K972107 · Medtronic Minimed · LZG · Aug 15, 1997 · General Hospital
Device Facts
Record ID
K972107
Device Name
MINIMED INFUSION PUMP, MODEL 507C
Applicant
Medtronic Minimed
Product Code
LZG · General Hospital
Decision Date
Aug 15, 1997
Decision
ST
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 507c is intended for continuous delivery of insulin, at set and variable rates, for the management of dialets mellius in persons requiring insulin. It is not intended for use with blood or blood products.
Device Story
MiniMed 507c is a programmable syringe infusion pump for continuous insulin delivery. Operated by patients or clinicians in home or clinical settings; prescribed by physicians. Device stores basal/bolus delivery history (90 days), event logs (200), and alarm history (50). Features include increased basal rate programming (up to 48 rates), increased delivery capacity (35.0 units/hr), priming bolus tracking, and 'Dual Wave' bolus delivery (immediate percentage plus diminishing remainder over time). Multilingual interface support added. Device provides automated insulin delivery based on user-programmed parameters to assist in diabetes management.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and design modifications.
Technological Characteristics
Syringe infusion pump; ASIC-based electronics; increased RAM (12K); 48 basal rate settings; 35.0 units/hr max delivery; Dual Wave bolus delivery; multilingual interface support. Materials, design, and energy source identical to predicate.
Indications for Use
Indicated for continuous insulin delivery for management of diabetes mellitus in persons requiring insulin.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
MiniMed model 507 Insulin Pump
Submission Summary (Full Text)
{0}------------------------------------------------
Minimed inc. Premarkel Notification [ SIM(k)] 507c Insulin Pump
K9-72107
## Part F. 510(k) Summary
In accordance with the requirements of SMDA 1990, and 21 C.FR 807.92, a 510(k) Summary follows:
Sulvmitter: MiniMed Inc. 12744 Sun Fernando Rd., Sylmar, California 91342
AUG 1 5 1997
Contact: Don Sclvcy, Regulatory Affairs (818) 362-5958, 3011; (520) 527-0107 (v/1).
Name of Device: MiniMcd model 507c Insulin Pump
Predicate Device: MiniMod model 507 Insulin Pump.
Description of the Device: The 507c career insulin punp is a rac-programmable syringe infusion punp, designed for continuous delivery of insulin, at set and variable ratos, as prescribed by the user's plysician. The 507c is restricted to sale by or on the order of a physician. It is not intended or indicated for the delivery of blood or blood products. The principal modifications described in this submission are:
1) Soveral of the purnp electronics have been integrated into an application spacific intograted circuit (ASIC) Approximately 12 K of momory has been added to the Random Access Memory. "The device now stores the lust 90 days of basel and bolus delivery information, compared with three days in the momory will also store the last 200 events and last 50 alarms, as well as the device serial number. 2) The basal rate option has been increasod. The user noy now program up to 48 basal rates, compared with 12, 3) The delivery capacity has been increased. The 507c may deliver up to 35.0 units per hour, compared to 25.0; 4) A sereen has been added to account for the priming bolus. This is a new feature that is intended to discount from the delivery history insulin used to primo the infusion sol(9) but not actually delivered to the patient; 5) A "Dual Wave" bolus delivery pattern has been added, which allows the device were to program the pump to deliver a percentage of the bolus immediately, with the rest to be delivered in a diminishing nauch over time: () A new screen has been added to allow the language the wishes the wishes the pump to use in presenting messages. Currently, only English is offered: novever, in the fitture other language options will be provided. The likely languages include French, German, Spanish, Judius, and possibly others. This Is designed to address the needs of device users more comfortable with languages other than English.
Intended Use of the New Device: The 507c is intended for continuous delivery of insulin, at set and variable rates, for the management of dialets mellius in persons requiring insulin. It is not intended for use with blood or blood products.
Comparison of the Technological Features of the New Device and Prodicate Device: 77hc technological features of the new device do not differ from the prodicate device. Both devices have the same materials, product design, and chergy source. The electronics have been callanced in the now device.
Don Selvey
Regulatory Affairs
MiniMed Inc.
7-31-97
date
12744 Sau Temando Road • Sylmar. CA 91349 • (800) 933 33327 • [816] 364-2246 • • Mrb Site: http://www.minined.com
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 5 1997
Mr. Don Selvey Regulatory Affairs MiniMed Technologies, Incorporated : 12744 San Fernando Road Sylmar, California 91342
Re: K972107 MiniMed Infusion Pump, Model 507C Regulatory Class: II Product Code: LZG Dated: June 3, 1997 Received: June 5, 1997
Dear Mr. Selvey:
This letter corrects our substantially equivalent letter dated August 1, 1997.
We have reviewed your Section 510(k) notification of intent to market the device reference above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (act). The general controls provisions of the act include requirements for registration, listing of devices, good manufacturing practices, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, the Food and Druq Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any
{2}------------------------------------------------
Page 2 - Mr. Selvey
obliqation you miqht have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulation.
Under Section 522(a) of the act, manufacturers of certain types of devices identified by the Act or designated by FDA are required to conduct postmarket surveillance studies. FDA has identified under Section 522(a) (1) (A) the device cleared for marketing by this letter as requiring postmarket surveillance.
Within thirty (30) days of first introduction or delivery for introduction of this device into interstate commerce you are required to submit to FDA certification of the date of introduction into interstate commerce, a detailed protocol which describes the postmarket surveillance study, and a detailed profile of the study's principal investigator that clearly establishes the qualifications and experience of the individual to conduct the proposed study. For your information, general quidance on preparing a protocol for a postmarket surveillance study is attached.
Submit five (5) copies to:
Center for Devices and Radiological Health Postmarket Surveillance Studies Document Center Suite 405 (HFZ-544) 1801 Rockville Pike 20852 Rockville, Maryland
Within sixty (60) days of receipt of your protocol, FDA will either approve or disapprove it and notify you of the Agency's action in writing. You should not begin your postmarket surveillance study of this device until the protocol has been approved. Data generated under an unapproved protocol may not satisfy your obligation under section 522. Please note that you must continue to collect and report data needed to maintain compliance with Medical Device Reporting regulations (21 CFR 803).
Failure to certify accurately the date of initial introduction of your device into interstate commerce, to submit timely an acceptable protocol, or to undertake and complete and FDA approved postmarket surveillance study consistent with the protocol will be considered violations of section 522. In accordance with the Medical Device Amendments of 1992, failure of a manufacturer to meet its obligations under section 522 is a prohibited act under section 301 (q) (1) (C) of the Act (21 U.S.C. 331 (q) (1) (C) . Further, under section 502 (t) (3) of the act (21 U.S.C. 352(t)(3)), a device is misbranded if there is a failure or refusal to comply with any requirement under
{3}------------------------------------------------
Page 3 - Mr. Selvey
section 522 of the act. Violations of sections 301 or 502 may lead to regulatory actions including seizure of your product, injunction, prosecution, or civil money penalties.
If you have questions specifically concerning postmarket surveillance study requirements, contact the Postmarket Surveillance Studies Branch at (301) 594-0639."
In addition, on August 16, 1993 the Final Rule for Device Tracking was published in the Federal Register, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirement of the requlation are found in 21 CFR 821 as described in the August 16, 1993 Federal Register beginning on page 43447.
This letter immediately will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but does not mean that FDA approves your device. Therefore, you may not promote or in any way represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809 for in vitro diagnostic devices), promotion, or advertising, please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-326) at (301) 594-4639. Other general information on your responsibilities under the act may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
by A. Ulatowski Tim Director
Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number: 《S7 2107
Device Name: MiniMed insulin pump, model S07c
Indications for Use: The MiniMed insulin pump, model 507c, is indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellius in persons requiring insulin.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | Patricia Crucente |
|---------------------------------------------------------------------|-------------------|
| Division of Dental, Infection Control, and General Hospital Devices | |
| 510(k) Number | K982107 |
| Prescription Use<br>(Per 21 CFR 801.109) | |
or Over-the-Counter Use్లో
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.