OPUS D-DIMER

K972316 · Behring Diagnostics, Inc. · GHH · Sep 9, 1997 · Hematology

Device Facts

Record IDK972316
Device NameOPUS D-DIMER
ApplicantBehring Diagnostics, Inc.
Product CodeGHH · Hematology
Decision DateSep 9, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7320
Device ClassClass 2

Indications for Use

OPUS D-Dimer is an in vitro fluorogenic enzyme immunoassay (ELISA) for the quantitative measurement of cross-linked fibrin degradation products containing D-Dimer in human plasma. OPUS D-Dimer is intended for use with the OPUS analyzers.

Device Story

OPUS D-Dimer is an in vitro fluorogenic enzyme immunoassay (ELISA) reagent kit used with OPUS immunoassay analyzers. It measures cross-linked fibrin degradation products (D-Dimer) in human plasma samples (EDTA, heparinized, or citrated). The device utilizes a mouse monoclonal antibody in a sandwich immunoassay format. The system is fully automated, performing fluorometric analysis using a stored calibration curve, eliminating the need for manual sample pretreatment or per-run calibration. It is intended for clinical laboratory use to aid in the diagnosis of thromboembolic events. Healthcare providers use the quantitative output to assess patient coagulation status and clinical risk.

Clinical Evidence

Bench testing only. Precision: Intra-assay %CV 7.3-8.9%; Inter-assay %CV 8.6-10.3%. Accuracy by recovery: 85-95% recovery in spiked human serum matrix. Accuracy by correlation: Compared to a commercial D-Dimer assay using 321 human plasma samples (range 37-8480 ng/ml); correlation coefficient 0.93, y-intercept 0.34, slope 0.89.

Technological Characteristics

In vitro fluorogenic enzyme immunoassay (ELISA). Reagents include mouse monoclonal antibody. Fully automated fluorometric instrument system. Supports EDTA, heparinized, and citrated plasma. Features stored calibration curve. No sample pretreatment required.

Indications for Use

Indicated for the quantitative measurement of cross-linked fibrin degradation products (XL-FDP) containing D-Dimer in human plasma, used in the diagnosis of thromboembolic events.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ \$\wedge / \subset \downarrow \sim\$ Behring Diagnostics Inc. OPUS® D-Dimer 510(k) Notification SEP 9 1997 ## 510(k) Summary for OPUS D-Dimer ## Manufactures Name, Address, Telephone, and contact person, date of 1 . preparation: Manufacturer: Behring Diagnostics Inc. 151 University Avenue Westwood, MA 02090 617-320-3117 Attn: Ruth Forstadt Preparation date: June 19, 1997 - Device Name/ Classification: 2. OPUS D-Dimer: Classification Number: Fibrinogen/fibrin degradation products assay Class II (864.7320) - 3 . marketed device: Identification of the legally ASSERACHROM® D-Di #### Description: 4 . Device OPUS D-Dimer is a set of reagents intended to be used together with the OPUS immunoassay analyzers for the quantitative measurement of cross-linked fibrin degradation products containing D-Dimer in human plasma. #### 5. Device Intended Use: OPUS D-Dimer is an in vitro fluorogenic enzyme immunoassay (ELISA) for the quantitative measurement of cross-linked fibrin degradation products containing D -Dimer in human plasma. OPUS D-Dimer is intended for use with the OPUS analyzers. **CONFIDENTIAL** {1}------------------------------------------------ Behring Diagnostics Inc. OPUS® D-Dimer 510(k) Notification ## Medical device to which equivalence is claimed and comparison 6. information: The OPUS D-Dimer assay is substantially equivalent in intended use to the ASSERACHROM® D-Di. The ASSERACHROM® D-Di, like the proposed product, employs the principle of two site or sandwich immunoassay. The OPUS D-Dimer and ASSERACHROM D-Di both quantitatively measure cross-linked fibrin degradation products that contain D-Dimer in human plasma. The OPUS D-Dimer differs from the ASSERACHROM® D-Di in that the enzyme labeled antibody is a rabbit polyclonal in the ASSERACHROM® D-Di while the enzyme labeled antibody is a mouse monoclonal in the OPUS test. Also, the OPUS D-Dimer includes a bilevel control, where as the ASSERACHROM® D-Di test does not include a control. Additionally, the OPUS D-Dimer is used with a fully automated fluorometric instrument system which includes a stored calibration curve while the ASSERACHROM® D-Di is a manual assay and uses a calibration curve with each run. Also, the OPUS D-Dimer does not require sample pretreatment for plasma samples while the ASSERACHROM® D-Di requires a predilution of the plasma samples. EDTA, heparinized and citrated plasma samples may be used with the OPUS D-Dimer, but only citrated plasma samples with the ASSERACHROM® D-Di. ### Device Performance Characteristics: 7 . ## Precision Intra-assay precision was determined by the evaluation of two levels of control material in replicates of twenty (20) each. %CV ranged from 7.3% to 8.9%. Inter-assay precision was determined by the evaluation of two levels of control material in duplicate, assayed over a five day period to total 20 replicates. %CV ranged from 8.6% to 10.3%. ## Accuracy by Recovery Recovery was determined by spiking previously assayed and pooled human serum matrix with five different levels of D-Dimer. The samples were assayed using OPUS D-Dimer in duplicate. Percent recovery ranged from 85 to 95%. #### Accuracy by Correlation OPUS D-Dimer was compared to a commercially available D-Dimer assay by evaluation of 321 human plasma samples ranging from 37 to 8480 ng/ml. A correlation coefficient of 0.93 was obtained, with a y-intercept value of 0.34 and a slope of 0.89. 0000224 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized lines. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 SEP 9 1997 Ruth Forstadt . Requlatory Affairs Associate Behring Diagnostics Inc. 151 University Avenue Westwood, Massachusetts 02090 Re : K972316 OPUS® D-Dimer Test System Regulatory Class: II Product Code: GHH Dated: June 19, 1997 Received: June 20, 1997 Dear Ms. Forstadt: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਰ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ run volu ത്രി വ ഗ്രഹ്മ Diagnostics Inc. Behring OPUS® D-Dimer 510(k) Notification Page of 510(k) Number (if known): K972316 OPUS D-Dimer Test System Device Name: J # Indications For Use: OPUS D-Dimer is an in vitro fluorogenic enzyme immunoassay (ELISA) for the quantitative measurement of cross-linked fibrin degradation products (XL-FDP) containing D-Dimer in human plasma, used in the diagnosis of thromboembolic events. OPUS D-Dimer is intended for use with the OPUS analyzers (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Clinical Laboratory Devifes 510(k) Number. K972316 Prescription Use (Per 21 CFR 801.109) OR Over-The Counter Use (Optional Format 1-2-96) 00026
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...