K972175 · Localmed, Inc. · QEY · Aug 28, 1997 · Cardiovascular
Device Facts
Record ID
K972175
Device Name
INFUSASLEEVE IIA (IS-2A)
Applicant
Localmed, Inc.
Product Code
QEY · Cardiovascular
Decision Date
Aug 28, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InfusaSleeve IIa is intended to infuse fluids such as heparinized saline, diagnostic agents such as contrast media, and thrombolytic agents such as urokinase into the coronary vasculature. The catheter is designed for use with a commercially available dilatation catheter to facilitate selective and subselective access and regional infusion.
Device Story
Multi-lumen infusion catheter; tracks coaxially over standard dilatation catheters. Input: infusion fluids (heparinized saline, contrast media, thrombolytic agents) via proximal luer port. Operation: fluid travels through four peripheral infusion lumens, exiting through multiple sideholes in distal infusion region into coronary vasculature. Used in clinical settings by physicians; facilitates regional drug delivery. Output: localized infusion of therapeutic/diagnostic agents. Benefits: enables targeted coronary delivery of agents during interventional procedures.
Clinical Evidence
No human clinical data provided. Evidence consists of non-clinical bench testing including material biocompatibility, bond joint tensile strengths, infusion pathway fatigue, burst pressure, infusate flow, torque strength, and in-vivo animal evaluations.
Indicated for patients requiring infusion of fluids, diagnostic agents, or thrombolytic agents into the coronary vasculature via selective or subselective access using a dilatation catheter.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
InfusaSleeve II
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The seal features an abstract design with lines and curves.
February 3, 2022
LocalMed, Inc. Ed Lee Director, Regulatory Affairs/quality Assurance 1820 Embarcadero Rd. Palo Alto, California 94303
Re: K972175
Trade/Device Name: InfusaSleeve IIa (IS-2a) Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: QEY, KRA
Dear Ed Lee:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 28, 1997. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075, Gregory.Oconnell(@FDA.HHS.gov.
Sincerely,
Image /page/0/Picture/9 description: The image shows a digital signature. The signature is from Gregory W. O'connell. The date of the signature is 2022.02.03, and the time is 14:35:39 -05'00'.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle's head is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856
Mr. Ed Lee Director, Requlatory Affairs and Quality Assurance LocalMed 1820 Embarcadero Road - -Palo Alto, California 94303
AUG 28 1997
Re: K972175 Kaplan-Simpson InfusaSleeve™ IIa Requlatory Class: II (two) Product Code: KRA Dated: June 6, 1997 Received: June 9, 1997
Dear Mr. Lee:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for
annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic (21 Crk Farc our and dadreronary %07.10 10-12-200 at 13:00 am
devices), please contact the Office of Cómpliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# 510(K) Number (if known): K972175
Device Name: InfusaSleeve IIa (IS-2a)
Indications for Use:…… -------------------------------------------------------------------------------------------------------------------------------------------------------
The InfusaSleeve IIa is intended to infuse fluids such as heparinized saline, diagnostic agents such as contrast media, and thrombolytic agents such as urokinase into the coronary vasculature. The catheter is designed for use with a commercially available dilatation catheter to facilitate selective and subselective access and regional infusion.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (Division Sign-Off) | |
| Division of Cardiovascular, Respiratory, and Neurological Devices | |
| 510(k) Number | ____________________________________ |
| Prescription Use (Per 21 CFR 801.109) | <span> <svg class="bi bi-check-lg" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M13.854 2.146a.5.5 0 0 1 0 .708l-7 7a.5.5 0 0 1-.708 0l-3.5-3.5a.5.5 0 1 1 .708-.708L6.5 9.793l6.646-6.647a.5.5 0 0 1 .708 0z" fill-rule="evenodd"></path> </svg> </span> |
| | OR |
| Over-The-Counter Use | ____________________________________ |
| | (Optional Format 1-2-96) |
{4}------------------------------------------------
K972175
510(K) SUMMARY
AUG 2 8
| SUBMITTER'S NAME AND ADDRESS: | LocalMed, Inc.<br>1820 Embarcadero Road<br>Palo Alto, CA 94303 |
|-------------------------------------|----------------------------------------------------------------|
| PHONE: | (415) 843-6770<br>(415) 843-6771 fax |
| CONTACT PERSON: | Edwin Lee |
| DATE SUMMARY PREPARED: | June 6, 1997 |
| DEVICE TRADE NAME: | Kaplan-Simpson InfusaSleeve <sup>TM</sup> IIa |
| DEVICE COMMON NAME: | Catheter, infusion |
| DEVICE CLASSIFICATION NAME: | Catheter, continuous flush |
| SUBSTANTIALLY EQUIVALENT DEVICE(S): | InfusaSleeve <sup>TM</sup> II |
### DEVICE DESCRIPTION AND FUNCTION:
The Kaplan-Simpson InfusaSleeve™ IIa is a multi-lumen catheter consisting of a proximal infusion port, a proximal shaft, and a main catheter shaft with a distal infusion region. An entry port for balloon and guide wire access is located at the proximal end of the main shaft which is proximal to the infusion region. See Figure 1.
Image /page/4/Figure/6 description: This image shows a medical device with labels pointing to its different parts. The device includes a proximal infusion port, proximal shaft, main catheter shaft, shaft transition marker, balloon/wire entry port, radiopaque markers, tip, and flushing tool. There is also a detailed view of the infusion region, highlighting expansion slits and sideholes.
### Figure 1
The infusion region consists of four separate infusion lumens each with multiple sideholes. Radiopaque markers are located within the infusion region. Infusion solution is delivered through the proximal infusion port, through the four infusion lumens and exits through the sideholes into the coronary vasculature. A standard luer adapter allows for attachment of a syringe with a manometer or equival infusion port. A flushing tool is included to facilitate flushing the main catheter shaft through the distal tip. The InfusaSleeve IIa is designed to track over standard dilatation catheters.
#### INTENDED USE:
The InfusaSleeve IIa is intended to infuse fluids such as heparinized saline, diagnostic agents such as contrast media, and thrombolytic agents such as urokinase into the coronary vasculature. The catheter is designed for use
{5}------------------------------------------------
with a commercially available dilatation catheter to facilitate selective access and regional infusion.
-
| COMPARISON OF TECHNOLOGICAL CHARACTERISTICS: | | |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| | InfusaSleeve II | InfusaSleeve IIa (premarket notification) |
| General indications: | General intravascular use, delivery of<br>therapeutic solutions. Selective and sub-<br>selective controlled infusion of solutions. | General intravascular use, delivery of<br>therapeutic solutions. Selective and sub-<br>selective controlled infusion of solutions. |
| Usage: | Single use, disposable | Single use, disposable |
| How supplied: | Sterile, non-pyrogenic | Sterile, non-pyrogenic |
| Sterilization method: | 100% EtO sterilized | 100% EtO sterilized |
| Materials: | | |
| Infusion port: | Polycarbonate | Polycarbonate |
| Proximal shaft: | Stainless Steel, Nylon | Stainless Steel, Nylon |
| Main shaft & Tip: | Nylon | Nylon |
| Lubricious<br>coating: | Silicone | Silicone |
| Radiopaque<br>marker: | Gold/Stainless steel | Gold/Stainless steel |
| Support elements: | Stainless steel | Stainless steel |
| Design &<br>Construction: | | |
| Catheter<br>configuration: | Introduced coaxially over a dilatation<br>device. Infusion through peripheral<br>lumens. | Introduced coaxially over a dilatation<br>device. Infusion through peripheral<br>lumens. |
| Shaft<br>configuration: | Multi-lumen device. Central lumen<br>accommodates dilatation device. Four<br>opposing peripheral infusion lumens. | Multi-lumen device. Central lumen<br>accommodates dilatation device. Four<br>opposing peripheral infusion lumens. |
| Distal infusion<br>region: | Nine delivery sideholes per peripheral<br>lumen. 0.04 mm sidehole diameters. | Nine delivery sideholes per peripheral<br>lumen. 0.028 mm sidehole diameters. |
| Mode of operation: | | |
| Solution delivery<br>mode: | Solution delivery via the peripheral lumen<br>sideholes into the vasculature. | Solution delivery via the peripheral lumen<br>sideholes into the vasculature. |
| Maximum<br>proximal infusion<br>pressures: | 50 psig | 50 psig |
| Vascular access: | Percutaneous | Percutaneous |
| Tracking<br>mechanism: | Over-the-balloon | Over-the-balloon |
| Size: | | |
| Total length: | 1.4 m | 1.4 m |
| Working length: | 1.3 m | 1.3 m |
| | InfusaSleeve II | InfusaSleeve IIa (premarket notification) |
| Infusion region<br>length: | 2 cm | 2 cm |
| Outer diameter: | 1.5 - 2.1 mm | 1.5 - 2.1 mm |
| Inner diameter: | 0.9 - 1.5 mm | 0.9 - 1.5 mm |
| Compatibility: | | |
| Max balloon<br>diameter | 3.5 mm | 4.0 mm |
| Max balloon<br>catheter shaft size: | 3.5 Fr | 3.8 Fr |
| Guiding catheter: | Standard | Standard |
| Infusion port<br>fitting: | ISO 594 luer conical fitting | ISO 594 luer conical fitting |
| Radiopacity: | via Radiopaque distal markers | via Radiopaque distal markers |
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
1
{6}------------------------------------------------
### NONCLINICAL TESTS:
Material biocompatibility, bond joint tensile strengths, infusion pathway fatigue, burst pressure, infusate flow, torque strength, histology, and in-vivo animal evaluations.
## TEST CONCLUSIONS:
The test results verified that the InfusaSleeve IIa performance is equivalent to, or exceeds the InfusaSleeve II device performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.