K972284 · Belmont Instrument Corp. · FRN · Nov 13, 1997 · General Hospital
Device Facts
Record ID
K972284
Device Name
FLUID MANAGEMENT SYSTEM
Applicant
Belmont Instrument Corp.
Product Code
FRN · General Hospital
Decision Date
Nov 13, 1997
Decision
ST
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery. Infusion of warmed fluid to rewarm patients after surgery or for hypothermia. Infusion of warmed fluid for irrigation in urology procedures.
Device Story
Belmont FMS2000 is a portable infusion pump and fluid warmer for high-speed delivery of blood, replacement IV fluids, or irrigation fluids. System inputs include fluid from bags; sensors monitor temperature, line pressure, and air in fluid path. Microprocessor-controlled pump delivers fluids at 10-500 ml/min; warming function maintains physiologic temperature. Device features touch screen for flow rate, volume, temperature, and pressure monitoring. Hardware override circuit prevents unsafe operation during computer failure. Battery backup enables mobile transport with pump and safety monitoring active (warming disabled). Used in surgical/trauma settings by clinicians. Output provides rapid fluid replacement to stabilize patients; alarms alert clinicians to unsafe conditions, facilitating immediate intervention.
Clinical Evidence
Bench testing only. Verification included pumping accuracy across flow rates, pressures, and viscosities; fluid warming capability to physiological temperatures; and safety/alarm functionality for sensor, software, and hardware failures. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity) passed per ANSI/AAMI 10993-1. Hemocompatibility testing (hemolysis and red cell fragility) confirmed the system is nonhemolytic per ASTM F756-93.
Technological Characteristics
Microprocessor-controlled infusion pump with integrated fluid warmer. Portable, 26 lb unit mountable on IV pole. Sensing: temperature, line pressure, air-in-line. Safety: hardware override circuit. Biocompatibility: tested per ANSI/AAMI 10993-1; hemocompatibility per ASTM F756-93. Connectivity: none specified. Power: AC with battery backup.
Indications for Use
Indicated for patients requiring volume replacement due to trauma or surgery, patients requiring rewarming after surgery or for hypothermia, and patients undergoing urology procedures requiring irrigation. Suitable for infusion of blood products, crystalloids, and colloids.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
K972284
Attachment III
Belmont Instrument Corporation
NOV 1 3 1997
## Belmont Instrument Corporation
780 Boston Road, Billerica, MA 01821 Tel (508) 663-0212 Fax (508) 663-0214
## PREMARKET NOTIFICATION 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS [As Required By 21 CFR 807.92(c)]
Date prepared: June 9, 1997
- Belmont Instrument Corporation 1. Submitter & 780 Boston Road Manufacturing Site: Billerica, MA 01821
Establishment Registration Number: 1219702
- Uraiwan P. Labadini, Quality Assurance/Regulatory Affairs 2. Contact Person: Manager
Fax: (508) 663-0214 (508) 663-0212 Ext. 28 Telephone:
- The Belmont Fluid Management System (7/MS2000) 3. Trade Name:
- Infusion Pump with Warmer. 4. Common name:
- Infusion Pump (per 21 CFR section 880.5725) 5. Classification name:
- 80 FRN Infusion Pump ଚ. Product Code: Device Class: Class II
7. Performance Standards: No performance standards have been officially adopted by the F.D.A.
- The Belmont Fluid Management System (7)MS2000) is substantially equivalent to 8. the Haemonetics Corporation Rapid Infusion System, R.I.S.®, which has been in commercial distribution since 1987, and which was the subject of Premarket Notification #K852645 submitted in June, 1985.
{1}------------------------------------------------
- Brief Description: The Belmont Fluid Management System (FMS2000) combines ு. advanced microprocessor technology with an efficient mechanical system to provide a high speed, simple and safe system for rapid infusion of warmed fluid. The Belmont I Jus 2000 infuses blood, replacement IV fluids or irrigation fluids warmed to physiologic temperature at user-set rates from 10 to 500 milliliters per minute (ml/min). A low infusion rate at 2.5 ml/min (150 ml/hr) is also available without heating.
The system monitors temperature, line pressure, and air in the fluid path to ensure safe operation and alarms at all unsafe conditions. A hardware override circuit prevents unsafe operation in case of system computer failure. A touch screen displays flow rate, total fluid infused, temperature, line pressure, alarm and status messages and proper procedures to proceed safely after an alarm situation.
A battery backup allows for mobile transport of the patient and system. During battery operation, fluid warming is disabled while pump operation and safety monitoring remain active.
- 10. Intended Use
The Belmont FMS2000 is for use in high blood loss surgical procedures, trauma and any situation where rapid replacement of warmed blood or replacement fluid at 10 -500 ml/min is required. It can also be used to deliver irrigation fluids at rates up to 500 ml/min.
- 11. Summary of the technological characteristics of the Belmont MS2000 compared to the Haemonetics R.I.S.®
The two systems perform the identical function, both contain a high flow pump, a fluid warmer, and a safety/surveillance system to ensure safe operation. Both can operate at high flow rates. The main differences between the two systems are as follows:
- Size and Weight: The Belmont 7MS2000 is a 26 pound compact, portable a. system which can be mounted on an IV pole. The Haemonetics R.I.S.® weighs 260 pounds, and is contained in a free standing console on casters.
- Maximum Infusion Rate: The Belmont JMS2000 unit will infuse up to 500 b. ml/min, and will empty a one unit of blood in less than a minute. The Haemonetics R.I.S.® unit will infuse up to 1500 ml/minute, and requires a large reservoir in the disposable set.
- Disposable Set. The Belmont FMS2000 and Haemonetics R.I.S.® C. disposable sets are similar in concept. The main difference is the reservoir
{2}------------------------------------------------
size. In the Belmont FMS2000 system, the reservoir serves to filter the fluid, to separate air and fluid, and to allow the air to vent via a hydrophobic filter. It serves as a conduit from the fluid bag to the system. In the Haemonetics R.I.S.® system, the reservoir acts as a filter, separates air and fluid, and also contains a large 3 liter volume for fluid storage, necessitated by the possibility of using a very high, 1500 ml/min. flow rate. In our system this volume is not needed.
- Summary of Nonclinical Tests and Results 12.
In order to verify performance of the Belmont JMS2000 in support of substantial equivalence, the following tests were carried out:
- The ability of the system to pump fluids accurately over the full range of flow a. rate and operating conditions including different input fluid temperatures, different back pressure, change in ambient temperature, and change in fluid viscosity.
- The ability of the system to warm cold fluids to physiological temperature b. over the full range of flow rate and operating conditions.
- The ability of the system to detect and alarm at unsafe or ineffective C. operating conditions including operator errors, the failure of the system sensors, the failure of the system software or computer, and other internal system malfunctions.
The Belmont 7MS2000 performed within specification in all of the above tests.
In order to verify biocompatibility, the system was tested for cytotoxicity, sensitization, irritation, and systemic toxicity, using the per ANSI/AAMI 10993-1: 1994 Biological evaluation of medical devices - Part 1: Guidance on selection of tests. All tests were passed. The system was tested for hemocompatibility by testing for red cell hemolysis, and red cell fragility. The system was found to have negligible effect on anticoagulated blood and the system is considered "nonhemolytic" according to ASTM F756-93: Standard Practice for Assessment of Hemolvtic Properties of Materials.
- 13. Conclusion: The Belmont Jus 2000 is substantially equivalent to the Haemonetics R.I.S® which received 510(k) approval in June 1985. Both systems are capable of high infusion flow rates, while maintaining infusate at physiological temperature, and both systems are suitable for use with blood products. Both systems monitor flow rate, temperature, line pressure, and the presence of air at the input line or within the system, and alarm and stop the system at all unsafe conditions.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wing segments, representing the department's mission to protect and promote the health and well-being of all Americans. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 3 1997
Mr. Uraiwan P. Labadini Quality Assurance/Regulatory Affairs Manager Belmont Instrument Corporation 780 Boston Road Billerica, Massachusetts 01821
K972284 Re : Fluid Management System Trade Name: Regulatory Class: II Product Code: FRN Dated: August 12, 1997 Received: August 15, 1997
Dear Mr. Labadini:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{4}------------------------------------------------
Page 2 - Mr. Babadini
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Ulchuski
Timothy A . Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health
্রি
. Enclosure
{5}------------------------------------------------
Attachment V
Page 1 Of 1
长972284 510(k) number: _______________________________________________________________________________________________________________________________________________________________
Device Name: Belmont Fluid Management System (JMS2000)
Indications For Use:
- Infusion of crystalloid, colloid, or blood product, including packed red blood cells, as A. volume replacement for patients suffering from blood loss due to trauma or surgery.
- Infusion of warmed fluid to rewarm patients after surgery or for hypothermia. B.
- Infusion of warmed fluid for irrigation in urology procedures. C.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patrizio Cusconte
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K972281/
Prescription Use 1 (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.