← Product Code [LGQ](/productcode/LGQ) · K972331

# INNOFLUOR GENTAMICIN ASSAY SYSTEM (K972331)

_Oxis Intl., Inc. · LGQ · Aug 1, 1997 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K972331

## Device Facts

- **Applicant:** Oxis Intl., Inc.
- **Product Code:** [LGQ](/productcode/LGQ.md)
- **Decision Date:** Aug 1, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3660
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The INNOFLUOR® Gentamicin Assay System is an in vitro diagnostic device intended for the quantitative determination of total Gentamicin in serum for therapeutic drug monitoring by fluorescence polarization immunoassay. The assay system is for use on the Abbott TDx® or the TDxFLx® analyzer. The INNOFLUOR® Gentamicin Reagent Set is intended for the quantitative determination of total gentamicin in serum for therapeutic drug monitoring. The reagent set is intended for use in the INNOFLUOR® Gentamicin Assay System. The INNOFLUOR® Gentamicin Calibrator Set is intended for use in the calibration of the INNOFLUOR® Gentamicin Assay System.

## Device Story

INNOFLUOR® Gentamicin Assay System is an in vitro diagnostic kit for therapeutic drug monitoring of gentamicin in serum. System comprises reagent set and calibrator set used on Abbott TDx® or TDxFLx® analyzers. Principle of operation is fluorescence polarization immunoassay (FPIA). User performs calibration on analyzer using calibrator set to generate curve before testing patient samples. Analyzer measures fluorescence polarization to quantify gentamicin concentration. Results assist clinicians in monitoring drug levels for therapeutic management. Used in clinical laboratory settings by trained personnel.

## Clinical Evidence

Bench testing only. Comparison study performed on 69 patient samples from individuals receiving gentamicin therapy. Subject device compared against Abbott Gentamicin Assay and existing INNOFLUOR® system using Abbott TDx® analyzer. Results demonstrated high correlation (r=0.9928 vs Abbott; r=0.9947 vs Existing) and linear regression equations confirming equivalency.

## Technological Characteristics

Fluorescence polarization immunoassay (FPIA) technology. Designed for use on Abbott TDx®/TDxFLx® analyzers. System components include reagents and calibrators for serum analysis. Quantitative measurement based on fluorescence polarization principles.

## Regulatory Identification

A phenobarbitol test system is a device intended to measure phenobarbital, an antiepileptic and sedative-hypnotic drug, in human specimens. Measurements obtained by this device are used in the diagnosis and treatment of phenobarbital use or overdose and in monitoring levels of phenobarbital to ensure appropriate therapy.

## Predicate Devices

- INNOFLUOR® Gentamicin Assay System, Existing
- Abbott Gentamicin Assay

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 972331

AUG - 1 - 1997

## 510(k) SUMMARY INNOFLUOR® GENTAMICIN REAGENT SET

Trade Name: INNOFLUOR® Gentamicin Assay System, which consists of two products that are packaged and sold separately: the INNOFLUOR® Gentamicin Reagent Set and the INNOFLUOR® Gentamicin Calibrator Set.

Gentamicin Fluorescence Polarization Immunoassay Common or Usual Name:

Classification Name: Fluorescence Polarization Immunoassay, Gentamicin

The INNOFLUOR® Gentamicin Assay System is a fluorescence polarization immunoassay intended for the quantitative determination of total gentamicin in serum for therapeutic drug monitoring. The assay system is for use on the Abbott TDx® or the TDxFLx® analyzer.

Substantial equivalence has been demonstrated between the INNOFLUOR® Gentamicin Assay System, Modified, the INNOFLUOR® Gentamicin Assay System, Existing and the Abbott Gentamicin Assay.

The technological characteristics, performance and intended use of the INNOFLUOR® Gentamicin Assay System, Modified are substantially equivalent to the INNOFLUOR® Gentamicin Assay System, Existing and the Abbott Gentamicin Assay.

Gentamicin concentrations measured by the INNOFLUOR® Gentamicin Assay System. Modified, (INNOFLUOR®, Modified), and the INNOFLUOR® Gentamicin Assay System, Existing, (INNOFLUOR®, Existing), were compared with those measured by the Abbott Gentamicin Assay (Abbott), using the Abbott TDx® analyzer, on 69 patient samples from patients receiving gentamicin therapy. Comparison of the patient sample results by linear regression analysis resulted in the regression equations: (INNOFLUOR®, Modified) = 0.985 x (Abbott) - 0.107, with a correlation coefficient of 0.9928 and (INNOFLUOR®, Modified) = 1.037 x (INNOFLUOR®, Existing) - 0.045, with a correlation coefficient of 0.9947, demonstrating equivalency of results.

Contact Person: Lvnda M. Tavlor Vice President, Technical Operations and Requlatory Affairs

Date Prepared: April 22, 1997

f

{1}------------------------------------------------

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

AUG - 1 1997 .

Lynda M. Taylor Vice President, Technical Operations OXIS International, Inc. 6040 N. Cutter Circle, Suite 317 Portland, Oregon 97217-3935

Re : K972331 INNOFLUOR™ Gentamicin Reagent Set Regulatory Class: II Product Code: LGQ Dated: June 20, 1997 Received: June 23, 1997

Dear Ms. Taylor:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

{2}------------------------------------------------

Paqe 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC). at (770) 488-7655.

This letter will allow you to begin marketing your device as ------described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,
Steven Litman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## STATEMENT OF INDICATIONS FOR USE

The INNOFLUOR® Gentamicin Reagent Set and the INNOFLUOR® Gentamicin Calibrator Set are packaged and sold separately, but are referred to collectively, in all associated product labeling, as the INNOFLUOR® Gentamicin Assay System.

The INNOFLUOR® Gentamicin Assay System is an in vitro diagnostic device intended for the quantitative determination of total Gentamicin in serum for therapeutic drug monitoring by fluorescence polarization immunoassay. The assay system is for use on the Abbott TDx® or the TDxFLx® analyzer.

The INNOFLUOR® Gentamicin Reagent Set is intended for the quantitative determination of total gentamicin in serum for therapeutic drug monitoring. The reagent set is intended for use in the INNOFLUOR® Gentamicin Assay System.

The INNOFLUOR® Gentamicin Calibrator Set is intended for use in the calibration of the INNOFLUOR® Gentamicin Assay System.

The INNOFLUOR® Gentamicin Reagent Set and Calibrator Set are used together to generate the calibration curve on the TDx®/TDxFLx® analyzer. The calibration curve must be established prior to assaying unknown samples. Prior to performing the calibration procedure, the correct analyzer operating parameters must be set by following the instructions provided in the Product Insert Supplement, which is included with every INNOFLUOR® Gentamicin Reagent Set.

(Division Sign-Off)
Division of Clinical Laboratory Devices

510(k) Numbe

TDx® and TDxFLx® are registered trademarks of Abbott Laboratories, Inc., Abbott Park, IL 60064

---

**Source:** [https://fda-staging.innolitics.com/device/K972331](https://fda-staging.innolitics.com/device/K972331)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com