K972138 · Carr Metal Products, Inc. · FRG · Nov 13, 1997 · General Hospital
Device Facts
Record ID
K972138
Device Name
CARR RIGID CONTAINER SYSTEM
Applicant
Carr Metal Products, Inc.
Product Code
FRG · General Hospital
Decision Date
Nov 13, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Container system designed to contain medical and dental (excluding dental hand pieces) instruments for sterilization and storage. This device is intended to be used and handled by healthcare personnel. This container system is suitable for Prevacuum steam and EO sterilization. Sterility assurance tested up to 30 days.
Device Story
Rigid sterilization container system; base (various depths: 2.5"-5.0") and universal lid; incorporates filter retainer for disposable filters, silicone seal, and latching mechanism; tamper-evident slot; used by central supply, OR nurses, and dental staff; instruments placed inside for sterilization via pre-vacuum steam or EO; container protects instruments during storage and transport; maintains sterility for at least 30 days; reusable; requires routine inspection after each use.
Clinical Evidence
Bench testing only. System tested per AAMI ST 33, ANSI/AAMI ST 41-1992, and ST 24-1992 standards. Validated for pre-vacuum steam (270°F–275°F) and ethylene oxide (EO) sterilization. Demonstrated sterility maintenance for at least 30 days.
Technological Characteristics
Rigid container system; materials: aluminum (lid/base) with coating; silicone seal; disposable filter media; latching mechanism; dimensions: 2.5", 3.0", 3.5", 4.0", and 5.0" base depths; sterilization methods: pre-vacuum steam and ethylene oxide; reusable.
Indications for Use
Indicated for healthcare personnel in hospitals, nursing homes, surgical centers, clinics, and dental offices to contain medical and dental instruments (excluding dental hand pieces) for sterilization, storage, and transportation via pre-vacuum steam or ethylene oxide (EO) sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
K972138
Carr Metal Products, Inc. 3735 N. Arlington Ave.
## Indianapolis, IN 46218
NOV 1 3 1997
## Non-Coalidential Summary of Safety and Effectiveness
page 1 of 4 November 7, 1997
Tel - (317) 542-0691 Fax - (317) 542-0694
| Official Contact: | Alan Booker - Operations Manager |
|----------------------------|----------------------------------------------------------|
| Proprietary or Trade Name: | Carr Sterilization Container System |
| Common/Usual Name: | Sterilization Container - rigid |
| Classification Name: | Sterilization wrap |
| Device: | Carr Sterilization Container system |
| Predicate Devices: | Aesculap - K792558, Medin - K833343, InstruMed - K844672 |
| | and Riley - K871202 |
#### Device Description:
Carr Metal Products, Inc.
3735 N. Arlington Ave. Indianapolis, IN 46218
A rigid container system include a base available in several depth sizes ( 2.5", 3.0", 3.5", 4.0" and 5.0") and a lid which fits any of the bases. The lid is a standard size which fits any size (depth) of base, it also incorporates a filter retainer unit which holds a disposable filter, has a seal and incorporates a latching mechanism which latches the lid and base together.
- Container system designed to contain medical and dental (excluding dental Indicated Use -hand pieces) instruments for sterilization, storage, and transportation. Suitable for Pre-vacuum steam and EO sterilization. Used and handled by healthcare personnel such as - central supply, operating room nurses, surgical nurses, dental assistants, or healthcare personnel who have responsibility for the cleaning, handling, sterilization and storage of medical and dental (excluding dental hand pieces) instruments.
- Environment of Use -- Used in health care facilities hospitals, nursing homes, extended care facilities, free standing surgical centers, clinics and medical and dental offices. Within these facilities the setting may be where medical and dental (excluding dental hand pieces) instruments which are required to be sterile
{1}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness
## page 2 of 4
## November 7, 1997
are utilized. Examples are - operating rooms - acute, sub-acute, outpatient
are utilized. Examples are - operating rooms - acutes services of a hospital. are utilized. Examples are - operating rooms - a2009.
surgery, physician and dental offices, wards, central services of a hospital.
Summary of Performance Testing --
ce Testing -The Carr Container system was independently tested ecording to AAMI ST
The Carr Container and ST 24-1192 for its performance under The Carr Container system was independently bases
33, ANSI / AAMI ST 41-1992 and ST 24-1192 for its performance under
33, ANSI / AAMI ST 41-1992 and ST 24-1192 for its perfor 33, ANSI / AAMI ST 41-1992 and ST 24-11-2 154 -12-11-11-11
two(2) sterilization methods - Pre-vacuum steam ( (2700 - 2750 F) and two(2) sterilization methods - Pre-vacuum siealis (1275
Ethylone Oxide (EO). It passed all criteria specific to sterilization containers
Ethylone Oxide (EO). It passed all c Ethylene Oxide (EO). It passed as .
and maintained a shelf life sterility of at least 30 days.
The staintained a shelf life sterility of at least 30 days.
maintained a shell like story.
| Comparison to Predicate Devices: | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|-------------------------------------------------------|
| Attribute | Carr Container | Aesculap<br>K792558 or other<br>referenced predicates |
| Use | | |
| Indicated for holding instruments<br>to be sterilized and stored | Yes | Yes |
| Intended to be reused | Yes | Yes |
| Methods of sterilization -<br>pre-vacuum steam (2700F - 2750F)<br>for a minimum of 4 minutes<br>ethylene oxide follow sterilizer<br>manufacturer's recommendations<br>Minimum 12 hours aeration | Yes | Yes |
| Design | | |
| Various sizes(2.5, 3.0, 3.5, 4.0, 5.0") depth<br>of bases offered | Yes | Yes |
| Incorporates a filter to permit entry<br>of steriliant agent and prevent microbial<br>migration during storage | Yes | Yes |
| Utilizes latch system to hold<br>lid in place | Yes | Yes |
| page 3 of 4<br>November 7, 1997<br>Carr Container | | |
| Attribute | | Aesculap<br>K792558 or other<br>referenced predicates |
| Design | | |
| Filter retainer does not have direct align-<br>ment holes preventing accidental puncture | Yes | No |
| media of filter | Yes | |
| Incorporates a silicone seal within lid | Yes | Yes |
| Utilizes trays to hold instruments | Yes | Yes |
| in place | Yes | Yes |
| Has a tamper evident device slot | Yes | |
| Outside labeling slot | | |
| Useful life subject to user handling<br>and routine inspection after each use | Yes | Yes |
| Performance Standards / Specifications | | |
| AAMI ST 33-R-11/95 Standard testing and requirements | Yes | Claimed by Aesculap<br>K792558 |
| 3.2 Permits transfer of contaminated materials | Yes | Claimed by<br>Aesculap K792558 |
| 3.3.1 Removable filter assembly disassembles | Yes | On Aesculap K792558 |
| 3.3.4 Labeling - reusable labels | Yes | Claimed by IntsruMed<br>K844672 |
| 3.4 - 3.5 Instructions provided for different cleaning<br>methods | Yes | Assumed provided |
| 3.6 - 4.2 Instructions for inspections | Yes | |
| 6.2.2 Sterilization - manufacturer documentation<br>available | Yes | Claimed by Medin<br>K833343, InstruMed<br>K844672 |
| 6.2.3 Drying time in labeling | Yes | Assumed |
| 6.2.4 EO residual removal - in labeling | Yes | Assumed |
| 6.2.5 Sterility maintenance - discussed in labeling,<br>claimed in independent testing | Yes | InstruMed K844672 |
{2}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness
Confidential Summary of Safety on
: ・・
{3}------------------------------------------------
# Non-Confidential Summary of Safety and Effectiveness
رالا
## November 7, 1997
| Attribute | Carr Container | Aesculap |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|-----------------------|
| | | K792558 or other |
| | | referenced predicates |
| The first of the first of the country of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the consisted on the | | |
| Performance Standards / Specifications | | |
|--------------------------------------------------------------------------------------------------------------------------------|--------|--------------------------------------------------|
| 6.3 User (basic) responsibilities listed in labeling | Yes | Assumed |
| 7.3.1 Routine inspection in labeling | Yes | Assumed |
| Testing of container under a challenge per<br>ANSI / AAMI ST 41-1992 and ST24-1992<br>for prevacuum steam and EO sterilization | Passed | Claimed by Medin in<br>literature - K833343 |
| Claim of at least 30 days sterility shelflife<br>supported by test data | Yes | InstruMed K844672 |
| Validation study performed to determine<br>that contents would be sterile utilizing pre-<br>vacuum steam and ethylene oxide | Yes | Claimed by Aesculap<br>K792558, Medin<br>K833343 |
| Materials | | |
| Aluminum for lid and base | Yes | Yes |
| Coating for lid and base | Yes | Yes |
| Approved filter media | Yes | Yes |
# Differences Between Other Legally Marketed Predicate Devices
There are no significant differences between the intended device and the predicate - Aesculap approved under K792558, Medin - K833343, InstruMed - K844672 and Riley - K871202.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Alan Booker Operations Manager Carr Metal Products, Incorporated 3735 N. Arlington Avenue Indianapolis, Indiana 46218
NOV 1 3 1997
K972138 Re : Carr Rigid Container System Trade Name: Regulatory Class: II Product Code: FRG Dated: September 5, 1997 Received: September 8, 1997
Dear Mr. Booker:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. →
{5}------------------------------------------------
Page 2 - Mr. Booker
This letter will allow you to begin marketing your device as first feet wire as 0 (k) premarket notification. The FDA described in your sit (i) privalence of your device to a legally Finding or subbeancear equite in a classification for your markets produce, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance obcathed from the Division of Smart Manufaceardin (201) 443-6597 milli its internet address "http://www.fda/gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Timothy A Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
11/12/199/ 12:20 31.1-335-9278
## INDICATIONS FOR USE
Page 1 of 1
510(k) Number: K972138
Carr Rigid Sterilization Container System Device Name:
Container system designed to contain medical and dental Intended Use : (excluding dental hand pieces) instruments for sterilization and storage. This device is intended to be used and handled by healthcare personnel. This container system is suitable for Prevacuum steam and EO sterilization. Sterility assurance tested up to 30 days.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Reorge L. Melle for Olin S. Zin Ph D
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices K977138 510(k) Number .
> Prescription Use (Per CFR 801.109)
OL
Over-the-counter use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.