K972353 · Ercons, Inc. · LKY · Oct 28, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K972353
Device Name
ERCO-RIBBON
Applicant
Ercons, Inc.
Product Code
LKY · Gastroenterology, Urology
Decision Date
Oct 28, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Erco-Ribbon™ is used to maintain penile rigidity in men with erectile dysfunction by restricting penile venous outflow after the patient has obtained an erection (either with the aid of a vacuum pump or naturally). Erco-Ribbon can be used for patients with following conditions: a) separately: premature ejaculation firm but short lasting erection b) as a part of a vacuum constriction system Diabetes Venous leakage Prostatectorny Hypertension Psychogenic condition Radiation therapy Spinal cord injury
Device Story
Erco-Ribbon is a penile constriction device consisting of an elastic silicone ribbon with a rectangular cross-section and a belt with a release loop. It is used by patients to maintain penile rigidity by restricting venous outflow. The device is applied by wrapping the ribbon multiple times around the base of the penis or a vacuum chamber; tension is adjusted by the user during wrapping. The release loop allows for quick removal. Unlike traditional solid rubber rings that require cone-shaped applicators and fixed sizing, the Erco-Ribbon offers adjustable pressure through variable tension and number of turns. It is intended for use in clinical or home settings to assist in managing erectile dysfunction and related conditions. The device provides a mechanical alternative to standard constriction rings, aiming to reduce discomfort, numbness, and bruising associated with excessive or uneven pressure.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Device consists of elastic, soft silicone ribbon with rectangular cross-section and acrylic mounting tube. Operates via mechanical constriction; tension is manually adjusted by the user through multiple wraps. No energy source or software components. Materials conform to FDA regulations for skin-contacting devices.
Indications for Use
Indicated for men with erectile dysfunction, premature ejaculation, or firm but short-lasting erections. Applicable to patients with diabetes, venous leakage, post-prostatectomy, hypertension, psychogenic conditions, radiation therapy history, or spinal cord injury. Used independently or as part of a vacuum constriction system.
Regulatory Classification
Identification
External penile rigidity devices are devices intended to create or maintain sufficient penile rigidity for sexual intercourse. External penile rigidity devices include vacuum pumps, constriction rings, and penile splints which are mechanical, powered, or pneumatic devices.
Special Controls
*Classification.* Class II (special controls). The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: External Penile Rigidity Devices.” See § 876.1(e) for the availability of this guidance document.
{0}------------------------------------------------
ERCONS, Inc.
•••
66 Overlook Terrace, suite 2E New York, NY, 10040 Tel. 1-212-927 3275 Fax. 1-212-927 7387
Ka72353
# ersum16s
#### 510(k) SUMMARY
for
OCT 2 8 1997
# ERCO-RIBBON™
Constriction Device for External Penile Rigidity System
Submitter: Dr. Yakov Altshuler, Vice President
Date: July 3, 1997
Trade name - Erco-Ribbon™
Common name - constriction device for external penille rigidity system
Equivalence is claimed to:
| 1. ErecAid | Osbon Medical Systems, Ltd. | K841257 | |
|------------------------------------------|-----------------------------|---------|--------------|
| 2. Revive | Revive System Corp. | K844445 | |
| 3. Confidence | Performance, Inc. | K891125 | |
| 4. VED | Mission Pharmacal Co. | K901223 | |
| 5. VET | VET-CO, Inc. | K902240 | |
| 3. Pos-T-Vac | POS-T-VAC | K960828 | |
| Code of Federal Regulation (CFR) Number: | | | Unclassified |
| Product Code: | | | 78 LKY |
Ercons, Inc.
PATENT PENDING
Erco-Ribbon™
CONFIDENTIAL
{1}------------------------------------------------
## Intended Use
Erco-Ribbon™ is used to maintain penile rigidity in men with erectile dysfunction by restricting penile venous outflow after the patient has obtained an erection (either with the aid of a vacuum pump or naturally).
Erco-Ribbon can be used for patients with following conditions:
a) separately:
premature ejaculation
firm but short lasting erection
b) as a part of a vacuum constriction system
Diabetes
Venous leakage
Prostatectorny
Hypertension
Psychogenic condition
Radiation therapy
Spinal cord injury
Ercons, Inc.
PATENT PENDING
Erco-Ribbon™
CONFIDENTIAL
{2}------------------------------------------------
#### DESCRIPTION
A constriction ribbon is made of elastic and soft silicone with generally rectangular cross section. A belt with release loop is used for keeping the constrictor in a prearranged tightened position and for unlocking the ribbon .
The material conforms FDA regulation for devices contacting human skin.
For unlocking of the constrictor. the belt's release loop is pulled forward. The belt slips off, allowing easy unwrapping of the constricting ribbon.
A mounting tube comprises a segment of acrylic tube.
For arrangement of Erco-Ribbon™ penile constriction device, the ribbon is wrapped with the tension by multyple turns over the segment of the proximal end of the mounting tube. (or a vacuum chamber, when Erco-Ribbon ™ is used as a part of a vacuum constriction system),
#### SUBSTANTIAL EQUIVALENCY COMPARISON
Comparison of design and performance of marketable constriction rings and Erco-Ribbon is made with consideration of the impact of device's design on its safety and effectiveness .
marketable Predicate devices and Erco-Ribbon™ are constriction Considered with the same intended use, but have different technological devices characteristics.
Ercons, Inc. Inc. Come
PATENT PENDING
Erco-Ribbon™
CONFIDENTIAL
{3}------------------------------------------------
Predicate devices:
To provide sufficient inward pressure on erected penis, one or more of constriction rings have to be placed at the edge of an open end of the vacuum chamber.
Rings are molded of natural or synthetic rubber with different durometer numbers, usually in the range of 30 to 60 and are manufactured in large, medium and small sizes. To provide acceptable inward pressure, selection for individual users is based on size, durometer and a number of constriction rings used together. Despite high cost due to a number of expensive molds, there is no way to provide smooth control of the pressure which happens to be excessive and causes discomfort, numbness, bruises.
Placement of rings presents certain technical difficulties. Recognizing this problem, special cone-shaped applicators are added to the system. Applicators facilitate the problem, but complicate the system and increase its cost.
Discomfort and pain may be caused by twisting of doubled rings during their transfer onto the penis and because of intertwining with pubic hair.
Removal of constriction ring from erected penis could be painful too, especially when two or more rings are used together. In this case after removal of the first ring the penis is still engorged, as the remaining ring prevents blood outflow.
Erco-Ribbon ™:
There are countless possibilities of providing proper pressure of the constricting device during fabrication: by selection of constricting ribbon's material, cross section, length.
Erco-Ribbon™ PATENT PENDING Ercons, Inc. CONFIDENTIAL
4
{4}------------------------------------------------
With the constrictor of given properties and dimensions, the user may prearrange desirable pressure by changing the tension and number of turns during wrapping.
The present design radically reduces major technical difficulties of constriction ring associated with its placement on the vacuum chamber. A multyturn constriction ribbon is much easier to apply than a solid constriction ring as the force necessary to extend multyturn constrictor is much lower than for solid ring.
### SUMMARY:
Erco-Ribbon ™ has the same intended use as Predicate devices.
Technological differences of Erco-Ribbon™ provide superior safety and
performance.
Erco-Ribbon™ is substantially equivalent to constriction devices that have been marketed for many years.
Ercons, Inc.
PATENT PENDING
Erco-Ribbon™
CONFIDENTIAL
5
{5}------------------------------------------------
Public Health Service . . . . . . .
DEPARTMENT OF HEALTH & HUMAN USA
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 8 1997
Yakov Altshuler, Ph.D. Vice President Ercons, Inc. 66 Overlook Terrace, Suite 2E New York, New York 10040
Dear Dr. Altshuler:
Re: K972353
Erco-Ribbon™ Dated: September 18, 1997 Received: September 22, 1997 Regulatory class: unclassified Product code: 78 LKY
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
510(k) Number K972353 Erco-Ribbon "" Device Name:
### Indications for use:
Erco-Ribbon™ is used to maintain penile rigidity in men with erectile dysfunction by restricting penile venous outflow after the patient has obtained an erection (either with the aid of a vacuum pump or naturally).
Erco-Ribbon can be used for patients with following conditions:
a) separately:
premature ejaculation
firm but short lasting erection
b) as a part of a vacuum constriction system
#### Diabetes
Venous leakage
Prostatectomy
Hypertension
Psychogenic condition
Radiation therapy
Spinal cord injury
Concurrence of CDRH2 Office of Device Evaluation (ODE)
Amed H.P.
(Division Sign, Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices K972353 510(k) Number.
**Prescription Use**
(Per 21 CFR 801-109)
OR
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.