LYPHOCHEK REMEDI CONTROL

K972174 · Bio-Rad · DIF · Jun 24, 1997 · Clinical Toxicology

Device Facts

Record IDK972174
Device NameLYPHOCHEK REMEDI CONTROL
ApplicantBio-Rad
Product CodeDIF · Clinical Toxicology
Decision DateJun 24, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3280
Device ClassClass 1

Indications for Use

Lyphochek REMEDi Control is intended for use as a quality control urine for the REMEDi HS Drug Profiling System.

Device Story

Lyphochek REMEDi Control is a quality control product consisting of human urine spiked with therapeutic drugs and drugs of abuse. It is provided in a lyophilized (freeze-dried) form to enhance stability. The device is used in clinical laboratory settings to monitor the performance and accuracy of the REMEDi HS Drug Profiling System. By providing a known sample with established drug concentrations, the control allows laboratory personnel to verify that the profiling system is functioning correctly, ensuring reliable results for patient drug testing. The control is reconstituted and processed through the system similarly to patient samples.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Matrix: Human urine. Form: Lyophilized. Storage: 2-8°C. Stability: 30 days open-vial at 2-8°C. Contains added therapeutic drugs and drugs of abuse.

Indications for Use

Indicated for use as a quality control urine for the REMEDi HS Drug Profiling System to monitor the performance of drug testing in human urine.

Regulatory Classification

Identification

A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K972174 Image /page/0/Picture/1 description: The image shows a partial view of the Bio-Rad logo. The logo is in black and white, with the letters "BIO-RAD" visible. The letters are bold and sans-serif. The logo is likely part of a larger document or advertisement. JUN 24 1997 **Bio-Rad** **Laboratories** ECS Division 26 E. Miraloma Avenue naheim. CA 92806 obona (714) 630-6400 all Free (800) 854-673. ## 510(k) Summary Submitter Bio-Rad Laboratories, ECS Division 3726 E. Miraloma Avenue Anaheim, CA 92806 (714)630-6400 Fax (714)666-1383 Contact Person Elizabeth Platt Date of Summary Preparation June 5, 1997 Device (Trade & Common Name) Lyphochek REMEDi Control Classification Name Class I, CFR 862.3280: Drug Mixture Control 91DIF Devices to Which Substantial Equivalence is Claimed Lyphochek Urine Toxicology Control Bio-Rad Laboratories, ECS Division Anaheim, CA K881989 Statement of Intended Use Lyphochek REMEDi Control is intended for use as a quality control urine for the REMEDi HS Drug Profiling System. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in a rounded rectangle shape. The words "BIO-RAD" are written in white letters on a black background. **Bio-Rad** **Laboratories** ECS Division 3726 E. Miraloma Avenue Anaheim. CA 92806 Telephone (714) 630-6400 Toll Free (800) 854-6737 ## Description of the Device Lyphochek REMEDi Control is prepared from human urine with added therapeutic drugs and drugs of abuse. The control is provided in lyophilized form for increased stability. ## Statement of How Technological Characteristics Compare to Substantial Equivalence Device A table is provided below comparing the similarities between the Bio-Rad Lyphochek REMEDi Control and the devices to which substantial equivalence is claimed. | | Bio-Rad Laboratories<br>Lyphochek REMEDI Control | Bio-Rad Laboratories Lyphochek<br>Urine Toxicology Control | |-----------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended<br>Use | Lyphochek REMEDI Control is<br>intended for use as a quality<br>control urine for the REMEDi HS<br>Drug Profiling System. | Lyphochek Urine Toxicology<br>Control is intended for use as a<br>quality control urine to monitor the<br>performance of an individual<br>laboratory's broad-scren testing of<br>drugs of abuse in urine. | | Levels | Two | One | | Form | Lyophilized | Lyophilized | | Open<br>Vial | 30 Days at 2-8°C | 10 Days at 2-8°C | | Matrix | Human Urine | Human Urine | | Storage | 2-8°C | 2-8°C | {2}------------------------------------------------ Image /page/2/Picture/10 description: The image shows the logo for the Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUN 2 4 1997 Elizabeth Platt Staff Requlatory Affairs Representative Bio-Rad Laboratories 3726 E. Miraloma Avenue Anaheim, California 92806 K972174 Re : Lyphochek REMEDi Control Requlatory Class: II Product Code: DIF June 5, 1997 Dated: Received: June 9, 1997 Dear Ms. Platt: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 510(k) Number: ---------------------------------------------------------------------------------------------------------------------------------------------------------------Device Name: Lyphochek REMEDi Control Indications for Use: Lyphochek REMEDi Control is intended for use as a quality control urine for the REMEDi HS Drug Profiling System. (PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) (Concurrence of CDRH, Office of Device Evaluation) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of Clinical Laboratory Devices | | | 510(k) Number | K972174 | | Prescription Use | | |------------------|--| |------------------|--| OR Over-The Counter Use 28
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...