IMATION SE-196 LASER IMAGER

K972163 · Imation Corp. · LMC · Sep 5, 1997 · Radiology

Device Facts

Record IDK972163
Device NameIMATION SE-196 LASER IMAGER
ApplicantImation Corp.
Product CodeLMC · Radiology
Decision DateSep 5, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2040
Device ClassClass 2

Indications for Use

The Imation™ SE-196 laser imager is intended use as a high quality hard copy device for output from digital imaging source modalities for use in medical imaging diagnosis and referral. Electronic image information signals are managed in the SE-196 and transformed optically to expose Imation imaging media. The system is intended for use with a variety of digital modalities including CT, MR and CR for diagnostic use by medical radiologists and communications to referring physicians and their patients.

Device Story

Imation SE-196 Laser Imager; receives digital image data from modalities (CT, MR, CR); manages signals via internal Image Management System (IMS); transforms signals optically to expose imaging media; prints hard copy film for diagnostic review. Used in clinical settings by radiologists/physicians. Features Automated Image Quality Control (AIQC) to match printing power with film characteristics; ensures density uniformity. Provides high-quality hard copy output; facilitates diagnostic interpretation and communication between clinicians and patients. No patient contact; device does not monitor or control patient-connected equipment.

Clinical Evidence

Bench testing only. Performance validated via spatial frequency response (325 dpi), gray-scale resolution, and density uniformity tests. AIQC system validated for maintaining density consistency over time and media changes. Reliability, qualification, and internal testing completed.

Technological Characteristics

Laser printer; 325 dpi resolution; 12-bit gray-scale. Uses daylight cartridges for imaging media. Integrated processor. Connectivity via modality host control or keypad. AIQC algorithm for density control. Safety standards: UL544, IEC601-1, IEC 825, UL1950.

Indications for Use

Indicated for medical radiologists and referring physicians to produce hard copy diagnostic images from digital modalities (CT, MR, CR) for diagnostic use and patient communication.

Regulatory Classification

Identification

A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary K972163 SEP - 5 1997 June 1, 1997 Imation Corp. 1 Imation Place Oakdale MN 55128 Contact: Stephen G. Slavens 3M Center, 235-2B-23 St. Paul MN 55144-1000 > Phone: 612-736-9536 FAX: 612-736-3116 | Device: | | |----------------------|---------------------------------------------------------| | Trade name: | Imation <sup>TM</sup> SE-196 Laser Imager | | Common name: | Laser Printer | | Classification name: | Magnetic Resonance Diagnostic Device<br>21 CFR 892.1000 | 3M(Imation™) 8700 Laser Imager Predicate devices: #### Description And Intended Use of Device: The Imation SE-196 Laser Imager provides high quality hard copy film output from digital imaging source modalities for use in medical imaging diagnosis and referral. Electronic image information signals are managed in the SE-196 and transformed optically to expose Imation imaging media. The system is intended for use with a variety of digital modalities including CT, MR and CR for diagnostic use by medical radiologists and communications to referring physicians and their patients. #### Technological Characteristics: The subject device and predicate devices use the same technical design base. The printers receive image data from the modality. Modality data and printing functions are performed by the IMS(Image management System). User control is performed by a keypad or directly by the modality through the host control. Imation imaging media is removed from a daylight cartridge and transported to the laser imaging station. Image data and media merge at the laser station and the film is scanned. The exposed media is transported through the integrated processor and exits the printer. Software is used to control the image management and machine functions. AIQC(Automated Image Quality Control) matches printing power with film characteristics to provide consistently high image quality. #### Performance Data: Important performance characteristics for medical hard copy devices are spatial frequency response, gray scale resolution and density information. The subject and predicate devices are similar with respect to spatial frequency response with pixel size of both devices being 325 dpi. Gray scale resolution for the predicate is wither 8 or 12 {1}------------------------------------------------ bit and 12 bit only for the subject device. Density uniformity of the subject device is the same or better than the predicate devices have built in density test patterns and AIQC (Automatic Image Quality Control) that maintains density uniformity over time and over changes in the film media. Safety and effectiveness are key activities in the commercialization and are assured via meeting voluntary standards, including UL544, IEC601-1, IEC 825 and Imation™ SE-196 Engineering specification(Part B). With the successful conclusion of both the field test and internal tests for qualification, validation and reliability, the Product Team will approve final release for production (including Software). After final approval, according to the established procedures, the software code label will be changed to production level label. #### Conclusion: The subject device, like the predicates, has no patient contact. The devices also do not control, monitor or otherwise effect any devices directly connected to or effecting the patient. Images displayed by the subject device and its predicates are reviewed by medical personnel, offering ample opportunity for competent human intervention in case of a malfunction or other failure. Films imaged by the subject device display the same or better image properties in the areas of spatial and gray-scale resolution and in density uniformity as the predicate. As with the predicate devices, a test pattern generator and automatic image quality control(AIQC) system are incorporated to assure consistency between input signals and output density. The subject and predicate device(s) have all been designed to the same safety standard, UL1950. Imation therefore concludes that the Imation™ SE-196 Laser Imager is as safe and effective as the predicate device {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows a black and white logo for the Department of Health. The logo features a stylized bird-like symbol with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH" are arranged vertically along the left side of the symbol, with the letters oriented to be read from top to bottom. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Stephen G. Slavens Regulatory Affairs Imation Corp. 1 Imation Place Oakdale, MN 55128-3414 SEP - 5 1997 Re: K972163 Imation™ SE-196 Laser Imager Dated: June 1, 1997 Received: June 9, 1997 Unclassified : Procode: 90 LMC Dear Mr. Slavens: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours htliau Yin Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## ATTACHMENT 2 ### Statement of Indications for Use: 510(K) Number (if known): K972163 Device Name: Imation™ SE-196 Laser Printer Indications for Use: The Imation™ SE-196 laser imager is intended use as a high quality hard copy device for output from digital imaging source modalities for use in medical imaging diagnosis and referral. Electronic image information signals are managed in the SE-196 and transformed optically to expose Imation imaging media. The system is intended for use with a variety of digital modalities including CT, MR and CR for diagnostic use by medical radiologists and communications to referring physicians and their patients. Concurrence of CDRH, Office of Device Evaluation (ODE) Perscription Use × (Per 21 CFR 801.109) OR Over the Counter Use Vorris C. klairr (Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological D 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...