ACS HI-TORQUE ALL STAR 0.014 GUIDE WIRE WITH MICROGLIDE COATING

K972142 · Guidant Corp. · DQX · Aug 25, 1997 · Cardiovascular

Device Facts

Record IDK972142
Device NameACS HI-TORQUE ALL STAR 0.014 GUIDE WIRE WITH MICROGLIDE COATING
ApplicantGuidant Corp.
Product CodeDQX · Cardiovascular
Decision DateAug 25, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). To facilitate the placement of equipment such as compatible stent devices during other diagnostic or therapeutic intravascular procedures.

Device Story

Steerable 0.014" guide wire; 190 cm or 300 cm lengths; used by physicians during intravascular procedures to navigate and place balloon catheters or stents. Features PTFE-coated proximal shaft; Microglide (silicon-based) coated distal working end; 3 cm flexible radiopaque tip (straight or J-shape). Includes TFE shrink tubing segment proximal to tip to maintain 0.014" diameter and reduce friction. Proximal markers at 90 and 100 cm. Operates as manual tool; provides tactile feedback to physician for vessel navigation. Benefits patient by enabling precise delivery of interventional equipment to target coronary or peripheral sites.

Clinical Evidence

Bench testing (distal tip pull, turns-to-failure, rotational accuracy, tip flexibility) and in vivo canine model testing performed. Results demonstrated performance equivalent to predicate device. No clinical data provided.

Technological Characteristics

0.014" diameter steerable guide wire. Materials: PTFE (proximal shaft), Microglide silicon-based coating (distal end), TFE shrink tubing. Radiopaque 3 cm distal tip. Lengths: 190 cm (extendable) and 300 cm (exchange). Manual operation; no energy source or software.

Indications for Use

Indicated for facilitating placement of balloon dilatation catheters during PTCA and PTA, and placement of equipment like compatible stents during diagnostic or therapeutic intravascular procedures. Contraindicated for use in cerebral vasculature and with atherectomy devices.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 11011/2 # 510(k) SUMMARY This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92. | 1. | Submitter's Name: | Guidant Corporation<br>Advanced Cardiovascular Systems, Inc. | | |----|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | | Submitter's Address: | 26531 Ynez Road<br>Temecula, CA 92591-4628 | | | | Telephone: | 909-914-2400 | AUG 25 1997 | | | Fax: | 909-914-6690 | | | | Contact Person: | Wayne R. Hohman | | | | Date Prepared: | June 5, 1997 | | | 2. | Device Trade Name: | ACS HI-TORQUE ALL STAR™ 0.014" Guide<br>Wire with Microglide® Coating | | | | Device Common Name: | Guide Wire | | | | Device Classification Name: | Catheter Guide Wire (74DQX) | | | 3. | Predicate Device: | ACS HI-TORQUE EXTRA S'PORT™ 0.014"<br>Guide Wire with Microglide® Coating<br>(K950156, cleared April 5, 1995, and<br>K942066, cleared July 18, 1994) | | #### Device Description: 4. The ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire with Microglide® Coating is a steerable guide wire with a nominal diameter of 0.014″ and two lengths: a 190 cm extendable length and a 300 cm exchange length. The proximal end of the 190 cm models is tapered to fit into the hypotube portion of the ACS DOC® Guide Wire Extension (K902755, cleared September 4, 1990). The proximal shaft of this guide wire is coated with polytetrafluoroethylene (PTFE) and the distal working end is coated with Microglide® Coating, a silicon based material. The distal end has a 3 cm, flexible, radiopaque tip that is provided either as a straight, shapeable configuration or as a preshaped J configuration. These guide wires have proximal markers at 90 and 100 cm from the distal tip. Two features distinguish the new guide wire from the predicate guide wire: the addition of a segment of tetrafluoroethylene (TFE) shrink tubing around the distal portion of the guide wire proximal to the 3 cm tip and a more flexible 3 cm tip. - 5. Intended Use: The ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire with Microglide® Coating has the following intended uses: {1}------------------------------------------------ - To facilitate the placement of balloon dilatation catheters during . percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). - To facilitate the placement of equipment such as compatible stent . devices during other diagnostic or therapeutic intravascular procedures The ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire with Microglide Coating has the following contraindications: - Use in cerebral vasculature. . - . Use with atherectomy devices. NOTE: The predicate guide wire is indicated for use with atherectomy devices; the new guide wire is contraindicated for use with atherectomy devices. - Technological Characteristics: 6. Comparison of the new and predicate devices shows that technological characteristics such as materials, biocompatibility, performance properties (see below), sterilization, and packaging are identical or substantially equivalent to the currently marketed predicate device. Two design features distinguish the new guide wire: an added segment of tetrafluoroethylene (TFE) shrink tubing around the distal portion of the guide wire just proximal to the 3 cm tip and a more flexible 3 cm tip. The TFE shrink tubing provides a continuous working diameter of 0.014″ in this distal portion of the guide wire. The decreased coefficient of friction of the TFE layer facilitates movement of the guide wire within 0.014" devices. The more flexible tip is designed to be more atraumatic in coronary vessels. - 7. Performance Data: Bench testing was performed to demonstrate that the ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire with Microglide Coating met acceptance criteria and performed similar to the predicate ACS HI-TORQUE EXTRA S'PORT™ 0.014" Guide Wire with Microglide "Coating. Four tests were performed as follows: - . Distal Tip Pull Test - . Distal Tip Turns-to-Failure Test - . Rotational Accuracy Test - Tip Flexibility Test . {2}------------------------------------------------ In vivo animal testing in a canine model with healthy coronary arteries was performed to demonstrate performance properties of the ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire in comparison to the predicate guide wire. The results showed that the ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire performed in a manner equivalent to the ACS HI-TORQUE EXTRA S'PORT™ 0.014" Guide Wire. The results from the four bench tests plus the animal testing showed that the new ACS HI-TORQUE ALL STAR™ Guide Wire met acceptance criteria and performed in a manner equivalent the predicate ACS HI-TORQUE EXTRA S'PORT™ 0.014″ Guide Wire. No new safety or effectiveness issues were raised during the testing program. - 8. Conclusions: Since the new guide wire has the same intended use (minus the atherectomy indication), similar design and technological characteristics, no new materials, equivalent performance properties, identical sterilization and packaging, and no new safety or effectiveness issues, the ACS HI-TORQUE ALL STAR™ Guide Wire with Microglide® Coating may be considered substantially equivalent to the predicate ACS HI-TORQUE EXTRA S'PORT™ Guide Wire with Microglide Coating. {3}------------------------------------------------ Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 Mr. Wayne R. Hohman Senior Regulatory Affairs Coordinator Guidant Corporation Advanced Cardiovascular Systems 26531 Ynez Road, P.O. Box 9810 Temecula, California 92591-4628 AUG 25 1997 K972142 Re: ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire with Microglide® Coating Requlatory Class: II (two) Product Code: DOX Dated: June 5, 1997 Received: June 6, 1997 Dear Mr. Hohman: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### INDICATIONS FOR USE STATEMENT # 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: ACS HI-TORQUE ALL STAR™ 0.014" Guide Wire with Microglide Coating Indications for Use: - . To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). - To facilitate the placement of equipment such as compatible stent devices . during other diagnostic or therapeutic intravascular procedures. Contraindications: - Use in cerebral vasculature. . - . Use with atherectomy devices. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use (Per 21 CFR 801.109) | <div> <img alt="checkbox" src="checkbox"/> </div> | |---------------------------------------|---------------------------------------------------| | OR over-the-counter | | ![signature](signature) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices | 510(k) Number | K972142 | |---------------|---------| |---------------|---------| (Optional Format 1-1-96) 120
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...