
Indications for Use
Indicated for patients requiring external cardiovascular pacing, ECG monitoring, and delivery of defibrillation shocks up to 360J via manual defibrillators.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "HEARTSTREAM" in all caps, with a stylized heart shape to the left of the word. The heart shape is formed by two curved lines that come together at the bottom. The text is in a bold, sans-serif font. The image is in black and white.
Image /page/0/Picture/1 description: The image shows the text "K972418 SEP 25". The text appears to be handwritten in a dark ink. The date "SEP 25" is located below and to the left of the number "K972418".
# 510(k) Summary Heartstream Electrode Adapters
| General Information | |
|------------------------------|-----------------------------------------------------------------------|
| Trade Name | Heartstream Defibrillation Pads with<br>Preattached Electrode Adapter |
| | Heartstream Electrode Adapter |
| FDA Panel and Classification | Cardiovascular, 74 DRO, Class III<br>MKJ |
| Contact Person | Cindy Pestka<br>Regulatory Affairs Manager |
| Address | Heartstream, Inc.<br>2401 4th Ave.<br>Suite 300<br>Seattle, WA 98121 |
## Substantially Equivalent Devices
| Manufacturer | Product |
|-------------------|------------------------------------------|
| Heartstream, Inc. | Heartstream External Defibrillation Pads |
| Katecho, Inc. | K-Defib/Pace Multifunction Electrodes |
| Darox | R2 series of adapters and cables |
#### Description of Device & Intended Use
Heartstream electrode adapters serve as an interface to allow Heartstream electrodes to be connected to various manual defibrillators for external cardiovascular pacing as well as monitoring and delivery of defibrillation shocks up to 360J.
Heartstream electrode adapters are made of rigid thermoplastic material (or equivalent), with varying configurations depending upon the manual defibrillator for which they are designed. The adapters will be provided as reusable stand-alone accessories (to be used in conjunction with standard Heartstream electrodes) or preattached to a Heartstream electrode set, for single use.
{1}------------------------------------------------
## Technological Characteristics
All Heartstream electrode adapters are designed to comply with applicable portions of relevant standards. including:
- IEC 601-1, Medical electrical equipment, Part 1: General requirements for safety, 1993 .
- IEC 601-2-4, Medical electrical equipment, Part 2: Particular Requirements for the Safety of Cardiac . Defibrillators and Cardiac Defibrillator/Monitors, 1993
- IEC 601-2-25, Medical electrical equipment, Part 2: Particular requirements for the safety of . electrocardiographs
- ANSVAAMI DF39-1993. Automatic external defibrillators and remote-control defibrillators. . September 16, 1993
Testing per these standards demonstrated that the Heartstream electrodes effectively deliver defibrillation energy up to 360J, allow for external transcutaneous cardiac pacing, demonstrate mechanical and electrical compatibility with manual defibrillators, and minimize safety risks to the user and patient.
### Summary of Substantial Equivalence
Heartstream electrode adapters were evaluated in terms of indications for use, technological characteristics and materials, and compared to equivalent devices currently on the market.
The Heartstream electrode adapter is very similar in form, fit and function to other comparable devices. Marketed adapters and adapter cables typically gonsist of two connectors (or a double-sided connector) joined by an insulated conductor or cable of varying length. One connector (or one side of the connector) fits into the connector socket of the defibrillator, while the other attaches to the proximal end of the electrode set. The connectors are typically made of a rigid thermoplastic material. Adapters and adapter cables are designed so that both connectors are firmly seated when in use, and so they, do not expose a user to a risk of accidental electrical contact. They are also designed to ensure that all pacing, monitoring and/or defibrillation functions are not adversely affected by their placement. These characteristics apply to both the stand-alone Heartstream electrode adapter and the Heartstream electrodes with a preattached adapter.
Heartstream electrodes with a preattached electrode adapter are substantially equivalent to the standard Heartstream electrodes regarding device design and usage for defibrillation, and Katecho K-Defib/Pace multifunction electrodes regarding device design and usage for defibrillation, monitoring and pacing.
Heartstream stand-alone electrode adapters are substantially equivalent to the Darox R2 series of adapter cables regarding device design and use as an interface between electrodes and defibrillators for defibrillation, monitoring and pacing, for use with numerous defibrillators.
Therefore, due to the similarity of design features, and the similarity of the indicated use to other predicate devices, Heartstream, Inc. believes this product does not raise any new safety or effectiveness issues.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with its head facing left and its wings represented by abstract lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 5 1997
Ms. Lori Glastetter Heartstream, Inc. 2401 4" Avenue, Suite 300 Seattle, Washington 98121
Re: K972418 Heartstream Electrode Adapters Regulatory Class: III (three) Product Code: 74 MKJ Dated: June 26, 1997 Received: June 27, 1997
Dear Ms. Pestka:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Paqe 2 - Ms. Lori Glastetter
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
. telephone 206.443.7630 facsimile 206.443.9694
Image /page/4/Picture/1 description: The image shows the logo for Heartstream. The logo consists of a stylized heart shape on the left, with two curved lines forming the upper part of the heart. To the right of the heart shape is the word "HEARTSTREAM" in all capital letters.
2401 Fourth Avenue, Suite 300 · Seattle, Washington 98121.1436 USA
Page 1 of 1 Submitted: July 22, 1997
- K972418 510(k) Number:
Heartstream Electrode Adapter, and Device Names:
> Preattached Heartstream External Defibrillation Pads with Electrode Adapter
- Heartstream Electrode Adapter: For use with Heartstream Indications for Use: Defibrillators or Manual Defibrillators with a Quick-Combo® Connection System. For Defibrillation, Monitoring and Pacing.
Heartstream External Defibrillation Pads with Preattached Electrode Adapter: For use with Heartstream Defibrillators or Manual Defibrillators with a Quick-Combo® Connection System. For Defibrillation, Monitoring and Pacing.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Atl. A. Cidl.
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
OR
510(k) Number
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.