HOME CARE BEAD BED

K972176 · Kinetic Concepts, Inc. · INX · Jan 16, 1998 · Physical Medicine

Device Facts

Record IDK972176
Device NameHOME CARE BEAD BED
ApplicantKinetic Concepts, Inc.
Product CodeINX · Physical Medicine
Decision DateJan 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5160
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Home Care Bead Bed is designed to provide Air Fluidized Therapy that helps prevent and treat complications of immobility such as skin breakdown and decubitus ulcers, while providing minimized dehydration. The Head Cushion System is an air inflatable cushion system designed to raise a patient's upper body. This cushion system is a replacement to the foam cushion wedges presently available on the FluidAir Bead Bed.

Device Story

Home Care Bead Bed provides air-fluidized therapy for home use. Device consists of tank filled with microspheres; air supply unit forces air upward through beads, creating fluid-like support surface to reduce interface pressure and maintain dry environment. System includes head cushion system (two air bladders) and knee gatch bladder for patient positioning. Caregiver operates device via power-control unit (fluidization rate, temperature) and hand control (cushion firmness, elevation). Top head cushion operates via pressure feedback; bottom cushion and knee gatch operate on binary ON/OFF logic. Device facilitates patient positioning up to 40 degrees. Designed for portability and home installation. Output is physical support surface; clinical benefit is prevention/treatment of skin breakdown and decubitus ulcers in immobile patients.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Air-fluidized therapy bed using microspheres. Components: head/foot pedestals, tank, soft diffuser, air supply (blowers, heat exchanger, thermostat). Head cushion system: vinyl Gore-Tex® bladders. Control: manual knobs for fluidization/temperature; hand control for cushion firmness/elevation. Connectivity: standalone. Power: electrical. Sterilization: not specified.

Indications for Use

Indicated for patients requiring advanced pressure relief and a dry environment, including those with pressure ulcers (especially draining wounds), skin flaps/grafts, burns, acute exfoliative dermatitis, and oncology patients (for pain relief and reduction of pathological fracture risk). Contraindicated for patients with acute/unstable spinal cord injury or those requiring traction.

Regulatory Classification

Identification

An air-fluidized bed is a device employing the circulation of filtered air through ceramic spherules (small, round ceramic objects) that is intended for medical purposes to treat or prevent bedsores, to treat severe or extensive burns, or to aid circulation.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for KCI. The logo consists of a series of curved lines on the left side, followed by the letters "KCI" in bold, sans-serif font. The curved lines are arranged in a semi-circular shape, with each line slightly offset from the others. A horizontal line is present below the logo. | 510 (k) Summary | JAN 1 6 1998 | |-----------------|--------------| |-----------------|--------------| | I. Name of Device: | Home Care Bead Bed | |-------------------------------|--------------------------------------------------------------| | II. Classification Name: | Bed, Air Fluidized Therapy, Powered<br>(per 21 CFR 890.5160) | | III. Substantial Equivalence: | FluidAir®, 510(k) No. K881917 | Device Description: IV. The Home Care Bead Bed is designed to provide Air Fluidized Therapy by supporting the patient on a surface of fluidized Microspheres (or "beads"). A large volume of air is forced upward through the microsphere mass causing the beads to behave as a fluid, helping reduce interface pressures while providing a dry patient environment with minimized dehydration. The Home Bead Bed is designed for home health care use allowing easy two-man transportation, installation and serviceability. The Home Bead Bed is a modification of the predicate fluidized bead bed, FluidAir® (by KCI), currently on the market and complies with the weight regulations for mobile and wood floor homes. The Home Care Bead Bed is comprised of the Head Pedestal (containing the Head Cushion System Control unit), the Foot Pedestal (containing the Power-Controller unit), the Tank unit with the Soft Diffuser, a wooden Floor Board, the Beads, Filter Sheet and the Head Cushion System incorporated into the Cover Sheet. The Foot Pedestal encloses the Air Supply unit that provides air to the soft diffuser and the Head Cushion System. The Air Supply unit contains similar components used in the FluidAir power unit such as blowers, heat exchanger, thermostat controls, cooling fans and filters. The Power-Control Unit located at the Foot Pedestal (refer to membrane diagram) has an "On/OFF" button and a "Fluidization" knob that allows the care giver to increase or decrease the fluidization. By turning the knob toward "MAX", the blower speed increases and sends more air through the Soft Diffuser thus increasing the fluidization rate. By turning the knob toward "MIN", the blower speed decreases and less air is blown through the diffuser thus decreasing the fluidization rate. The "Temp" knob on the Power-Control Unit allows the care giver to adjust the temperature of the air flowing to the beads. This heating and cooling regulation system is similar to the one used on the FluidAir. The Head Cushion System consists of a head cushion (consisting of two air bladders) and a knee gatch bladder made of vinyl Gore-Tex® fabric and sewn onto a Gore-Tex® fabric cover sheet, similar to the FluidAir bed cover sheet, and placed atop the Home Care Bead bed. The two head cushion bladders and the knee gatch bladder are controlled by small bore servo valves located in the Head Pedestal. These valves are connected to individual port pins for appropriate control. The top head cushion bladder is operated in a pressure feedback mode, while the bottom head cushion operates only in an ON or OFF mode (no pressure feedback control). The knee gatch valve is plumbed in such a way as to draw air from the bottom head cushion bladder whenever Corporate: 8023 Vantage Drive San Antonio, Texas 78230-4726 (210) 524-9000 {1}------------------------------------------------ the "High" setting is selected. Otherwise, the knee gatch bladder remains deflated. Forty (40) degrees of elevation can be achieved when both head cushions are inflated. The knee gatch bladder provides greater patient positioning when the bottom head cushion is elevated. The Hand Control has an ON and OFF control for fluidization, and SOFT and FIRM and HIGH and LOW controls for the Head Cushion System. When the Head Cushion System's power is turned ON, the system defaults to the LOW setting of mid-range firmness for the Top Head cushion. The Bottom Head cushion and Knee Gatch cushion remain deflated. The FIRM and SOFT control buttons adjusts the firmness level of the Top cushion. The HIGH control button fully inflates the Bottom Head cushion and the Knee Gatch bladder. The head will elevate to a maximum of 40 degrees, depending on the firmness of the Top Head Cushion. The LOW control button returns the Low setting, which will return the head elevation to approximately 10 degrees. When the Head Cushion System is turned OFF, all cushions in the Head Cushion System are deflated. Date Prepared: June 1, 1997 {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 6 1998 Mr. William H. Quirk · Director of Regulatory Affairs Kinetic Concepts, Inc. 3440 East Houston Street San Antonio, Texas 78265-9508 Re: K972176 Home Care Bead Bed Trade Name: Requlatory Class: II INX Product Code: Dated: September 23, 1997 October 20, 1997 Received: Dear Mr. Quirk: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ ## Page 2 - Mr. William H. Quirk This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Summary Page 3 ## INTENDED USE OF HOME CARE BEAD BED The Home Care Bead Bed is designed to provide Air Fluidized Therapy that helps prevent and treat complications of immobility such as skin breakdown and decubitus ulcers, while providing minimized dehydration. The Head Cushion System is an air inflatable cushion system designed to raise a patient's upper body. This cushion system is a replacement to the foam cushion wedges presently available on the FluidAir Bead Bed. ## INDICATIONS The Home Care Bead Bed is indicated for patients who would benefit from advanced pressure relief and a dry patient environment. Particular classes of such patients include (among others), victims of the following: - Pressure Ulcers especially draining wounds . - . Skin Flaps and Grafts - · Burns - · Acute Exfoliative Dermatitis - · Oncology: - · Aids in pain relief - · Reduces risk of Pathological Fractures ## CONTRAINDICATIONS Patient conditions for which the application of Air Fluidized Therapy on the Home Care Bead Bed is contraindicated include: - · Acute or Unstable Spinal Cord Injury - · Traction Prescription Use (Per 21 CFR 801.109) bcolly (Division Sign-Off) Division of General Restorative Devices K972176 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...