← Product Code [OYU](/productcode/OYU) · K972200

# ANEUVYSION (K972200)

_Vysis · OYU · Oct 20, 1997 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K972200

## Device Facts

- **Applicant:** Vysis
- **Product Code:** [OYU](/productcode/OYU.md)
- **Decision Date:** Oct 20, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.4700
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The AneuVysion™ (CEP 18, X, Y-alpha satellite, LSI 13 and 21) Multicolor Probe Panel is intended to use CEP 18/X/Y probe to detect alpha satellite sequences in the centromere regions of chromosomes 18, X, and Y, and LSI 13/21 probe to detect the 13q14 region and the 21q22.13 to 21q22.2 region. The AneuVysion™ kit is indicated for use as an adjunct to standard cytogenetic metaphase analysis for identifying and enumerating chromosomes 13, 18, 21, X, and Y via fluorescence in situ hybridization (FISH) in metaphase cells and interphase nuclei obtained from amniotic fluid in subjects with presumed high risk pregnancies. It is not intended to be used as a stand alone assay for test reporting. FISH results are intended to be reported and interpreted only in conjunction with results of standard cytogenetic analysis, performed concurrently, utilizing the same patient specimen. FISH results should not be reported prior to standard cytogenetic results except in instances where reporting of FISH results alone is medically indicated or standard cytogenetic results are not available, e.g., culture failure. Reporting and interpretation of FISH should be consistent with professional standards of practice! This device is intended for use only with amniocyte cells; it is not intended for and has not been validated for use with other test matrices. This FISH assay will not detect the presence of structural chromosome abnormalities frequently associated with birth defects. This FISH assay will be performed in cytogenetics laboratories.

## Device Story

AneuVysion™ kit consists of fluorescent DNA probe mixtures (CEP 18/X/Y and LSI 13/21) for fluorescence in situ hybridization (FISH). Input: amniotic fluid specimens (cultured or uncultured). Process: probes hybridize to specific chromosomal regions (centromeres 18, X, Y; loci 13q14 and 21q22.13-21q22.2); laboratory technicians perform signal enumeration via microscopy. Output: visual identification and enumeration of chromosomes 13, 18, 21, X, and Y in metaphase cells and interphase nuclei. Used in cytogenetics laboratories. Results are interpreted by clinicians alongside standard karyotyping to identify aneuploidy. Benefits: rapid adjunct to standard cytogenetics for prenatal screening of high-risk pregnancies; provides diagnostic information when standard culture fails.

## Clinical Evidence

Multi-center, blinded, controlled study of 1,516 patients (2,238 specimens). Primary endpoint: concordance between FISH and standard cytogenetic analysis. Results: 99.9% (860/861) detection of true aneuploid cases; 100% (504/504) detection of euploid cases. Analytical sensitivity: limit of detection estimated at 3% in interphase cells. Analytical specificity: 705 metaphase spreads showed no cross-hybridization. Reproducibility: no significant inter-site, inter-lot, inter-day, or inter-observer variations observed.

## Technological Characteristics

Fluorescent DNA probe mixtures (CEP 18/X/Y and LSI 13/21). CEP 18/X/Y: aqua, green, orange tri-color mixture. LSI 13/21: green, orange dual-color mixture. Sensing principle: fluorescence in situ hybridization (FISH) for target sequence detection. Form factor: liquid reagent kit. Analysis: manual signal enumeration via fluorescence microscopy.

## Regulatory Identification

An automated FISH enumeration system is a device that consists of an automated scanning microscope, image analysis system, and customized software applications for FISH assays. This device is intended for in vitro diagnostic use with FISH assays as an aid in the detection, counting and classification of cells based on recognition of cellular color, size, and shape, and in the detection and enumeration of FISH signals in interphase nuclei of formalin-fixed, paraffin-embedded human tissue specimens.

## Special Controls

The device is classified as Class II under regulation 21 CFR 866.4700 with special controls. The special control guidance document " Class II Special Controls Guidance Document: Automated Fluorescence in situ Hybridization (FISH) Enumeration Systems" is available at www.fda.gov/cdrh/oivd/guidance/1550.pdf.

*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Automated Fluorescence*in situ* Hybridization (FISH) Enumeration Systems.” See § 866.1(e) for the availability of this guidance document.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 4

#### SUMMARY: SAFETY AND EFFECTIVENESS INFORMATION FOR AneuVysion™ kit

The AneuVysion™ kit is a combination of two DNA probe mixtures; CEP 18/X/Y and LSI 13/21. The CEP 18/X/Y probe is a mixture of directly labeled fluorescent DNA probes specific for the D18Z1, DXZ1 and DYZ3 regions of chromosomes 18, X, and Y respectively. The LSI 13/21 probe contains a mixture of unique DNA sequences that hybridize in the 13q14 region of chromosome 13, and unique DNA sequences complementary to the D21S259, D21S341, and D21S342 loci contained within the 21q22.13 to 21q22.2 region on the long arm of chromosome 21. The LSI 13 probe was created from a set of overlapping clones which contain the entire RB-1 gene as well as regions extending beyond the gene on both sides. The probe extends beyond the 180 kb RB-1 gene for 110-170 kb in the 5' direction and approximately 120 kb in the 3' direction; the entire probe is 410-470 kb. CEP 18/X/Y is an aqua, green, and orange tri-color probe mixture and LSI 13/21 is a green and orange dual-color probe mixture.

The Aneu Vysion™ (CEP 18, X, Y-alpha satellite, LSI 13 and 21) Multicolor Probe Panel is intended to use CEP 18/X/Y probe to detect alpha satellite sequences in the centromere regions of chromosomes 18, X, and Y, and LSI 13/21 probe to detect the 13g14 region and the 21q22.13 to 21q22.2 region. The Aneu Vysion™ kit is indicated for use as an adjunct to standard cytogenetic metaphase analysis for identifying and enumerating chromosomes 13, 18, 21, X, and Y via fluorescence in situ hybridization (FISH) in metaphase cells and interphase nuclei obtained from amniotic fluid in subjects with presumed high risk pregnancies. It is not intended to be used as a stand alone assay for test reporting. FISH results are intended to be reported and interpreted only in conjunction with results of standard cytogenetic analysis, performed concurrently, utilizing the same patient specimen. FISH results should not be reported prior to standard cytogenetic results except in instances where reporting of FISH results alone is medically indicated or standard cytogenetic results are not available, e.g., culture failure. Reporting and interpretation of FISH should be consistent with professional standards of practice! This device is intended for use only with amniocyte cells; it is not intended for and has not been validated for use with other test matrices. This FISH assay will not detect the presence of structural chromosome abnormalities frequently associated with birth defects. This FISH assay will be performed in cytogenetics laboratories.

Standard cytogenetic analysis detects cytogenetic abnormalities by karyotyping metaphase spreads after staining the chromosomes with a dye in cultured tissue cells.

Safety and effectiveness issues relevant to FISH assays such as the AneuVysion™ assay may include cross-reactivity, poor sensitivity, poor specificity, or poor reproducibility.

<sup>1</sup> American College of Medical Genetics. Prenatal interphase fluorescence in situ hybridization (FISH) policy statement. Am J Hum Genet. 1993;53:526-527.

{1}------------------------------------------------

# Analytical Sensitivity and Specificity

## Hybridization Efficiency

In the pivotal study, among the human amniotic fluid specimens, the average percentage of cells with no hybridization signal was 0.42% for LSI 13, CEP 18, and LST 21. The average percentage of cells with only one hybridization Y signal was 0.06% for CEP X/Y. Thus, <2% cells with no or only one signal for each probe is a realistic standard of acceptance.

### Analytical Sensitivity

The analytical sensitivity of the AneuVysion™ kit probes was tested in the reproducibility study described below. In that study, the 100% XY (or 0% XO) specimen was estimated with a mean of 0.04% (40.2%), and the 10% XO specimen was estimated with a mean of 9.10% (±1.79%) X-signaled nuclei. The upper 95% CI was 0.43 for the 0% X0 specimen and the lower 95% CI for the 10% XO specimen was 5.59%. Thus, the limit of detection for the AneuVysion™ kit in interphase cells is estimated to be 3%.

## Analytical Specificity

Locus specificity studies were performed with metaphase spreads according to standard Vysis QC protocols. A total of 705 metaphase spreads were examined sequentially by G-banding to identify chromosomes 13, 18, 21, X, and Y, followed by FISH. No cross-hybridization to other chromosome loci was observed in any of the 705 cells examined; hybridization was limited to the target regions of chromosomes 13, 18, 21, X, and Y.

## Reproducibility

ん

A pivotal study was conducted to assess the reproducibility of the AneuVysion™ assay interphase analysis for the percentage of aneuploid cells. The Aneu Vysion™ assay were assessed for inter-site, inter-lot, inter-day and inter-observer reproducibility. One normal and three mosaic cultured human amniocyte specimens were evaluated for the percentage of aneuploid cells according to the instructions for signal enumeration in the package insert. Using ANOVA, no significant variations were observed in any of the inter-assay reproducibility parameters. The intra-assay mean, S.D., and percent C.V. of the observed percentage of aneuploid nuclei for all samples are shown in Table 1.

| Precision of % Aneuploid Cells by Level of Mosaicism |    |                          |                       |                      |                       |                      |                      |                      |                     |                     |
|------------------------------------------------------|----|--------------------------|-----------------------|----------------------|-----------------------|----------------------|----------------------|----------------------|---------------------|---------------------|
| Specimen                                             | N  | Summary<br>Statistics    | %X                    | %XX                  | %XXX                  | %XY                  | + 21                 | %2-sig<br>CH-21      | %2-sig<br>CH-18     | %2-sig<br>CH-13     |
| 100% XY                                              | 24 | mean<br>S.D.<br>C.V. (%) | 0.04<br>0.20<br>--    | 0.04<br>0.20<br>--   | 0.00<br>0.00<br>--    | 98.88<br>1.14<br>1.2 | 0.79<br>0.77<br>--   | 96.7<br>1.01<br>1.0  | 94.6<br>1.98<br>2.1 | 97.1<br>1.53<br>1.6 |
| 10% X/<br>90% XX                                     | 24 | mean<br>S.D.<br>C.V. (%) | 9.10<br>1.51<br>16.6  | 88.17<br>1.61<br>1.8 | 0.85<br>0.67<br>--    | 0.00<br>0.00<br>--   | 1.25<br>0.69<br>--   | 94.8<br>1.44<br>1.5  | 94.9<br>1.62<br>1.7 | 95.6<br>1.69<br>1.8 |
| 17% X/<br>47% XX/<br>36% XXX                         | 24 | mean<br>S.D.<br>C.V. (%) | 19.48<br>4.06<br>20.8 | 42.56<br>3.65<br>8.6 | 36.68<br>3.74<br>10.2 | 0.00<br>0.00<br>--   | 1.16<br>0.72<br>--   | 96.1<br>1.72<br>1.8  | 95.7<br>1.16<br>1.2 | 97.3<br>1.03<br>1.1 |
| 50% XY+21/<br>50% XY                                 | 24 | mean<br>S.D.<br>C.V. (%) | 0.04<br>0.12<br>--    | 0.00<br>0.00<br>--   | 0.00<br>0.00<br>--    | 98.13<br>1.34<br>1.4 | 52.03<br>3.25<br>6.2 | 45.98<br>3.69<br>8.0 | 96.1<br>0.90<br>0.9 | 97.1<br>1.33<br>1.4 |

Table 1 Precision of % Aneuploid Cells by Lavel of Mossicism

Property of Vysis Inc.

{2}------------------------------------------------

# Methods Comparison; Clinical Specimens

A multi-center, blinded, controlled, comparative study was conducted to further define A multi-center, onnoon, contronou, ochipe in relative to standard cytogenetic analysis, the standard of care, in cultured and uncultured amniotic fluid speciments. Thirty one investigation sites analyzed amniotic fluid specimens obtained from a total of 1516 patients.

All study sites conducted the trial according to the prescribed assay procedures and signal enumeration guides.

A total of 2,238 amniocyte specimens were obtained and analyzed from 1,516 patients. Of these 2,238 specimens, 55 were deemed uninformative. These 55 uninformative specimens included three due to maternal cell contamination, four due to FISH assay specificals included unce date to machine (c40) number of nucles for analysis. Thus on per specimen basis, the rate of informativeness is 97.5% (2183/2238). Note that among the Specifich basis, the rate of insufficient (<40) number of nuclei available for one of both of the probes, 31 were partially uninformative for either the CEP 18/X/Y (22) or the LSI 13/21 (9).

Of the 1516 patients, thirteen patients with either cultured or uncultured speciments were included in the 55 uninformative specimens. Thus on per patient basis, the rate of informativeness is 99.8% (1503/1516). One species per patient was included in the miorinativeness is 97.0% (1958) 1939) were cultured and 927 were uncultured specimens.

The maternal age ranged from 13 to 52 years, with a mean (±S.D.) age of 33.2 years The matchial age ranged from 11 to 38 weeks, with a mean (tS.D.) of (10.6 years). The gestarional age varied age varied among study sites, while the mean gestational age did not.

Each site performed FISH analyses according to the instructions in the Ane Vysion™ kit package insert. The percentage of aneuploid cells was determined by FISH after enumerating a minimum of 50 interphase nuclei per target; a minimum of 40 evaluable nuclei was deemed informative.

From this pivotal multi-center comparative study described above, the results of interphase FISH analysis were compared to standard cytogenetics.

## True Aneuploid Cases

Among the 861 aneuploid cases, there were 75 +13; 192 +18; 322 +21; 107 45,X;
Among the 861 aneuploid cases, there were 75 +13; 192 +18; 322 +21; 107 45,X; 44 47,XXX; 57 47,XXY, 24 47,XYY; one -21; one XXY +18; one XXX +18; 44 47,XXXXY; one 48,XXYY; one tetraploid, 92,XXYY; and 33 triploids (19 69,XXXX and 14 69,XXY) and one 46,XXX idic(18). Of which, 860 had % 09,AXA and 14 07,AXX / and one 35%, which was due to long storage of prepared slide. Thus, under the worst case scenario, for determination of true aneuploidy, FISH is able to detect 99.9% (860/861) of cases identified by or true and cytogenetic analysis. Note also that one male and one female trisomy 21 Standard Cycligencie analysis. - Il contamination were deemed informative.

{3}------------------------------------------------

## Mosaic Cases

There were 62 true mosaic cases, as identified by standard cytogenetic analysis. There were 23 cases associated with X0, 10 with XXX, 8 with XXY, 7 with +21, 4 with +18, 2 with -18, 3 with +13, and 5 with X0 complexes. Even though aneuploid cell lines were detected in 60 cases, twelve of the 62 mosaic cases showed less than 10% % aneuploid cells by FISH, and 15, showed greater than 60%. The correlation of the % aneuploid cells is 0.76, between FISH assay and standard cytogenetic analysis.

#### Pseudomosaic Cases

There were 76 pseudomosaic cases, as identified by standard cytogenetic analysis. Among these 76 cases, 7 +13, 9 -13, one +18, 24 -18, 11 +21, 19 -21, 25 X0, one XXX, one XXY, one XYY, and 6 XXYY were observed. FISH assay results showed less than 10% aneuploid cells which is consistent with the euploid state.

#### Euploid Cases

Among 504 euploid cases identified by standard cytogenetic analysis, FISH also found each to have % aneuploid cells less than 10%. Thus, for determination of euploidy, FISH is able to detect 100% (504/504) of cases identified by standard cytogenetic analysis. Note also that all four male fetal cases with mild maternal cell contamination were deemed informative.

## Pairwise Comparison between Cultured and Uncultured Specimens There were 722 patients in the pivotal study with FISH assay performed on both uncultured and cultured samples of the same specimen. In mosaic cases, the aneuploid cell lines were detected by FISH with varying %aneuploid cells between cultured and uncultured samples, leading to a few discordance. The FISH test results in aneuploid, euploid and pseudomosaic cases were concordant between cultured and uncultued samples.

## Conclusions

Performance of Aneu Vysion™ is supported by the Vysis Quality Control Procedures and is demonstrated in the clinical studies. When the AneuVysion™ DNA Probe Kit is used as instructed in the package insert, the above statements describe its performance.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around it. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of curved lines.

JAN 2 5 2012

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

Vysis c/o Dr. Russel K. Enns Vice President, Regulatory Affairs 3100 Woodcreek Dr. Downers Grove, IL 60515

Re: k972200

Trade/Device Name: AneuVysion (CEP 18, X, Y-alpha satellite, LSI 13 and 21) Multicolor Probe Panel Regulation Number: 21 CFR §866.4700 Regulation Name: Automated fluorescence in situ hybridization (FISH) enumeration systems. Regulatory Class: Class II Product Code: OYU, KIR Dated: September 18, 1997 Received: September 19, 1997

Dear Dr. Enns:

This letter corrects our substantially equivalent letter of October 20, 1997.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);

{5}------------------------------------------------

Page 2 – Dr. Russel K. Enns

labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), 11 Jou active of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 Fires, part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of

Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Maria M Clan

Maria M. Chan, Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## 510(k) Number (if known): K972200

#### Device Name: Aneu Vysion™ Multicolor DNA Probe Kit CEP® 18/X/Y-alpha satellite, LSI® 13/21

### Indications For Use:

The AneuVysion™ (CEP 18, X, Y-alpha satellite, LSI 13 and 21) Multicolor Probe Panel is intended to use CEP 18/X/X probe to detect alpha satellite sequences in the centromere regions of chromosomes 18, X, and Y, and LSI 13/21 probe to detect the 13q14 region and the 21q22.13 to 21q22.2 region. The AneuVysion™ kit is indicated for use as an adjunct to standard cytogenetic metaphase analysis for identifying and enumerating chromosomes 13, 18, 21, X, and Y via fluorescence in situ hybridization (FISH) in metaphase cells and interphase nuclei obtained from uncultured amniotic fluid in subjects with presumed high risk pregnancies. It is not intended to be used as a stand alone grossy for test reporting. FISH results are intended to be reported and interpreted only in conjunction with results of standard cytogenetic analysis, performed concurrently utilizing the same patient specimen. FISH results should not be reported prior to standard cytogenetic results except in instances where reporting of FISH results alone is medically indicated or standard cycogenetic results are not available, e.g., culture failure. Reporting and interpretation of FISH should be consistent with professional standards of propotice [1] ] This device is intended for use only with amniocyte cells; it is not intended for and has not been validated for use with other test matrices. This FISH assay will not dee the presence of structural chromosome abnormalities frequently associated with birth defects. This FISH assay will be performed in cytogenetics laboratories.

Tita E. Masini

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number 1972200

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR

Over-The-Counter Use_

---

**Source:** [https://fda-staging.innolitics.com/device/K972200](https://fda-staging.innolitics.com/device/K972200)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
