OSTEO BOS SYSTEM

K972323 · Osteonics Corp. · HRS · Sep 10, 1997 · Orthopedic

Device Facts

Record IDK972323
Device NameOSTEO BOS SYSTEM
ApplicantOsteonics Corp.
Product CodeHRS · Orthopedic
Decision DateSep 10, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Osteo BOS™ System components are intended for long and small bone fracture fixation.

Device Story

Osteo BOS System consists of titanium alloy bone plates and screws used for internal fracture fixation of long and small bones. Components include various plate geometries (wide, narrow, T-plate, T-buttress, L-plate) and screw types (cortex, cancellous, washers, nuts). Device is used by surgeons in clinical/OR settings to stabilize bone fractures. System provides mechanical fixation to support bone healing. Not indicated for pedicular screw fixation or attachment to the cervical, thoracic, or lumbar vertebral column.

Clinical Evidence

bench testing only

Technological Characteristics

Materials: Titanium alloy. Components: Plates, cortex screws, cancellous screws, nuts, washers. Dimensions: Various diameters (2.7mm to 6.5mm). Principle: Mechanical internal bone fixation. Sterilization: Not specified.

Indications for Use

Indicated for internal fracture fixation of long and small bones in patients requiring surgical stabilization.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Osteo BQ5" System ____________________________________________________________________________________________________________________________________________________________ K972323 ______________________________________________________________________________________________________________________________________________________________________________ #### 510(k) PREMARKET NOTIFICATION SUMMARY OF SAFETY AND EFFECTIVENESS SEP 1 0 1997 OSTEO BOS* SYSTEM ### Submission Information | Name and Address of the Sponsor<br>of the 510(k) Submission: | Osteonics Corporation<br>59 Route 17<br>Allendale, NJ 07401-1677<br>201-825-4900 | |--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------| | Contact Person: | Kate Sutton<br>Regulatory Affairs Specialist | | Date Summary Prepared: | September 3, 1997 | | Device Identification | | | Proprietary Name: | Osteo BOS™ System | | Common Name: | The Titanium Plating System | | Classification Name and Reference: | Plate, Fixation, Bone<br>21 CFR §888.3030<br>Smooth Or Threaded Metallic Bone Fixation Fastener<br>21 CFR §888.3040 | # Predicate Device Identification The subject Osteo BOS" System components are substantially equivalent to similar bone plates and bone screws offered by Synthes. ## Device Description The Osteo BOS" System components are used for internal fracture fixation of long and small bones. All Osteo BOS" System components are manufactured from Titanium alloy and include the following: - � Plate - Wide (06.5mm/04.5mm) - Plate Narrow (06.5mm/04.5mm) . - . T-Plate (06.5mm/04.5mm) - T-Buttress Plate (06.5mm/04.5mm) . {1}------------------------------------------------ - L-Plate (Ø6.5mm/Ø4.5mm) . - Plate Wide (@4.0mm/03.5mm) � - Plate Narrow (e4.0mm/03.5mm) & Osteo Plate (04.0mm/03.5mm) � - Small T-Plate (04.0mm/03.5mm) ◆ - Plate (@2.7mm) � - Cortex Screw (02.7mm) . - . Cortex Screw (03.5mm) - t Cortex Screw (03.5mm with shaft) - Cortex Screw (04.5mm) . - Cortex Screw (04.5mm with shaft) . - Nut for @4.5mm cortex screw . - Ø8mm Washer for Ø4.0mm cancellous screws & . @3.5mm cortex screws - 06.5mm Washer for 02.7mm cortex screws � - Cancellous Screw (ø4.0mm) ↓ - Cancellous Screw (\$4.0mm with short thread) . - Cancellous Screw (06.5mm) ♥ - Cancellous Screw (06.5mm, thread lengths 16mm & 32mm) . - 013.0mm Washer for cancellous screws 26.5mm & . cortex screws 04.5mm # Intended Use The Osteo BOS" System components are intended for long and small bone fracture fixation. # Statement of Technological Comparison The subject Osteo BOS" System components are substantially equivalent in design and intended use to the predicate bone plates offered by Synthes. The subject plates are manufactured from Titanium alloy, and the predicate plates are manufactured from CP Titanium. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three distinct head profiles, suggesting a sense of community or collaboration. The bird is positioned within a circular frame, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 0 1997 Ms. Kate Sutton Regulatory Affairs Specialist Osteonics 59 Route 17 Allendale, New Jersey 07401-1677 - Re: K972323 Osteo BOS™ System Regulatory Class: II Product Codes: HWC and HRS Dated: June 19, 1997 Received: June 20, 1997 Dear Ms. Sutton: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting a device for pedicular screw fixation/attachment would cause the device to be adulterated under 501(f)(1) of the Act. This device, if intended for use in pedicular screw fixation/attachment, would be found not substantially equivalent and would be a class III device under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly: - The package insert must prominently state that the device 1. is intended for the specific use(s) described in the enclosure only; and {3}------------------------------------------------ Page 2 - Ms. Kate Sutton - You may not label or in any way promote this device for 2. pedicular screw attachment to, or fixation of the cervical, thoracic or lumbar vertebral column. If this device is a screw with outer diameters of 3 mm - 10 mm and overall lengths of 10 mm - 75 mm inclusively, the package insert must include the following statement, This device is not approved for screw "WARNING: attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine." Any pedicular screw fixation/attachment to the cervical, thoracic or lumbar vertebral column of this device is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under 21 CFR, Part 812. All users of the device for pedicular screw fixation/attachment must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conduct of the investigation. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of {4}------------------------------------------------ Page 3 - Ms. Kate Sutton Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, S. Colum M. Witten, Ph.D., M.D. a M. Witten, Ph.D., M.D. Celi Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): K972323 Device Name: Osteo BOS" System Indications For Use: The indications for use of the Osteo BOS" System bone plates and bone screws, in keeping with those of other legally marketed bone plates and bone screws, are as follows. - The Osteo BOS™ System components are intended for long and small bone fracture fixation. � (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | (Division Sig. Off) | | |---------------------|---------| | Division | | | 510(k) Number | K972323 | Prescription Use_X ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...