REAADS VON WILLEBRAND FACTOR ANTIGEN TEST KIT

K972005 · Reaads Medical Products, Inc. · GGP · Aug 13, 1997 · Hematology

Device Facts

Record IDK972005
Device NameREAADS VON WILLEBRAND FACTOR ANTIGEN TEST KIT
ApplicantReaads Medical Products, Inc.
Product CodeGGP · Hematology
Decision DateAug 13, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2

Indications for Use

The REAADS von Willebrand Factor Antigen (vWF:Ag) Test kit is an in vitro diagnostic assay for the quantitative determination of human plasma levels of vWF:Ag (as a percent concentration) by enzyme linked immunosorbent assay (ELISA). Plasma levels of vWF:Ag may be used as an aid in diagnosing von Willebrand disease. The REAADS vWF:Ag Test Kit is intended to be used by clinical (hospital and reference) laboratories.

Device Story

The REAADS von Willebrand Factor Antigen (vWF:Ag) Test Kit is an ELISA-based in vitro diagnostic assay. It utilizes 96-microwell polystyrene plates coated with capture antibodies specific for human vWF. Citrated human plasma samples are incubated in the wells, allowing vWF:Ag to bind to the immobilized antibodies. After washing, an HRP-conjugated anti-human vWF detection antibody is added. A chromogenic substrate (TMB and H2O2) is then introduced to produce a colorimetric reaction. The intensity of the color is measured spectrophotometrically at 450nm. The device is intended for use in clinical or reference laboratories by trained personnel. Results are calculated as a relative percent concentration against a provided reference plasma curve. These quantitative results are used in conjunction with functional assay determinations to assist clinicians in diagnosing von Willebrand disease, characterized by vWF levels typically between 0-50% of normal.

Clinical Evidence

Clinical performance was evaluated by comparing the REAADS vWF:Ag Test Kit against the Helena Von Willebrand Factor Antigen Rocket EIA. Testing was performed on clinical samples. Results showed a coefficient of correlation of 0.962. Statistical analysis using single factor ANOVA yielded a p-value of 0.739, indicating the methods are statistically similar. The study concluded that the performance of the two methods is comparable, with minor differences in value recovery attributed to the improved specificity of the ELISA technology.

Technological Characteristics

Sandwich ELISA; 96-microwell polystyrene plates; HRP-conjugated detection antibody; TMB/H2O2 chromogenic substrate; spectrophotometric detection at 450nm. Quantitative assay for vWF:Ag in citrated human plasma.

Indications for Use

Indicated for the quantitative determination of vWF:Ag in human plasma to aid in the diagnosis of von Willebrand disease in patients suspected of having quantitative or functional vWF deficiencies.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## SUMMARY OF 510(K) SAFETY AND EFFECTIVENESS ## REAADS von Willebrand Factor Antigen Test Kit July 30, 1997 AUG 1 3 1997 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The REAADS von Willebrand Factor Antigen (vWF:Ag) Test Kit is compared to a legally marketed predicate device and a substantial equivalence claim is made. The predicate device is the Helena Von Willebrand Factor Antigen Rocket EIA (former name: Helena Factor VIII-Related Antigen Rocket EID System) currently manufactured and marketed by Helena Laboratories, Beaumont, Texas. The REAADS von Willebrand Factor Antigen Test Kit is a sandwich enzyme linked immunosorbent assay (ELISA). The capture antibody specific for human vWF is immobilized to 96-microwell polystyrene plates. Diluted patient plasma is incubated in the wells, allowing any available vWF:Ag to bind to the anti-human vWF antibody bound to the plastic. The plates are rinsed to remove any unbound plasma vWF:Ag molecules. Bound vWF:Ag is quantitated using an HRP conjugated anti-human vWF detection antibody. Any unbound conjugated anti-human vWF is washed away after an incubation period. A chromogenic substrate of tetramethylbenzidine (TMB) and hydrogen peroxide (H2O2) is added to develop a colored reaction. The intensity of the color is measured spectrophotometrically at 450nm in optical density (O.D.) units. vWF:Ag relative percent concentrations of patient plasma is determined against a curve made from a reference plasma provided. The intended use of the device is to quantitatively determine vWF:Ag in citrated human plasma. Normal levels of vWF:Ag are generally accepted to be between 50-160% as compared to a standard or pooled normal plasma. Patients suffereing from von Willebrand Disease can have either quantitative levels or decreased function of vWF between 0 - 50% as compared to a standard or pooled normal plasma. The quantitative results from REAADS vWF ELISA may be used in conjunction with functional assay determinations to establish a diagnosis of vWF. Test results for clinical samples demonstrate that the performance of the REAADS vWF:Ag Test Kit and the Helena Von Willebrand Factor Antigen Rocket EIA Method is substantially equivalent. The coefficient of correlation for the entire population is 0.962, with a P-value of 0.739 (by single factor ANOVA), indicating the results by the two methods are statistically similar. Although a few minor differences in value recovery were observed between the assays, in general the performance was comparable. The differences may be attributed to the improved specificity of REAADS ELISA technology when compared to EIA. Maria Dette 2-30-97 Date Nanci Dexter Director, Quality and Regulatory Affairs {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 AUG 1 3 1997 Ms. Nanci Dexter Director Quality and Regulatory Affairs REAADS Medical Products, Inc. 12061 Tejon Street Westminster, CO 80234 Re : K972005/S1 REAADS® von Willebrand Factor Antigen Test Kit Regulatory Class: II Product Code: GGP Dated: July 30, 1997 Received: August 1, 1997 Dear Ms. Dexter: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as sec forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {2}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as --described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Gutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Statement 510(k) Number: _ Device Name: REAADS von Willebrand Factor Antigen Test Kit Indications for Use: ... ........................................... The REAADS von Willebrand Factor Antigen (vWF:Ag) Test kit is an in vitro diagnostic assay for the quantitative determination of human plasma levels of vWF:Ag (as a percent assay to throncentration) by enzyme linked immunosorbent assay (ELISA). Plasma levels of vWF:Ag may be used as an aid in diagnosing von Willebrand disease. The REAADS vWF:Ag Test Kit is intended to be used by clinical (hospital and reference) laboratories. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Divisio (Din-Off) Division of Clinical Laboratory Devices 510(k) Number K72285 prescription use 8/12/97 CA
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...