The device is intended to be moved in a hospital or clinic or other facility to provide capability of X-ray exposures in remote areas, or in instances where stationary devices are unsuitable. The device is designed to be used in conjunction with a cassette film or other image receptor, in order to provide a radiological exposure of the human body, to serve as an aide in diagnosing illnesses and fractures. The unit is not intended to treat, prevent, cure, mitigate or otherwise medicate ailments.
Device Story
Mobile X-ray unit (Compact/Comp-X) provides diagnostic imaging in hospitals/clinics; used where stationary units are unsuitable. Operator (technician/physician) positions unit via wheeled base and articulated arm (45-inch extension, ±90-degree pivot, ±180-degree gimbals). System uses X-ray tube with rotating anode (40-120 kVp) and Beam Limiting Device (BLD). Operator selects kVp/mAs settings; two-step push-button initiates 'prep' (filament warming, anode spinning at 3000rpm) followed by high-voltage exposure. Inverter transforms mains current to high-frequency, medium-voltage, then to high-voltage via monoblock generator. Controller monitors parameters; aborts exposure if errors occur or if button released. Provides radiological exposure on film/receptor to aid diagnosis of fractures/illnesses. Benefits include portability for remote/bedside imaging.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Mobile X-ray unit; zinc-coated metal mechanics; lead (Pb) radiation shielding. High-frequency generator; rotating anode X-ray tube. Input: 40-120 kVp, 0.5-100 mAs. Connectivity: manual control via 13ft corded push-button. Exposure time: 5ms to 4s. Mechanical: triangular base, 4-wheel configuration, articulated arm with gimbals.
Indications for Use
Indicated for general radiographic procedures on patients requiring diagnostic imaging of illnesses and fractures.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
5972365
#### Summary of the safety and effectiveness information in the premarket notification (510(k) Summary)
DEC 1 2 1997
Submitted by: Eurotec, srl Address: Via Aldo Moro 5/7, 24020 Scanzorosciate (Bergamo) Italy Telephone: (++39.35) 655-557 Contact Person: Christian Bonazzi (Official Correspondent, Technical Director of Electronic Design Department) 510(k) Number: K972365 Owner Operator ID Number: 9030797
This Summary was prepared on August 23, 1997
(1) This 510(k) summary is part of the premarket notification documentation for the unit "Compact" or "Comp-X", an X-ray mobile unit intended to be moved in a hospital or clinic or other facility to provide capability of X-ray exposures in remote areas, or in instances where stationary devices are unsuitable. Its classification name is X-ray mobile unit (RAD only). Its classification number is IZK.
(2) The device for which this 510(k) Summary is prepared is substantially equivalent to the model MU125, manufactured by Shimadzu (510(k) number: K955036) . This device was found to be substantially equivalent through the 510(k) premarket notification process.
(3) The device that is subject to this 510(k) premarket notification document will be labeled on all promotional material and within all documents as "X-ray mobile unit with HF generator". The device appears as a mobile unit and is composed of three parts:
The X-ray generator (monoblock) is constituted of an X-ray tube with rotating anode, encased in metal casing which also contains the high voltage generator. High voltage is required by the X-ray tube to generate a usable X-ray beam. The X-ray tube emits radiations when supplied with between 40,000 and 120,000 Volts. The monoblock is attached in a non-permanent manner to a Beam Limiting Device (BLD), which will limit the X-ray beam exiting the window of the monoblock by means of adjustable shutters. The combination of monoblock and BLD create the X-ray source.
The monoblock and the X-ray emission are controlled by an X-ray controller an electronic device connected to the monoblock through low tension cables.
X-ray generator and X-ray controller are enclosed in mechanics which allow the device to be moved on wheels, as well as the X-ray generator to be positioned as required by the operator.
The material used are metal and radiation insulating material (lead (Pb() for the X-ray head (monoblock and BLD). The mechanics are constructed of zinc coated metal.
{1}------------------------------------------------
The device is designed to be transported on wheels: the main structure contains the electronics and sits on a triangular base with two set of wheels; two larger wheels are located at the rear of the base, while two smaller wheel (nose gear) are located at the front of the device. The top of the main structure is equipped with a column (perpendicular to the ground), which supports the arm structure; this (at the end of which is located the X-ray source) extends for 45 inches and allows vertical movements (up and down) and pivotal movements. The latter is achieved by moving the arm around its column for a total of ±90 degrees.
The X-ray source is attached to the arm by mean of gimbals, which allow z-axis rotation and x-axis rotation for a total of ±180 degrees each.
The BLD is also capable of movements, swiveling on its vertical axis for a total of ±90 degrees.
The device is designed to perform X-ray exposures by means of a two-step push button (attached to the unit by mean of a 13ft coiled cord). The operator chooses the preferred setting by selecting kVp (ranging from 40 to 120) and the mAs (ranging from 0.5 to 100). After selecting the appropriate settings and positioning the device, the operator will initiate exposure by 'prepping' the exposure. This is achieved by pressing the first step of the push button. This step will feed the appropriate mA (current) to the filament of the tube, thus warming it, and will begin spinning the anode at 3000rpm At this stage, however, there is no radiation emission, since the tube is not fed with high voltage current. Emission will begin once the unit has prepped, and will prompt (by means of a led located on the control panel) the operator to push the second step of the push-button. At this point, the current drawn from the mains and stored by capacitors will be released in one-350 Vac shot, which will be directed to the electronics and processed by an Inverter. The inverter fulfills the duty of transforming this current in High Frequency 350 Vac shot, which will be directed in this form and fed to the monoblock. This high frequency, medium-voltage current is transformed to high voltage current through a HV generator located in the monoblock. The generator will create the high voltage technique (40 to 120 kVp) set by the operator. Only after the HV reaches this point (and after all settings and parameters have been controlled and stabilized by the X-ray controller on their way to the tube) the tube emits its radiation is emitted for a time which varies, according to the technique pre-set by the operator, between 5 milliseconds and 4 seconds.
Any problem encountered during the control by the X-ray controller, will result in an abortion of the exposure and an indication (both visual and audible) of such action.
Any exposure can also be manually aborted by the operator by releasing the push-button before the emission is terminated. This action will also result in an audible and visual signal to the operator.
After an exposure is terminated, the operator can perform another exposure after 8 to 10 seconds.
2
{2}------------------------------------------------
#### (4) INTENDED USE STATEMENT
The device is intended to be moved in a hospital or clinic or other facility to provide capability of X-ray exposures in remote areas, or in instances where stationary devices are unsuitable. The device is designed to be used in conjunction with a cassette film or other image receptor, in order to provide a radiological exposure of the human body, to serve as an aide in diagnosing illnesses and fractures. The unit is not intended to treat, prevent, cure, mitigate or otherwise medicate ailments.
(5) In brief, the device has the same technological characteristics and uses as the predicate devices stated in paragraph (2) of this Summary, in that all devices referenced in paragraph (2) are X-ray mobile units, which move on wheels, composed of a main structure, containing electronics for the control of the radiographic exposures, used within hospitals or other structures to provide capability of radiographic exposures used in the diagnosis of ailments. Furthermore, all predicate devices are furnished with an Xray head, a high tension transformer to supply the tube with the necessary high tension, and a BLD to limit the projection of the X-ray beam as well as to give a visual indication of the area where the beam is going to be projected. These devices might, however, present some minor differences. These differences might be:
- k Vp range: might slightly vary from device to device I.
- II. focal spot: might vary from device to device depending on the maximum kW output of the unit
- III. mAs range: might vary from device to device
- IV. Shortest exposure time: might vary depending on the technique settings range of a specific device
- V. motorized movements: some devices might have motorized movements in light of their weight
- VI. operating frequency: some devices might operate on low frequency, which implies a lower quality image and longer X-ray emission
- VII.mechanical movements: some devices might be equipped with mechanical movements that differ from other devices (i.e. telescopic movement of the Xray head).
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 2 1997
Christian Bonazzi Official Correspondent c/o Eurotec, srl Via Aldo Moro 5/7 Scanzorosciate (Bergamo) 24020 Italy
Re: K972365
Compact or Comp-X (Mobile X-Ray Unit) Dated: September 25, 1997 Received: September 26, 1997 Regulatory Class: II 21 CFR 892.1720/Procode: 90 IZL
Dear Mr. Bonazzi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your deving regulation (21-CFR Part 80t and additionally 809.10 for in viro ' diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at.(301) 594-4639. Also, please note the regulation entitled, " "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W.Liliau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
5972365
#### Summary of the safety and effectiveness information in the premarket notification (510(k) Summary)
DEC 1 2 1997
Submitted by: Eurotec, srl Address: Via Aldo Moro 5/7, 24020 Scanzorosciate (Bergamo) Italy Telephone: (++39.35) 655-557 Contact Person: Christian Bonazzi (Official Correspondent, Technical Director of Electronic Design Department) 510(k) Number: K972365 Owner Operator ID Number: 9030797
This Summary was prepared on August 23, 1997
(1) This 510(k) summary is part of the premarket notification documentation for the unit "Compact" or "Comp-X", an X-ray mobile unit intended to be moved in a hospital or clinic or other facility to provide capability of X-ray exposures in remote areas, or in instances where stationary devices are unsuitable. Its classification name is X-ray mobile unit (RAD only). Its classification number is IZK.
(2) The device for which this 510(k) Summary is prepared is substantially equivalent to the model MU125, manufactured by Shimadzu (510(k) number: K955036) . This device was found to be substantially equivalent through the 510(k) premarket notification process.
(3) The device that is subject to this 510(k) premarket notification document will be labeled on all promotional material and within all documents as "X-ray mobile unit with HF generator". The device appears as a mobile unit and is composed of three parts:
The X-ray generator (monoblock) is constituted of an X-ray tube with rotating anode, encased in metal casing which also contains the high voltage generator. High voltage is required by the X-ray tube to generate a usable X-ray beam. The X-ray tube emits radiations when supplied with between 40,000 and 120,000 Volts. The monoblock is attached in a non-permanent manner to a Beam Limiting Device (BLD), which will limit the X-ray beam exiting the window of the monoblock by means of adjustable shutters. The combination of monoblock and BLD create the X-ray source.
The monoblock and the X-ray emission are controlled by an X-ray controller an electronic device connected to the monoblock through low tension cables.
X-ray generator and X-ray controller are enclosed in mechanics which allow the device to be moved on wheels, as well as the X-ray generator to be positioned as required by the operator.
The material used are metal and radiation insulating material (lead (Pb() for the X-ray head (monoblock and BLD). The mechanics are constructed of zinc coated metal.
{5}------------------------------------------------
The device is designed to be transported on wheels: the main structure contains the electronics and sits on a triangular base with two set of wheels; two larger wheels are located at the rear of the base, while two smaller wheel (nose gear) are located at the front of the device. The top of the main structure is equipped with a column (perpendicular to the ground), which supports the arm structure; this (at the end of which is located the X-ray source) extends for 45 inches and allows vertical movements (up and down) and pivotal movements. The latter is achieved by moving the arm around its column for a total of ±90 degrees.
The X-ray source is attached to the arm by mean of gimbals, which allow z-axis rotation and x-axis rotation for a total of ±180 degrees each.
The BLD is also capable of movements, swiveling on its vertical axis for a total of ±90 degrees.
The device is designed to perform X-ray exposures by means of a two-step push button (attached to the unit by mean of a 13ft coiled cord). The operator chooses the preferred setting by selecting kVp (ranging from 40 to 120) and the mAs (ranging from 0.5 to 100). After selecting the appropriate settings and positioning the device, the operator will initiate exposure by 'prepping' the exposure. This is achieved by pressing the first step of the push button. This step will feed the appropriate mA (current) to the filament of the tube, thus warming it, and will begin spinning the anode at 3000rpm At this stage, however, there is no radiation emission, since the tube is not fed with high voltage current. Emission will begin once the unit has prepped, and will prompt (by means of a led located on the control panel) the operator to push the second step of the push-button. At this point, the current drawn from the mains and stored by capacitors will be released in one-350 Vac shot, which will be directed to the electronics and processed by an Inverter. The inverter fulfills the duty of transforming this current in High Frequency 350 Vac shot, which will be directed in this form and fed to the monoblock. This high frequency, medium-voltage current is transformed to high voltage current through a HV generator located in the monoblock. The generator will create the high voltage technique (40 to 120 kVp) set by the operator. Only after the HV reaches this point (and after all settings and parameters have been controlled and stabilized by the X-ray controller on their way to the tube) the tube emits its radiation is emitted for a time which varies, according to the technique pre-set by the operator, between 5 milliseconds and 4 seconds.
Any problem encountered during the control by the X-ray controller, will result in an abortion of the exposure and an indication (both visual and audible) of such action.
Any exposure can also be manually aborted by the operator by releasing the push-button before the emission is terminated. This action will also result in an audible and visual signal to the operator.
After an exposure is terminated, the operator can perform another exposure after 8 to 10 seconds.
2
{6}------------------------------------------------
#### (4) INTENDED USE STATEMENT
The device is intended to be moved in a hospital or clinic or other facility to provide capability of X-ray exposures in remote areas, or in instances where stationary devices are unsuitable. The device is designed to be used in conjunction with a cassette film or other image receptor, in order to provide a radiological exposure of the human body, to serve as an aide in diagnosing illnesses and fractures. The unit is not intended to treat, prevent, cure, mitigate or otherwise medicate ailments.
(5) In brief, the device has the same technological characteristics and uses as the predicate devices stated in paragraph (2) of this Summary, in that all devices referenced in paragraph (2) are X-ray mobile units, which move on wheels, composed of a main structure, containing electronics for the control of the radiographic exposures, used within hospitals or other structures to provide capability of radiographic exposures used in the diagnosis of ailments. Furthermore, all predicate devices are furnished with an Xray head, a high tension transformer to supply the tube with the necessary high tension, and a BLD to limit the projection of the X-ray beam as well as to give a visual indication of the area where the beam is going to be projected. These devices might, however, present some minor differences. These differences might be:
- k Vp range: might slightly vary from device to device I.
- II. focal spot: might vary from device to device depending on the maximum kW output of the unit
- III. mAs range: might vary from device to device
- IV. Shortest exposure time: might vary depending on the technique settings range of a specific device
- V. motorized movements: some devices might have motorized movements in light of their weight
- VI. operating frequency: some devices might operate on low frequency, which implies a lower quality image and longer X-ray emission
- VII.mechanical movements: some devices might be equipped with mechanical movements that differ from other devices (i.e. telescopic movement of the Xray head).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 2 1997
Christian Bonazzi Official Correspondent c/o Eurotec, srl Via Aldo Moro 5/7 Scanzorosciate (Bergamo) 24020 Italy
Re: K972365
Compact or Comp-X (Mobile X-Ray Unit) Dated: September 25, 1997 Received: September 26, 1997 Regulatory Class: II 21 CFR 892.1720/Procode: 90 IZL
Dear Mr. Bonazzi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your deving regulation (21-CFR Part 80t and additionally 809.10 for in viro ' diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at.(301) 594-4639. Also, please note the regulation entitled, " "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
W.Liliau Yu
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{8}------------------------------------------------
Page 1 of
510(k) Number (if known): __x972365
Device Name: Company of Comp. X MOBILE X AN UNIT
THE UNIT IS INTENDED FOR USE ON Indications For Use: GENERAL RADIOGRAPHIC PROCEDURES.
> CHRISTIAN BONAZZI OFFICIAL CORRESPONDENT
Bonali Illan
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|---------------------|--|
|---------------------|--|
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
| 510(k) Number | K972365 |
|---------------|---------|
|---------------|---------|
Prescription Use_X (Per 21 CFR 801.109) OR
Over-The-Counter Use______________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.